Traditional and emerging technologies for washing and volume reducing blood products

, , &
Pages 37-46 | Published online: 03 Jan 2019

Abstract

Millions of blood components including red blood cells, platelets, and granulocytes are transfused each year in the United States. The transfusion of these blood products may be associated with adverse clinical outcomes in some patients due to residual proteins and other contaminants that accumulate in blood units during processing and storage. Blood products are, therefore, often washed in normal saline or other media to remove the contaminants and improve the quality of blood cells before transfusion. While there are numerous methods for washing and volume reducing blood components, a vast majority utilize centrifugation-based processing, such as manual centrifugation, open and closed cell processing systems, and cell salvage/autotransfusion devices. Although these technologies are widely employed with a relatively low risk to the average patient, there is evidence that centrifugation-based processing may be inadequate when transfusing to immunocompromised patients, neonatal and infant patients, or patients susceptible to transfusion-related allergic reactions. Cell separation and volume reduction techniques that employ centrifugation have been shown to damage blood cells, contributing to these adverse outcomes. The limitations and disadvantages of centrifugation-based processing have spurred the development of novel centrifugation-free methods for washing and volume reducing blood components, thereby causing significantly less damage to the cells. Some of these emerging technologies are already transforming niche applications, poised to enter mainstream blood cell processing in the not too distant future.

Introduction

Millions of blood components including red blood cells (RBCs), platelets, and granulocytes are transfused each year in the United States. The transfusion of stored blood products has been associated with adverse clinical outcomesCitation1Citation3 brought on by residual plasma proteins, excess potassium, free hemoglobin, and other contaminants that accumulate during storage.Citation4Citation6 To avoid these effects, washing or volume reduction of blood products is often performed to remove contaminants and improve the quality of a unit before transfusion. These procedures conventionally use centrifugation-based cell processors that separate blood cells from the suspending medium based on the difference in their respective densities. Although widely used, these cell processors have limitations that may outweigh their advantages in some settings. There has, therefore, been a significant research effort in recent years for the development of novel centrifugation-free approaches to wash and volume reduce blood products.

Volume reduction is conventionally performed through the centrifugation of a blood component and the subsequent removal of the supernatant. A reduction in sample volume is often required in instances of platelet transfusions, specifically for neonates and infants, where the recipient is unable to withstand transfusions of large volumes.Citation7 Volume reduction is also used for patients prone to severe transfusion-related allergic reactions brought on by plasma proteins and inflammatory mediators released by residual white blood cells (WBCs).Citation8,Citation9 Blood washing involves the dilution of a blood component and subsequent volume reduction of the diluted unit to a desired concentration of blood cells. While stored RBCs are the most common washed blood product, platelet washing is also indicated for neonates receiving transfusions, patients undergoing chemotherapy, and patients prone to transfusion-related allergic reactions.Citation1,Citation2 Whole blood is washed during cell salvage procedures to remove foreign bodies and contaminants suctioned from the incision site. Any blood product cryopreserved to prolong its shelf life is also washed to remove toxic preservatives.Citation10,Citation11 The ultimate goal of volume reduction and washing is to provide an adequately pure sample of a component of blood for transfusion or transplantation, with reduced risk for adverse clinical outcomes. Here we present various methods for blood washing and volume reduction of cellular blood products and highlight new developments in this field (advantages and disadvantages of the technologies are summarized in ).

Table 1 A summary of the advantages and disadvantages of each of the technologies described in this study

Centrifugation-based technologies

Manual centrifugation

Manual volume reduction and washing typically involves pelleting blood cells in a unit using a swinging bucket centrifuge, and discarding the supernatant. As previously mentioned, neonates and infants require volume-reduced platelet concentrates to avoid large transfusion volumes and reduce the risk of fetomaternal alloimmune thrombocytopenia. Volume reduction can also reduce transfusion-related allergic reactions in susceptible patients.Citation12 Since there is currently no standard for the preparation of volume reduced platelets, manual centrifugation using a high-capacity centrifuge is most often used.Citation13,Citation14 However, since volume reduction procedures are not standardized, there are varying spin speeds reported to concentrate platelets. The mechanical stresses placed on platelets, via centrifugal forces, have been shown to lead to significant platelet activation.Citation15 Therefore, discrepancies in spin speeds make the process of manual centrifugation less robust and produces platelet concentrates with wide ranges of quality and purity.Citation16 Furthermore, the manual centrifugation of blood components adds an additional layer of user manipulation, increasing the risk of contamination.Citation12 While the procedure of manually washing platelets is somewhat laborious and has limitations, when performed correctly, the process can yield high recoveries at a fraction of the cost of automated processing systems.Citation15

Automated systems: open and closed

Most blood banking centers utilize automated systems to wash or volume reduce the blood products they produce and distribute. Automated systems are pre-programmed with a variety of cell processing procedures, from density gradient separation to the deglycerolizing of thawed products. In a survey of 15 children’s hospitals (150–400 beds) across America, 14 out of 15 hospitals used the COBE 2991 cell processor (Terumo BCT, Lakewood, CO, USA), making it one of the most popular processors in the field.Citation17 This centrifugation system separates each blood component by their respective density and each layer of fluid is expressed from the disposable insert by connected tubing. After separation, normal saline can be added to dilute the blood product, which is then removed via a connector tubing to obtain a volume-reduced and/or washed product.

