Comparison of rechargeable versus battery-operated insulin pumps: temperature fluctuations

, , , &
Pages 371-376 | Published online: 14 Oct 2016

Abstract

The role of continuous subcutaneous insulin infusion (insulin pumps) has become increasingly important in diabetes management, and many different types of these systems are currently available. This exploratory study focused on the reported heating issues that lithium-ion battery-powered pumps may have during charging compared with battery-operated pumps. It was found that pump temperature increased by 6.4°C during a long charging cycle of a lithiumion battery-operated pump under ambient temperatures. In an environmental-chamber kept at 35°C, the pump temperature increased by 4.4°C, which indicates that the pump temperature was above that of the recommended safety limit for insulin storage of 37°C. When designing new pumps, and when using currently available rechargeable pumps in warmer climates, the implications of these temperature increases should be taken into consideration. Future studies should also further examine insulin quality after charging.

View responses to this article:
Reported insulin pump temperature fluctuations lack clinical relevance

Introduction

Continuous subcutaneous insulin infusion (insulin pump) therapy is an effective method of managing glycemic control in people with diabetes mellitus who require insulin. As such, accurate and reliable insulin delivery is a top priority when using an insulin pump.Citation1 As the use of insulin pumps has increased, many different rechargeable and battery-operated systems have been created.

Rechargeable lithium-ion batteries are becoming popular power sources for portable electronic devices, including insulin pumps.Citation2 The electrolytes contained within the battery have excellent conductivity and life-cycle performance. However, there are risks associated with the use of lithium-ion batteries, such as leakage and fire hazards. Furthermore, excessive heat exposure may cause the separator (a membrane inside the battery) to shrink or melt, causing a short circuit between the two electrodes.Citation2,Citation3 Prolonged exposure to high temperatures, such as those that may occur if a rechargeable battery is repeatedly overcharged or if the pump is exposed to direct sunlight, leads to increased heat production within the battery.

Insulin manufacturers specify that insulin temperatures should not exceed 37°C;Citation4,Citation5 temperatures above this cutoff carry the risk of insulin degradation, with the associated consequences of prolonged hyperglycemia.Citation6,Citation7 Previous studies have shown that exposure to temperatures higher than 37°C may increase the risk of insulin fibrillation, a form of protein misfolding that leads to the development of precipitates in insulin solutions.Citation8 Therefore, temperature fluctuations in rechargeable insulin pumps may have significant implications for the stability of the insulin in the pump. However, temperature fluctuations in nonrechargeable battery-operated pumps have not yet been examined compared with those using rechargeable batteries.

Changes in insulin temperature may potentially affect insulin action, and thus have the potential to compromise glycemic control and/or safety. Therefore, the goal of this study was to evaluate temperature fluctuations during the charging period of rechargeable insulin pumps, and compare them with temperature fluctuations in battery-operated insulin pumps, in both ambient temperatures (eg, usual home/office temperature in the US) and in a high-temperature environment (eg, outdoors on a summer day in the US).

Materials and methods

Materials and equipment

The materials and equipment required for this study included an extruded polystyrene rigid-foam insulation board (61×45.7×2.5 cm, Foamular®; Owens Corning, Toledo, OH, USA) to prevent stray heat conduction between the experimental pumps. The design also included a TiS40 infrared (IR) camera and two 54 II B thermometers for measuring air and insulin-pump temperatures (Fluke Corporation, Everett, WA, USA). The 54 II B thermometers, together with Chromega–Alomega thermocouples (SA1-K-72-SC, gauge 30 [0.025 cm], length 182.9 cm; Omega Engineering Inc, Stamford, CT, USA) were used to monitor insulin-pump and ambient temperatures. A temperature-controlled environment was simulated in a Hotpack environmental-chamber (417522; SP Industries, Warminster, PA, USA). All devices were calibrated and certified at time of use.

Six t:slim pumps (Tandem Diabetes Care Inc, San Diego, CA, USA) were used as insulin pumps with rechargeable batteries. Six Vibe pumps (Animas Diabetes Care LLC, West Chester, PA, USA) were used as insulin pumps with regular batteries. High-purity water with food dye (McCormick, Sparks, MD, USA) was used in lieu of insulin to fill cartridges and prime pumps.

Experimental design: pump testing

The study set up was constructed and conducted on the extruded polystyrene rigid-foam insulation board in ambient temperature () and in a temperature-controlled environmental-chamber (). Two thermometers (54 II B) were used to measure pump and air temperatures. One thermometer was used to measure pump temperatures and required two thermocouples: one attached to the rear of the rechargeable pump located in the center of the set up (), and one attached to the rear of the battery-operated pump. Both thermocouples were attached at the location of the insulin cartridge. A separate thermometer, with one thermocouple, provided reference air-temperature measurements in both settings. For both ambient-air and environmental-chamber temperature measurements, the IR camera was set up on a tripod, targeting the rechargeable pump located in the center of the set up, and capturing all 12 pumps in the IR camera frame.

Figure 1 Ambient (A) and environmental-chamber (B) temperature-measurement set up and sensor placement (C).

Note: Images show infrared camera on tripod, foam board containing six battery-operated pumps, six rechargeable pumps, and two sensors.
Figure 1 Ambient (A) and environmental-chamber (B) temperature-measurement set up and sensor placement (C).

