MRI-conditional pacemakers: current perspectives

, , , , &
Pages 115-124 | Published online: 07 May 2014

Abstract

Use of both magnetic resonance imaging (MRI) and pacing devices has undergone remarkable growth in recent years, and it is estimated that the majority of patients with pacemakers will need an MRI during their lifetime. These investigations will generally be denied due to the potentially dangerous interactions between cardiac devices and the magnetic fields and radio frequency energy used in MRI. Despite the increasing reports of uneventful scanning in selected patients with conventional pacemakers under close surveillance, MRI is still contraindicated in those circumstances and cannot be considered a routine procedure. These limitations prompted a series of modifications in generator and lead engineering, designed to minimize interactions that could compromise device function and patient safety. The resulting MRI-conditional pacemakers were first introduced in 2008 and the clinical experience gathered so far supports their safety in the MRI environment if certain conditions are fulfilled. With this technology, new questions and controversies arise regarding patient selection, clinical impact, and cost-effectiveness. In this review, we discuss the potential risks of MRI in patients with electronic cardiac devices and present updated information regarding the features of MRI-conditional pacemakers and the clinical experience with currently available models. Finally, we provide some guidance on how to scan patients who have these devices and discuss future directions in the field.

Introduction

Magnetic resonance imaging (MRI) is a versatile imaging technique capable of producing high quality images of the human body. MRI provides excellent spatial resolution and unparalleled tissue characterization without exposing patients to the potential risks of ionizing radiation and iodinated contrast agents. Beyond morphological assessment, MRI is capable of providing valuable information on tissue perfusion, function, and metabolism. For these reasons, MRI has become the imaging modality of choice in a wide spectrum of clinical situations, and is now used for the diagnosis, staging, and follow-up of numerous diseases, including a large number of neurological, musculoskeletal, oncological, and cardiovascular disorders. As a consequence, the number of MRI scans has grown strongly in recent decades. In the USA, the number of procedures rose from 7.7 million in 1993 to nearly 22 million in 2002.Citation1 More recent data indicate that, in 2011, 32 million scans were performed in that country.Citation2 Worldwide, it is estimated that approximately 60 million scans are performed each year.Citation3 The use of MRI is likely to continue to grow due to a combination of an aging population and the expanding indications and increasing accessibility to this technique.

Parallel to the growth of MRI, the number of patients with implantable cardiac devices, including pacemakers, implantable cardioverter defibrillators, and cardiac resynchronization therapy, has been steadily increasing. Between 1993 and 2009, 2.9 million patients received permanent pacemakers in the USA and overall use increased by more than 50%.Citation4 A worldwide survey showed that, in 2009 alone, more than one million pacemakers were implanted, and virtually all countries reported increases in implant numbers.Citation5 The combination of these two growing phenomena results in an estimated 50%–75% probability of a patient needing an MRI over the lifetime of the device.Citation6 These patients will generally see their MRI studies denied due to safety concerns.Citation7Citation9 This “clash of technologies” thus creates a paradoxical effect in which application of a useful technique is limited by the presence of another technology in the same patient.Citation10 The recent introduction of MRI-conditional pacemakers overcomes this important limitation of MRI and cardiac pacing, allowing physicians to take full advantage of this imaging technique in the growing number of patients with cardiac devices. In this review, we discuss the potential hazards of the interactions between conventional pacemakers and the MRI environment, and present updated information regarding the features of MRI-conditional pacemakers and clinical experience with currently available models. Finally, we provide some advice on how to scan patients with these devices and discuss future directions in the field.

Potential risks of MRI in patients with conventional pacemakers

The risks associated with MRI can be broadly categorized by the interaction between the pacemaker system and three essential components of MRI, ie, the static magnetic field, gradient magnetic fields, and radiofrequency energy.Citation11

Static magnetic field

Most modern clinical MRI scanners operate at static magnetic fields of 1.5 or 3 tesla (T), roughly corresponding to 30,000 and 60,000 times the strength of the earth’s magnetic field, respectively. The greatest risk from this strong magnetic field is the attraction of ferromagnetic objects into the scanner causing movement, torque, dislodgment, or even a “projectile effect” that can result in patient injury and damage to the MRI system. However, except for older devices with a higher content of ferromagnetic material (approved by the US Food and Drug Administration [FDA] prior to 1998), the mechanical forces exerted on pacemaker systems are usually negligible at 1.5 T.Citation12 Of greater concern is the possibility of magnetic sensor activation and unpredictable reed-switch behavior, causing the device to revert to asynchronous pacing.Citation13,Citation14 The static magnetic field is also responsible for the magnetohydrodynamic effect. Since blood contains electrically charged ions, its flow in the presence of a powerful static magnetic field produces small voltages, which are superimposed on the patient’s electrocardiogram. This may simulate life-threatening arrhythmias and produce other electrocardiographic changes, including T-wave abnormalities and elevation of the ST segment.Citation15 The magnetohydrodynamic effect hinders electrocardiographic monitoring during scanning,Citation16 and may lead to inappropriate inhibition of pacemaker function.Citation17 The stronger the magnetic field, the greater the magnitude of this phenomenon.