RBCs are known to experience significant deterioration while in storage leading to leakage of potassium and free hemoglobin into the supernatant.Citation6,Citation18 For this reason, transfusion of older units (stored longer than 7 days) is often not recommended for neonates and infants undergoing cardiac surgery, in order to prevent hyperkalemic cardiac arrest and other complications. Transfusion of older units is also contraindicated for infants born preterm who suffer from severe anemia.Citation19,Citation20 Washing RBC units using COBE 2991 has been shown to reduce pro-inflammatory markers and to decrease the number of transfusions necessary for pediatric patients undergoing cardiac bypass.Citation21 Cortés-Puch et al also found, in a canine study, that washing RBC units approaching their expiration date (42 days) using the COBE 2991 system improves outcomes when compared to unwashed 42-day-old RBC units.Citation22 This result indicates that washing older units for later transfusion can be advantageous when the supply and availability of fresh blood is low. This same study, however, also found that washing 7-day-old RBCs led to increased organ failure and mortality and that the quality of 42-day-old RBCs could not be restored to the level of the 7-day-old, unwashed RBCs.Citation22 These findings suggest that the COBE 2991 washing system should be reserved for RBC units nearing the end of the allowable storage.

Although stored RBCs washed using the COBE 2991 can be transfused within 24 hours, O’Leary et al found that the concentration of extracellular potassium in washed units at 24 hours actually exceeds prewash potassium levels, and RBC hemolysis increases steadily over the 24-hour post-wash storage,Citation17,Citation23 calling for a reduction in expiration time from 24 hours to 6 hours. For patients receiving these washed units, the excess free hemoglobin can exacerbate clinical complications and lead to more negative prognoses. For cryopreserved blood products (eg, hematopoietic stem cells [HSC]), dimethyl sulfoxide (DMSO) that is used to prevent the formation of ice crystals must be washed off prior to transfusion because of its toxicity.Citation10 Although cryopreserved products are typically washed using automated cell processors like COBE 2991, the system is unable to sufficiently reduce the DMSO concentration in a single run.Citation24

Unlike open cell-processing systems, closed systems, such as the ACP 215 (Haemonetics Corporation, Braintree, MA, USA), significantly reduce the risk of bacterial contamination and allow for the addition of storage media to washed units. The addition of storage media prolongs the shelf life of washed RBCs from 24 hours to 7 days, while still meeting all standard regulatory metrics for washed units.Citation25,Citation26 In clinical studies, the use of the ACP 215 extended the shelf life of RBC units and led to a decrease in the number of units that had to be discarded due to expiration.Citation25 Unfortunately, both COBE 2991 and ACP 215 were found to have caused significant RBC damage, indicating that centrifugation-based methods of washing with cell processors can reduce the quality of washed RBCs. Irradiating RBCs is also known to cause similar membrane damage, potassium leakage, and hemolysis,Citation27 even though the technique is necessary to prevent transfusion-associated graft-versus-host disease and is indicated for neonates. Imposing additional stress on irradiated blood products by centrifugation may also worsen the quality of irradiated products further.Citation23,Citation28,Citation29

Centrifugation-based cell processors are unable to consistently reduce the concentration of IgA, below the level acceptable for transfusion to IgA-deficient patients, in a single-wash run. While washing RBCs twice with the ACP 215 led to a significant decrease in IgA, incorporating a second washing step introduced further damage to the cells and increased the total processing time.Citation30 Additional evidence suggests that washed deglycerolized RBCs have hemolysis levels that are not consistently within the range of regulatory standards after washing.Citation31 ACP 215 has also been used to wash and concentrate platelets with a reported 99% plasma removal rate, compared to manual centrifugation, which yielded 95%–97% plasma removal.Citation32,Citation33 However, the intense mechanical stress associated with passing through a centrifugation-based cell processor inevitably activates plateletsCitation34 as well as residual WBCs.Citation35 These activated cells adhere to parts of the centrifugation chamber and tubing, and continually release cytokines and pro-inflammatory markersCitation36 into the suspending medium of the washed blood product. Neither COBE 2991 nor ACP 215 can remove these activated cells and inflammatory markers, consequently diminishing the quality of washed blood products and causing a range of adverse outcomes at the point of care.Citation37

Autotransfusion systems: discontinuous and continuous flow

Intraoperative blood salvage for autotransfusion decreases the need for allogeneic blood through the recovery and subsequent reinfusion of a patient’s own blood shed during surgery. Blood salvage is particularly useful when a substantial (>20%) blood loss is anticipated, such as in cardiothoracic or orthopedic surgery, and during transplantations.Citation38,Citation39 Auto-transfusion is also utilized for patients who cannot accept donated blood for cultural or religious reasons.Citation39 During the cell salvage process, shed blood is aspirated from the surgical site and collected in a container with anticoagulant (heparin or citrate). Because the blood is suctioned from the surgical cavity and blood loss varies greatly depending on the patient and the type of surgical procedure, a wide range of contaminants (eg, fat, coagulation factors, skin and bone cells, excess plasma, and colloid solution used for fluid resuscitation) are also collected with shed blood. Therefore, salvaged blood is filtered, and then washed and volume reduced to generate a unit of concentrated RBCs before it can be reinfused back into the patient. Because of this functionality, the machines used for processing salvaged RBCs can also be used for washing and volume reduction of allogeneic cellular blood products.

