Comparative analysis of anterior corneal curvature and astigmatism measurements obtained with three different devices

, PhD, , PhD, , PhD, , PhD & , PhD
Pages 618-624 | Received 30 Apr 2019, Accepted 09 Oct 2019, Published online: 15 Apr 2021
 

Abstract

Background

The objective of this study was to compare the central corneal curvature and astigmatism measurements obtained with three different systems in healthy eyes and to assess the level of interchangeability between them.

Methods

This was a comparative study examining 30 healthy eyes of 30 patients (age 15–53-years). A complete eye examination was performed in all cases including analysis of anterior corneal curvature and astigmatism with three devices: the colour‐LED topography system Cassini (i‐Optics) (CAS), the Scheimpflug‐based system Pentacam (Oculus Optikgeräte) (PTC) and the optical biometer IOL‐Master 500 (Carl Zeiss Meditec) (IOLM). Differences between devices in terms of curvature in the flattest (flat K) and steepest meridians (steep K) as well as in the magnitude (AST) and power vector components of astigmatism (J0 and J45) were evaluated. The interchangeability between devices was evaluated with the Bland–Altman method.

Results

Statistically significant differences between devices were found in steep K and flat K (p < 0.001). No statistically significant differences between devices were found in AST (p = 0.057) and J0 power vector (p = 0.185). However, differences between devices in J45 did reach statistical significance (p = 0.039). Ranges of agreement for curvature measures ranged from 0.123 (flat K, CAS‐PTC) to 0.165-mm (steep K, CAS‐PTC). Ranges for the magnitude of astigmatism were 0.868, 1.059 and 0.739 D in CAS‐IOLM, PTC‐IOLM and CAS‐PTC comparisons, respectively. For J0 and J45, ranges of agreement were below 0.522 D.

Conclusions

Measurements of central corneal curvature and astigmatism obtained with the three devices evaluated cannot be used interchangeably.

ACKNOWLEDGEMENT

David P. Piñero has been supported by the Ministry of Economy, Industry and Competitiveness of Spain within the program Ramón y Cajal, RYC‐2016‐20471. The authors have no proprietary or commercial interest in the medical devices described in this manuscript.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.