During the ambient-temperature study, IR pictures were taken every 60 seconds, and insulin-pump temperature and air temperature were logged every 60 seconds over the course of a 3-hour period. Measurements were started just prior to the power strip being plugged in. IR imaging was used to register and demonstrate temperature changes of all 12 pumps, but thermocouple measurements were conducted on only one sample of each type, to quantify the extent of heating. During the elevated-temperature study, the environmental-chamber was heated to 35°C, and the equipment was set up as depicted in . The chamber’s design allows the pulling of a power cable through to the outside, which can then be plugged in without opening the chamber. The environmental-chamber door was securely closed, and 2 hours were allowed before the start of the study to let the temperature return to 35°C. Once the proper temperature was reached, the power strip was plugged in, initiating the charging period for the rechargeable pumps. IR pictures were taken every 60 seconds, and insulin pump temperature and air temperature were logged every 60 seconds over the course of a 3-hour period.

For both experiments, rechargeable pumps were either low on battery or drained when charging initiated, while battery-operated pumps were equipped with new batteries. Pumps stayed plugged in for 3 hours. The rechargeable pumps were fully charged after approximately 70 minutes, as judged at the end of the temperature climb.

Results

For the rechargeable pumps, the sensor readouts indicated temperature increases in both the ambient- and elevated-temperature settings. An increase of 6.4°C from the initial temperature was measured in the ambient-temperature setting (). The sensor readouts also indicated an increase of 4.4°C from the initial temperature in the environmental-chamber setting, which was kept at a constant 35°C (), with a maximum measured temperature of 40.4°C. The IR camera also revealed an apparent increase from the initial temperature in the environmental-chamber setting (). The IR camera showed a 2.7°C-higher starting temperature compared with the actual chamber temperature. No significant temperature differences were found between the battery-operated pumps in either the environmental-chamber (35.9°C) or the ambient setting (22.6°C).

Figure 2 Ambient-temperature results.

Notes: Battery and rechargeable battery pump temperatures, insulin degradation temperature, and ambient air temperatures shown, measured over 180 minutes.
Figure 2 Ambient-temperature results.

Figure 3 Enviromental-chamber temperature results.

Notes: Battery and rechargeable battery pump temperatures, insulin degradation temperature, and chamber temperatures are shown, measured over 180 minutes in a controlled (35°C) environment.
Figure 3 Enviromental-chamber temperature results.

Figure 4 Infrared image of temperature increases in rechargeable pumps, with the layout of the devices in the top-left corner.

Note: Measurements were performed in the environmental-chamber which was maintained at a constant temperature (35°C, 95°F).
Figure 4 Infrared image of temperature increases in rechargeable pumps, with the layout of the devices in the top-left corner.

Discussion

In this study, temperatures measured during a full charging period of the rechargeable lithium-ion battery pump (70 minutes) were shown to increase substantially over time in both the ambient-temperature setting and in the temperature-controlled environmental-chamber. The results further showed that the temperatures of the rechargeable insulin pumps often exceeded 37°C in an environment with an elevated temperature of 35°C during a full charging period. In contrast, the battery-operated pumps showed little or no temperature fluctuation in either environment. Manufacturers of rechargeable insulin pumps suggest charging the battery for no longer than 10–15 minutes per day, possibly to avoid overcharging and the associated consequences described here. However, manufacturers do not actually prohibit charging longer,Citation9 and thus a full-charge cycle as tested here is not an unlikely scenario.

Insulin-analog manufacturer specifications indicate that their product should not be exposed to temperatures over 37°C.Citation4,Citation5 The labeling information containing instructions for diabetes patients using insulin pumps states that the temperature of the insulin in the pump cartridge may exceed ambient temperature when the pump housing, cover, tubing, or sport case is exposed to sunlight or radiant heat.Citation4 During exploratory experiments in which temperature sensors were placed inside the pump cartridge, additional heating beyond what was observed in the current study was noted (data not shown; due to the invasive nature of these exploratory experiments, which required manipulation of the insulin pumps, it was decided that these experiments were not representative of appropriate pump use). In the study described here, it was shown that during charging, insulin pumps with rechargeable lithium-ion batteries were subjected to temperature increases that may exceed the recommended temperatures for insulin storage and handling, especially in warm environments. These temperature variations may result in unexplained glycemic variability in the patient, as higher temperatures may result in compromised insulin quality.

Conclusion

By measurements, we have shown that lithium-ion rechargeable batteries, such as those currently used to power insulin pumps, showed substantial temperature increases during a full charging period in both ambient conditions and warm environments compared with battery-operated pumps. When charging in warm environments, rechargeable pump temperatures may exceed the temperature recommended for insulin storage and handling, potentially affecting the reliability of insulin action. When using lithium-ion rechargeable batteries in the design of new pumps, the implications of these temperature increases should be taken into consideration, as well as the effect heating may have on insulin quality, especially in warmer climates, as mentioned earlier, as the effects were largest in the environmental-chamber. The effect on insulin will have to be further examined in future studies.

Acknowledgments

This study was sponsored by Animas Corporation. The authors received editorial/writing support from Excerpta Medica in the preparation of this manuscript.

Disclosure

This study was sponsored by Animas Corporation. The authors report no other conflicts of interest in this work.