A typical autotransfusion device suctions blood into a bowl (eg, Latham, Baylor, or disk shaped) which accommodates the variable intake of blood during the salvage process. Because the blood is exposed to air and high pressure during suction, massive RBC hemolysis and activation of platelets and WBCs may occur. Activated cells adhere to the surface of the bowl during centrifugation and release leukoattractants and thromboplastic materials into the unit.Citation37 In a comparative study, whole blood washed with a discontinuous-flow autotransfusion device, Cell Saver Elite (Haemonetics Corporation), was found to have more hemolysis and microparticle formation than blood washed with COBE 2991,Citation40 likely from the exposure of blood to both air and high suction pressures. However, in terms of potassium, free hemoglobin, platelet, and WBC removal, autotransfusion machines, such as the Cell Saver 5 (Haemonetics Corporation), have demonstrated a level of performance on par with conventional cell processors.Citation37

The washing/volume reduction performance varies among autotransfusion devices even when they used the same Latham-bowl technology – for example, the Cell Saver 5 produces RBCs with lower hematocrits while achieving a higher RBC recovery than Sequestra 1000 (Medtronic, Minneapolis, MN, USA).Citation41 Still, an even more significant source of variability in cell recovery rates is the ability to manually set and change centrifugation parameters,Citation42 and therefore, the ultimate impact from processing on the quality of washed RBC product would likely depend on user preferences.

Autotransfusion devices based on the pneumatic disk technology, such as the OrthoPAT (Haemonetics Corporation), have a disposable disk of tunable size that allows for optimal processing of salvaged blood regardless of volume. Although these devices can process smaller volumes, the washing procedure typically takes longer than when performed using the conventional Latham bowl-based devices.Citation43,Citation44 Additionally, both pneumatic disk and Latham bowl-based discontinuous-flow devices retain a significant volume of blood in the bowl and tubing,Citation43,Citation44 which further decreases the yield of washed RBCs. Regardless, the ability to process small volumes of blood significantly expands the applicability of autotransfusion devices.Citation43

Unlike the discontinuous-flow devices that must employ multiple disposable inserts to wash any volume of blood larger than the bowl size, continuous-flow technologies can fill, wash, and separate blood components continuously in a single disposable washing chamber. Fresenius Continuous Autotransfusion System (CATS; Terumo BCT) is the first Food and Drug Administration approved and commercially available autotransfusion device that washes and separates salvaged RBCs via continuous-flow centrifugation. Blood is driven through the machine and allowed to separate in a rotating double spiral washing chamber. During rotation, RBCs are forced into the outer wall of the spiral, while all other components move towards the inner wall. Platelets, plasma, and WBCs are removed as washing solution (eg, normal saline) is added to the concentrated RBCs. CATS can concentrate RBCs from whole blood or from stored RBC units before resuspension in normal saline to the desired hematocrit (and can also separate platelet-rich plasma from whole blood and concentrate platelets). The CATS disposable washing chamber requires a minimum fill volume of 30 mL, allowing small-volume washing without any significant modifications to the system. Since the chamber is constantly rotating during processing, there are virtually no residual RBCs left in the washing chamber, providing a better recovery than conventional cell processors and autotransfusion devices that utilize a bowl or pneumatic disk.

In a direct comparison, CATS system generated washed RBCs with unchanged potassium levels and little hemolysis, unlike units washed with the COBE 2991.Citation45 CATS was also able to produce RBC concentrates at a higher hematocrit than COBE 2991 and most other conventional discontinuous-flow autotransfusion devices.Citation37,Citation45 When testing for the removal of fat cells from salvaged blood, CATS surpassed discontinuous systems (eg, Cell Saver 5) and was on par with the Cell Save Elite featuring an integrated fat reduction program.Citation46 Although the CATS system represents a significant improvement upon current centrifugation-based cell processors and discontinuous-flow autotransfusion devices, it is not without limitations. Washing RBCs with CATS yields inconsistent levels of free hemoglobin removalCitation41 and has a negative effect on both RBC deformability and cell 2,3-DPG content after processing.Citation47 Similar to other centrifugation-based cell processors, the negative effects of CATS processing on RBC properties may increase in a time-dependent manner following the washing procedure.Citation48

Although autotransfusion devices are traditionally used to wash fresh blood shed during surgery, washing stored RBCs via autotransfusion has also been explored. Washing with the discontinuous-flow Dideco Electa 740E (Dideco SPA, Mirandola, Italy) aged stored RBCs. De Vroege et al found that ATP and 2,3-DPG contents of RBCs were reduced after washing. Furthermore, while there was free hemoglobin in the waste bag, free hemoglobin levels were not reduced in the final washed unit.Citation49 This indicated that the washing procedure potentially damaged and lysed some RBCs resulting in the release of intracellular hemoglobin. It is possible that the constant deterioration RBCs undergo in storage makes them more susceptible to lysis during centrifugation-based processing. Interestingly, washing stored RBCs with the Cell Saver also did not reduce the amount of microparticles in the final unit,Citation50 while CATS system was able to remove microaggregates with a wide range of sizes.Citation50 Considering autotransfusion technologies were developed to wash relatively fresh salvaged blood, their design and default settings may not be optimal for washing stored RBCs. Furthermore, autotransfusion utilizes specific spinning regimes to remove contaminants not typically found in stored RBC units (eg, fat, bone, etc) and, therefore, may not be completely safe for stored, fragile RBCs.

Finally, autotransfusion devices have also been employed to separate platelets from platelet-rich plasma. Three types of autotransfusion devices utilizing either a Latham bowl, Baylor bowl, or closed-spiral chamber, showed only 50%–60% platelet recovery,Citation51 which was consistently lower than that for conventional centrifugation.Citation52 The platelet recovery performance of autotransfusion devices will have to be significantly improved before they could provide a practical alternative to specialized cell processors or manual centrifugation.

Centrifugation-free technologies

Membrane filtration

Membrane filtration technologies present an attractive alternative to the centrifugation-based cell processors for volume reduction and washing of blood products because membrane-based filtration systems typically do not have moving parts and are significantly simpler and less expensive. Simple filtration, such as using a hollow fiber membrane cartridge, is usually employed for filtering out excessive amounts of the suspending media (such as washing solution or plasma) and retaining the blood cells in suspension. A hollow fiber device, the PSN 120 (Baxter International Inc., Deerfield, IL, US) made from polysynthane, was successfully used to remove 95% of DMSO from cryopreserved platelets, showing a level of performance similar to that of the conventional centrifugation-based washing.Citation53 In comparison, a cellulose-acetate dialyzer (#23.05; Baxter International Inc.) was only able to remove 52% of DMSO from cryopreserved platelets, highlighting the importance of the material that the hollow fiber device employs.Citation53 Both polysynthane and polyacrylonitrile-based (800HF, AKZO; Mission Medical, Fremont, CA, USA) hollow fibers were found to cause a level of platelet activation (assessed via P-selectin expression) similar to that of conventional centrifugation.Citation53 It is likely, therefore, that hollow fiber filtration devices made of these materials may also be suitable for washing and volume reducing conventional platelet concentrates.

Brune et al presented a gravity-based, hollow fiber filtration system utilizing a Sangofer (Sangofer neonatal; Heim Group, Gladbeck, Germany) hollow fiber system for separating RBCs from placental whole blood of full-term newborns. Approximately 60 mL or more of blood from a donation bag flows, via gravity, through a leukoreduction filter and then passes through a hollow fiber system that separates plasma from RBCs. The RBCs reached a final hematocrit of ~56% and were on par, in terms of quality, with adult RBCs, both stored for 35 days.Citation54 Based on this early work, ErySep Classic Blood Separation System (Lmb Technologie GmbH, Schwaig, Germany) and 3M Blood Separation System (3M, Maplewood, MN, USA) are two commercially available hollow fiber systems for separating RBCs and plasma from whole blood. Both the ErySep and 3M Blood Separation Systems operate by gravity and consist of a sampling bag, a whole blood bag, and two collection bags for RBCs and plasma, as well as an attached leukoreduction filter. 3M can process a unit of whole blood within 64 minutes and produce a 293-mL RBC unit.Citation55 ErySep takeŝ140 minutes to produce a 306-mL RBC unit.Citation56 Although RBCs met all regulatory standards, platelets are completely lost during processing, and plasma produced using these hollow fiber systems has significantly diminished concentrations of important clotting factors (eg, von Willebrand factor, antithrombin III, factor V, factor VIII, and factor XI).Citation55Citation58 This important limitation notwithstanding, it is likely that these hollow fiber systems based on polyethersulfone membranes could be readily adapted for washing or volume reducing stored RBCs.

Recent developments in microfiltration utilize membranes for the absorption of unwanted solutes located in solution, such as excess potassium, free hemoglobin, and clotting residuals. Potassium absorbing filters (PAF), developed by Yamada et al have been shown to effectively reduce the amount of potassium in packed RBCs, decreasing the risk of hyperkalemia posttransfusion.Citation59 Such a filter could ameliorate the buildup of potassium that occurs after centrifugation or irradiation.Citation60 These filters have also been shown to remove excess ammonium from RBC units (76%–87% reduction), fresh frozen plasma (21%–31% reduction), platelets (53% reduction), and albumin solutions (49%–92% reduction) reducing the risk of ammonium acting as a neurotoxin during traumatic brain injury or in neonates with underdeveloped blood–brain barriers.Citation61 However, PAF filters are not currently compatible with rapid blood transfusion devices, and more research is still needed to determine the effect of PAF filtration on RBC quality.Citation62,Citation63 Although evidence is limited on the clinical feasibility and advantages of incorporating PAF filtration to standard blood transfusions, the apparent permissible flow rates could allow for high-throughput blood banking applications beyond the specific clinical significance of this technology for neonatal and pediatric hematology practices.

HemoSep (Brightwake, Nottinghamshire, UK) is a hemoconcentration system developed to soak up to 250 mL of excess plasma during blood salvage without the use of centrifugation.Citation64Citation66 The primary component of the HemoSep system is a blood bag separated into two compartments by a filter membrane, which prevents cells in the blood compartment of the bag from entering the super-absorber pad contained in the other compartment of the bag. The blood plasma and other fluids are free to pass through the membrane, however, to be soaked up by the super-absorber pad. Robertson et al developed a sensor capable of measuring the hematocrit of packed RBCs, further increasing the utility of HemoSep hemoconcentration system.Citation67 The low-cost, simplicity, and portability of the HemoSep system in comparison to conventional autotransfusion devices makes it an attractive alternative for blood salvage, particularly in resource-limited and military settings.Citation68 Furthermore, HemoSep has been shown to provide a superior platelet recovery in a direct comparison to CATS (although CATS system was found to concentrate RBCs significantly better than HemoSep).Citation69 Systems implementing physical filters for washing or volume reduction of RBCs (such as the HemoSep bag) provide an exciting alternative to conventional centrifugation-based machines, while also potentially reducing the damage to cells during processing.

Spinning membrane filtration

Filtration using spinning membrane is another method used for blood cell fractionation. Spinning membrane separators provide excellent filtration rates by generating Taylor vortices in the gap between the membrane and the shell of the device. This unique flow pattern creates flow at the membrane, constantly sweeping the surface to prevent the cellular components from depositing on and clogging or fouling the membrane, while continually replenishing the medium to be filtered.Citation70 CytoMate (Baxter International Inc.) is an automatic cell washing device that utilizes a spinning membrane to ensure cell filtration against a counter-flow buffer circulation.Citation71 The closed system has automatic user-definable programming that allows for efficient separation with low risk of contamination.Citation72 CytoMate has been used for a wide variety of applications such as for washing and recovery of viable cytokine-induced killer cells after extended culture,Citation71 for removal of cryoprotective agents from HSC before infusion,Citation72 and for reducing platelet contamination prior to selection of immunomagnetic CD34+ stem cells used in autologous transplants.Citation73 CytoMate was shown to reproducibly remove 96% or more of DMSO from frozen HSC.Citation74 With such a high efficiency, it is conceivable that CytoMate could also be very effective for washing and volume reducing cellular blood components as well, although a special protocol will likely need to be developed to limit RBC lysis and platelet activation within the system.

The LOVO (Fresenius Kabi, Bad Homburg vor der Höhe, Germany) device similarly incorporates a non-pelletizing spinning membrane, made of 4-µm track-etched polycarbonate, to rapidly separate out target cells, remove supernatant, and perform dilution or concentration of cells.Citation75 LOVO has been utilized to achieve automated washing and purification of various cellular products. Schlinker et al adapted the LOVO device to separate in vitro-derived platelets from nucleated cells and later recover immature megakaryocytes for expansion in culture.Citation76 Following membrane filtration, they were able to obtain platelets with high purity and low activation, and good recovery of megakaryocytes. LOVO has also been used to wash mobilized peripheral blood leukapheresis products for CD34+ cell isolationCitation77 and to remove the cryoprotectant from hematopoietic progenitor cell grafts for bedside infusion.Citation75 The use of LOVO for volume reduction and washing of classical transfusion products (RBCs, platelets, and granulocytes) may significantly reduce the cellular damage associated with centrifugation-based processing. However, similar to the centrifugation-based cell processors, LOVO requires a relatively large capital investment and costly disposable cartridges to operate, and therefore, may not be available in resource-limited settings.

Novel centrifugation-free technologies

Washing stored RBCs through centrifugation-free methods can prove challenging since RBCs have a dynamically changing shape and can pass through very narrow pores.Citation78 This makes size-based filtration for washing and volume reduction of RBCs difficult to achieve. Many methods to filter out RBCs without centrifugation use an aggregation agent, like hydroxyethyl starch or dextran.Citation79,Citation80 These agents accelerate aggregation of RBCs, allowing them to form rouleaux and sediment faster in normal gravity at the bottom of the container, while leaving platelet-rich plasma at the top layer. Our laboratory has developed a washing system that utilizes RBC sedimentation in normal gravity (but does not require any added aggregation agents) to wash stored RBCs, producing a product with a final hematocrit of about 37% while also removing 80% of free hemoglobin and 90% of total free protein.Citation81 In this system, a dilute suspension of stored RBCs is flowing through a coiled narrow bore tubing, oriented in such a way as to allow RBC sedimentation perpendicular to the direction of flow. A series of bifurcations built-in along the length of the tubing siphons off washing waste while allowing sedimented RBCs to continue flowing in the tubing until they are collected at the end of the system.

Further development has made the next generation of our centrifugation-free washing system significantly faster and more efficient, by incorporating “inertial focusing” (to enable high-throughput separation of washing solution from washed RBCs), and by adding a membrane pouch with a biocompatible hydrogel to absorb the excess fluid and concentrate washed RBCs to the desired hematocrit.Citation82 This washing system was able to reduce IgA level to about 0.02 mg/dL in a single pass,Citation82 while conventional centrifugation-based washing typically requires two cycles to reduce IgA levels to a concentration that is safe for IgA deficient recipients (<0.05 mg/dL).Citation30 Importantly, our centrifugation-free washing system significantly reduced the buildup of extracellular potassium and free hemoglobin in the supernatant of washed RBCs during the 24-hours short-term storage. In fact, the levels of potassium and free hemoglobin in samples washed using our system after 24 hours were as good as in paired samples immediately after centrifugation-based washing.Citation83 Excess potassium and free hemoglobin are markers of RBC damage and deterioration, and our findings, therefore, indicate that washing stored RBCs with our centrifugation-free system leads to less damaged RBCs in comparison to conventional centrifugation-based cell processing.

While more research and development is needed before our centrifugation-free washing system becomes truly competitive, it may address key limitations of centrifugation-based washing systems, including high cost, complexity, and the significant damage to cells occurring during centrifugation. Our system also has several important advantages over existing centrifugation-free approaches. For example, unlike hollow fiber filtration, our system is able to remove, concurrently, both large contaminants (eg, microparticles and cellular debris) and soluble contaminants (eg, potassium, free hemoglobin, and IgA) when washing stored RBCs. An important limitation of our work so far is that our centrifugation-free washing system has been tested only with relatively small volumes of blood (~50 mL), and increasing the capacity (and throughput) of the system further will require multiplexing several of the separation tubing coils (which is a nontrivial engineering task). However, even the current small-volume prototype of our system could be useful for neonatal and infant patients, who require relatively small volumes when transfused. For these patients, conventional washing using most centrifugation-based cell processors is not practical due to the large dead volume within the disposable sets.

Conclusion

The washing and volume-reduction of blood products may prevent a host of clinical complications caused by the adverse effects associated with residual plasma proteins, and other contaminants found in units of blood component after processing and storage. Here, we reviewed several traditional cell processing techniques based on centrifugation, including manual centrifugation, automated cell processors, and various autotransfusion devices. Although these methods are commonly used to process blood, there is accumulating evidence that centrifugation leads to increased hemolysis, activation of platelets and leukocytes, cellular damage, and various other negative effects that reduce the quality and efficacy of cellular blood products. We also discussed several classes of centrifugation-free approaches, including membrane-and inertia-based separation systems that aim to produce high-quality washed or volume reduced blood products with minimal cellular damage. While these centrifugation-free approaches are very promising, they typically serve niche applications and many are still in early stages of development. Further research is needed to move these emerging technologies into the mainstream of blood cell processing, to provide the highest quality blood products for transfusion and ultimate improve clinical outcomes for the patients.

Acknowledgments

This work was supported in part by a 2012 NIH Director’s Transformative Research Award (NHLBI R01HL117329, PI: SSS). Study sponsor had no role in the writing of the article, and in the decision to submit the paper for publication.

Disclosure

The authors report no conflicts of interest in this work.

References

  • CholetteJMPietropaoliAPHenrichsKFLonger RBC storage duration is associated with increased postoperative infections in pediatric cardiac surgeryPediatr Crit Care Med201516322723525607740
  • RemyKEHallMWCholetteJMechanisms of red blood cell transfusion-related immunomodulationTransfusion201858380481529383722
  • KarafinMSCarpenterEPanASimpsonPFieldJJOlder red cell units are associated with an increased incidence of infection in chronically transfused adults with sickle cell diseaseTransfusion and Apheresis Science201756334535128279592
  • KentMWKelherMRWestFBSillimanCCThe pro-inflammatory potential of microparticles in red blood cell unitsTransfus Med201424317618124786047
  • MakrooRNRainaVBhatiaAEvaluation of the red cell hemolysis in packed red cells during processing and storageAsian J Transfus Sci2011511521572708
  • FlattJFBawazirWMBruceLJThe involvement of cation leaks in the storage lesion of red blood cellsFront Physiol2014536521424987374
  • CrightonGLEstcourtLJWoodEMA therapeutic-only versus prophylactic platelet transfusion strategy for preventing bleeding in patients with haematological disorders after myelosuppressive chemotherapy or stem cell transplantationCochrane Database Syst Rev20151210CD010981
  • AnaniWTriulziDYazerMHQuLRelative IgA-deficient recipients have an increased risk of severe allergic transfusion reactionsVox Sang2014107438939225220631
  • TobianAASavageWJTischDJThomanSKingKENessPMPrevention of allergic transfusion reactions to platelets and red blood cells through plasma reductionTransfusion20115181676168321214585
  • ChenCSharmaSDooHWashing stem cells cryopreserved in Dimethyl Sulfoxide (DMSO) before infusion prevents DMSO-related side effects with little loss of CD34 positive cells and prompt engraftmentAm Soc Hematology2004
  • AbonnencMPesseBTissotJDBarelliSLionNAutomatic washing of thawed haematopoietic progenitor cell grafts: a preclinical evaluationVox Sang2017112436737828337763
  • HirayamaJAzumaHFujiharaMComparison between bacterial growth in platelets (PLTs) washed with M-sol and that in PLT-rich plasmaTransfusion20115171592159421736575
  • SchoenfeldHMuhmMDoepfmerURKoxWJSpiesCRadtkeHThe functional integrity of platelets in volume-reduced platelet concentratesAnesth Analg20051001788115616055
  • VeeraputhiranMWareJDentJAA comparison of washed and volume-reduced platelets with respect to platelet activation, aggregation, and plasma protein removalTransfusion201050
  • Fernández-MuñozHCastilla-LlorenteCPlazaEMQuality assessment and transfusion efficacy of buffy coat-derived platelet concentrates washed with platelet additive solutionBlood Transfus201816327328488971
  • BrambillaMRossettiLZaraCDo methodological differences account for the current controversy on tissue factor expression in platelets?Platelets201829440641428643538
  • O’LearyMFSzklarskiPKleinTMYoungPPHemolysis of red blood cells after cell washing with different automated technologies: clinical implications in a neonatal cardiac surgery populationTransfusion201151595596021091957
  • RapidoFBrittenhamGMBandyopadhyaySProlonged red cell storage before transfusion increases extravascular hemolysisJ Clin Invest2017127137538227941245
  • SwindellCGBarkerTAMcguirkSPWashing of irradiated red blood cells prevents hyperkalaemia during cardiopulmonary bypass in neonates and infants undergoing surgery for complex congenital heart diseaseEur J Cardiothorac Surg200731465966417291775
  • LeeACReduqueLLLubanNLNessPMAntonBHeitmillerESTransfusion-associated hyperkalemic cardiac arrest in pediatric patients receiving massive transfusionTransfusion201454124425423581425
  • CholetteJMHenrichsKFAlfierisGMWashing red blood cells and platelets transfused in cardiac surgery reduces post-operative inflammation and number of transfusions: Results of a prospective, randomized, controlled clinical trialPediatr crit care med201213329029921926663
  • Cortés-PuchIWangDSunJWashing older blood units before transfusion reduces plasma iron and improves outcomes in experimental canine pneumoniaBlood201412391403141124366359
  • WeiskopfRBSchnappSRouine-RappKBostromAToyPExtracellular potassium concentrations in red blood cell suspensions after irradiation and washing (vol 45, pg 1295, 2005)Transfusion2006461461
  • Chiche-LapierreCETramalloniDChaputNLapierreVComparative analysis of sepax S-100, COBE 2991, and manual DMSO removal techniques from cryopreserved hematopoietic stem cell apheresis productCytotherapy2016186S47
  • HansenAYiQ-LAckerJPQuality of red blood cells washed using the ACP 215 cell processor: assessment of optimal pre- and postwash storage times and conditionsTransfusion20135381772177923521180
  • AckerJPHansenALYiQLIntroduction of a closed-system cell processor for red blood cell washing: postimplementation monitoring of safety and efficacyTransfusion2016561495726444143
  • HarmSKRavalJSCramerJWatersJHYazerMHHaemolysis and sublethal injury of RBCs after routine blood bank manipulationsTransfus Med201222318118522188550
  • WeissDRGoehringJWeisbachVCentrifugation after irradiation of red blood cells does not accelerate haemolysisClin Lab2011577-852352621888016
  • SerranoKPambrunCLevinEDevineDVSupernatant reduction of stored gamma-irradiated red blood cells minimizes potentially harmful substances present in transfusion aliquots for neonatesTransfusion201757123009301828782124
  • HansenALTurnerTRKurachJDAckerJPQuality of red blood cells washed using a second wash sequence on an automated cell processorTransfusion201555102415242125988774
  • TurnerTHansenAKurachJAckerJPFrom development to implementation: adjusting the hematocrit of deglycerolized red cell concentrates to meet regulatory standardsTransfusion Medicine and Hemotherapy2017441303828275331
  • IwamaAHirayamaJNogawaMComparison between in vitro properties of washed platelet concentrates suspended in M-sol and those in BRS-A, both of which were prepared with an automated cell processorTransfusion and Apheresis Science201756224124428237840
  • OikawaSMinegishiMEndoKKawashimaWSuzukiKShimizuHWashing of platelets can be fully automated using a closed-system cell processor and BRS-A platelet additive solutionVox Sang2016111443744027509576
  • SöderströmACNyboMNielsenCVinholtPJThe effect of centrifugation speed and time on pre-analytical platelet activationClinical Chemistry and Laboratory Medicine201654121913192027227708
  • Dohan EhrenfestDMPintoNRPeredaAThe impact of the centrifuge characteristics and centrifugation protocols on the cells, growth factors, and fibrin architecture of a leukocyte- and platelet-rich fibrin (L-PRF) clot and membranePlatelets201829217118428437133
  • SchneiderSORensingHHartmannLGrundmannUVolkTBiedlerAImpact of intraoperatively salvaged and washed blood on stimulated cytokine release in vitroTransfusion20145410pt22782279025294235
  • SerrickCJScholzMPartial bowls using the Haemonetics Cell Saver 5: does it produce a quality product?J Extra Corpor Technol200537216116117453
  • YoonCNohSLeeJCKoSHAhnWKimHCInfluence of the washing program on the blood processing performance of a continuous auto-transfusion deviceJournal of Artificial Organs201417111812224292854
  • EsperSAWatersJHIntra-operative cell salvage: a fresh look at the indications and contraindicationsBlood Transfus20119213921251468
  • Bennett-GuerreroEKirbyBSZhuHHermanAEBandarenkoNMcmahonTJRandomized study of washing 40- to 42-day-stored red blood cellsTransfusion201454102544255224735194
  • SerrickCJScholzMMeloASinghONoelDQuality of red blood cells using autotransfusion devices: a comparative analysisJournal of Extracorporeal Technology2003352834
  • NaumenkoKSKimSFCherkanovaMSNaumenkoSEThe Haemonetics Cell Saver 5 washing properties: effect of different washing pump and centrifuge speedsInteract Cardiovasc Thorac Surg20087575976318635585
  • WatersJHTuohyMJHobsonDFProcopGBacterial reduction by cell salvage washing and leukocyte depletion filtrationAnesthesiology200399365265512960550
  • SeyfriedTBreuAGruberMReipertJHansenEProcessing of small volumes in blood salvage devicesTransfusion20145410pt22775278124953124
  • SasakiJTirottaCLimHComparison of stored red blood cell washing techniques for priming extracorporeal circuitsPerfusion201833213013528925857
  • SeyfriedTFHaasLGruberMBreuALoiblMHansenEFat removal during cell salvage: a comparison of four different cell salvage devicesTransfusion20155571637164325702832
  • GuYJVermeijdenWJde VriesAJHagenaarsJAGraaffRvan OeverenWInfluence of mechanical cell salvage on red blood cell aggregation, deformability, and 2,3-diphosphoglycerate in patients undergoing cardiac surgery with cardiopulmonary bypassAnn Thorac Surg20088651570157519049752
  • CheJTianMDingGEffects of cell salvage on erythrocyte 2,3-disphosphoglycerate and G-6-PD levels and phosphatidylserine expressionInt J Lab Hematol201335438539223176294
  • de VroegeRWildevuurWRMuradinJAGravesDvan OeverenWWashing of stored red blood cells by an autotransfusion device before transfusionVox Sang200792213013517298575
  • Westphal-VargheseBErrenMWestphalMProcessing of stored packed red blood cells using autotransfusion devices decreases potassium and microaggregates: a prospective, randomized, single-blinded in vitro studyTransfus Med2007172899517430464
  • ScholzMSerrickCNoelDSinghOMeloAA prospective comparison of the platelet sequestration ability of three autotransfusion devicesJ Extra Corpor Technol200537328616350382
  • SeyfriedTFStreithoffFGruberMPlatelet sequestration with a new-generation autotransfusion deviceTransfusion201858498999729380387
  • ArnaudFKapnikEMerymanHTUse of hollow fiber membrane filtration for the removal of DMSO from platelet concentratesPlatelets200314313113712850836
  • BruneTFillSHeimGRabsilberAWohlfarthKGarritsenHSPQuality and stability of red cells derived from gravity-separated placental blood with a hollow-fiber systemTransfusion200747122271227517714412
  • HacksteinHMöllerAGerlachMSachsUBeinGProspective quality control study of a novel gravity-driven whole blood separation system suitable for humanitarian crisesVox Sang2017112880680928952151
  • JohnsonLKwokMMarksDCPreparation of red blood cell concentrates and plasma units from whole blood held overnight using a hollow-fibre separation systemTransfusion Medicine2015251131925808374
  • BruneTHannemann-PohlKNißleKEckerNGarritsenHQuality, stability, and safety data of packed red cells and plasma processed by gravity separation using a new fully integrated hollow-fibre filter deviceAdvances in Hematology20092009316
  • HornseyVSMccollKDrummondOProwseCVSeparation of whole blood into plasma and red cells by using a hollow-fibre filtration systemVox Sanguinis2005892818516101688
  • YamadaCHeitmillerESNessPMKingKEReduction in potassium concentration of stored red blood cell units using a resin filterTransfusion20105091926193320561298
  • MorrisonAMcmillanLCampbellJDMPetrikJEvaluation of a potassium removal filter on irradiated red cells stored in SAGMTransfusion Medicine201525532032526174724
  • FujitaHShiotaniYTakadaYNishimuraSEffects of potassium adsorption filters on the removal of ammonia from blood productsBlood Transfus201816217328287384
  • MatsuuraHAkatsukaYMuramatsuCEvaluation of the potassium adsorption capacity of a potassium adsorption filter during rapid blood transfusionVox Sang2015108442843125754102
  • CidJVillegasVCarbasséGAlbaCPereaDLozanoMTransfusion of irradiated red blood cell units with a potassium adsorption filter: A randomized controlled trialTransfusion20165651046105126923301
  • GunaydinSGourlayTNovel ultrafiltration technique for blood conservation in cardiac operationsThe Annals of Thoracic Surgery20139562148215123706433
  • HoganMNeedhamAOrtmannEHaemoconcentration of residual cardiopulmonary bypass blood using Hemosep®: a randomised controlled trialAnaesthesia201570556357025692545
  • BoyleGKuffelAParmarKA comparison of haemostatic biomarkers during low-risk patients undergoing cardiopulmonary bypass using either conventional centrifugal cell salvage or the HemoSep devicePerfusion201895026765911878905
  • RobertsonCAGourlayTDevelopment of a diagnostic sensor for measuring blood cell concentrations during haemoconcentrationPerfusion201732212613227591743
  • GourlayTSimpsonCRobertsonCADevelopment of a portable blood salvage and autotransfusion technology to enhance survivability of personnel requiring major medical interventions in austere or military environmentsJ R Army Med Corps201816429610229079661
  • GunaydinSRobertsonCBudakABGourlayTComparative evaluation of blood salvage techniques in patients undergoing cardiac surgery with cardiopulmonary bypassPerfusion201833210510928856967
  • WegenerCOlsonBMinKMethod for sized-based cell separation using spinning membrane filtration: Google Patents2017
  • GiancolaROliosoPdi RitiMEvaluation of an automated closed fluid management device for processing expanded cytokine-induced killer cells to use in immunotherapy programs for cancerTransfusion200848462963918208417
  • ShuZHeimfeldSGaoDHematopoietic SCT with cryopreserved grafts: adverse reactions after transplantation and cryoprotectant removal before infusionBone Marrow Transplantation201449446947624076548
  • ZinnoFLandiFAureliVPositive immunomagnetic CD34+ cell selection in haplo-identical transplants in β-thalassemia patients: removal of platelets using an automated systemCytotherapy2010121606619929454
  • CalmelsBHouzéPHengesseJCDucrotTMalenfantCChabannonCPreclinical evaluation of an automated closed fluid management device: Cytomate, for washing out DMSO from hematopoietic stem cell grafts after thawingBone Marrow Transplant200331982382812732892
  • MfarrejBBouchetGCouquiaudJPre-clinical assessment of the Lovo device for dimethyl sulfoxide removal and cell concentration in thawed hematopoietic progenitor cell graftsCytotherapy201719121501150829037941
  • SchlinkerACRadwanskiKWegenerCMinKMillerWMSeparation of in-vitro-derived megakaryocytes and platelets using spinning-membrane filtrationBiotechnology and Bioengineering2015112478880025312394
  • SchlinkerACAutomated, spinning membrane filtration for preparation of mobilized leukapheresis products for CD34+ cell selectionCell and Gene Therapy Insights201738623637
  • YangXForouzanOBrownTPShevkoplyasSSIntegrated separation of blood plasma from whole blood for microfluidic paper-based analytical devicesLab Chip201212227428022094609
  • GannonTMcconnellKRileyJThe use of hydroxyethyl starch sedimentation for autologous buffy coat preparationJ Extra Corpor Technol200537331116350387
  • MuñozMGarcía-SegoviaSArizaDCobosAGarcía-ErceJAThomasDSedimentation method for preparation of postoperatively salvaged unwashed shed blood in orthopaedic surgeryBr J Anaesth2010105445746520639211
  • KhanalGHuynhRATorabianKXiaHVörösEShevkoplyasSSTowards bedside washing of stored red blood cells: a prototype of a simple apparatus based on microscale sedimentation in normal gravityVox Sang20181131313929067695
  • VörösEPietyNZStrachanBCLuMShevkoplyasSSCentrifugation-free washing: A novel approach for removing immunoglobulin A from stored red blood cellsAm J Hematol201893451852629285804
  • VörösELuMLezzarDShevkoplyasSSCentrifugation-free washing reduces buildup of potassium and free hemoglobin in washed red blood cells after the procedureAm J Hematol20189312E389E39130184257