Gradient magnetic fields

During MRI scanning, gradient coils create additional, linearly varying magnetic fields that add or subtract from the main magnetic field. Gradient magnetic fields are essential for spatial encoding and are measured in millitesla per meter (mT/m). High performance gradient coils used in MRI are typically capable of producing gradient magnetic fields of 20–100 mT/m or higher for a 1.5 T scanner. These gradients are repeatedly and rapidly turned on and off (the slew rate of a gradient system is a measure of how quickly this can happen). These rapidly changing magnetic fields can induce electrical currents in pacemaker leads, causing oversensing, undersensing, or even life-threatening arrhythmias.Citation8,Citation13,Citation18,Citation19

Radiofrequency energy

In order to produce images, MRI scanners use short bursts of electromagnetic waves at very specific frequencies (radiofrequency pulses), which interact with spinning protons. At the end of each pulse, protons return to their previous spinning orientation and, in doing so, release energy in the form of radiofrequency waves that are used for reconstructing an image. During this process, the body will absorb some of the radiofrequency energy causing resistive heating. The specific absorption rate (measured in W/kg) is the dosimetric term used to describe the amount of radiofrequency energy employed during an MRI scan. Pacemaker leads can act as “antennae” which concentrate this radiofrequency energy, producing heat and electrical currents,Citation20 which may cause tissue destruction at the lead tip,Citation21 myocardial stimulation (including life-threatening arrhythmias), and damage to the pulse generator circuitry and battery.Citation22Citation24 This may produce pacemaker reset, battery depletion, and adverse effects on sensing, pacing thresholds, and lead impedances, causing inappropriate pacing acceleration or inhibition.Citation10,Citation25 Abandoned or fractured leads are especially prone to tip heating.Citation26,Citation27

MRI in patients with conventional pacemakers: feasible?

Given the potential risks of the interactions between the MRI environment and conventional pacemakers, these devices are classified as MRI-unsafe and their presence has long been considered a contraindication for MRI. Over the years, these safety concerns have been substantiated by at least 17 supposed MRI-associated deaths among patients with pacemakers.Citation28 This number is probably an underestimation of the real number of fatalities, since there are several cases of patients with a cardiac pacemaker who died after exposure to MRI that have never been reported in the medical literature but have come to light via the general press or the legal system.Citation29 However, it should also be noted that a firm causal relationship has seldom been established, and that most of these deaths occurred in patients with older pacemaker models undergoing MRI without appropriate programming or physician-supervised monitoring.Citation28,Citation30

Despite all safety concerns, many pacemaker patients without an acceptable alternative imaging modality have undergone MRI uneventfully.Citation31Citation45 A recent review of 15 human studies involving 1,419 MRI scans (mostly nonthoracic) reported no serious adverse events, although 65% of these were performed in patients with MRI-conditional devices.Citation30 Without accounting for these, 49% of the examinations did not result in any significant changes in pacemaker function after the scan.Citation30 A multicenter registry of clinically indicated nonthoracic MRI at 1.5 T for patients with non-MRI-conditional pacemakers and implantable cardioverter defibrillators implanted after 2001 is currently underway.Citation46 This registry is expected to encompass up to 1,500 examinations, provide further insights on the risks of MRI in such conditions, and hopefully validate a clinical protocol for screening and scanning these patients, who will for many years constitute the majority of those with implanted cardiac devices.

It should be emphasized that special precautions were taken in these studies, including careful patient selection and device programming, rigorous monitoring, reanimation readiness during the procedure, and exclusion of thoracic MRI (in most cases). So, performing MRI in patients with conventional pacemakers cannot be considered a routine procedure.Citation7,Citation8,Citation47Citation49 Moreover, the number of patients examined in these conditions is still largely insufficient to determine safety, and possible long-term effects are unknown.Citation50 Rather than establishing the clinical safety of MRI in patients with non-MRI-conditional devices, these data have shown that the risk may be lower than previously thought if a number of conditions are met and special precautions are taken.Citation51 A practical implication of these findings is that, when the information provided by MRI seems vital for the management of a patient with a non-MRI-conditional pacemaker, the risk of scanning under close surveillance is probably lower than the risk of removing existing leads, since lead extraction is associated with a significant risk of complications.Citation52,Citation53

Position statements issued by the American Heart Association and the European Society of Cardiology continue to advise against MRI in patients with conventional pacemakers, although they do not establish it as an absolute contraindication.Citation7,Citation8 The American Heart Association guidelines discourage MRI in non-pacemaker-dependent patients, except in cases with a strong clinical indication and in which the benefits clearly outweigh the risks.Citation7 Among pacemaker-dependent patients, MRI should not be performed unless there are highly compelling circumstances.Citation7 The European guidelines also strongly recommend against MRI in non-pacemaker-dependent patients, except in the presence of a documented, extremely serious, life-threatening, or severely “quality of life-limiting” condition.Citation8 In pacemaker-dependent patients, the indication for MRI should be seriously reconsidered if the underlying rhythm is too slow, and the threshold for imaging and safety requirements are higher.Citation8 Both sets of guidelines restrict MRI in patients with conventional pacemakers to specialized centers with specific expertise and equipment for close monitoring. In 2013, the American College of Radiology issued a Guidance Document on Safe MR Practices,Citation54 where a case-by-case and site-by-site appraisal is recommended before scanning patients with non-MRI-conditional devices. These guidelines also emphasize the need for careful risk-benefit assessment and close collaboration between radiology and cardiology staff and industry representatives of the device manufacturers.

MRI-conditional pacemakers: “safe by design”

The increasing importance of MRI as a diagnostic tool and the limitations imposed by conventional pacemakers prompted a large amount of research and efforts to develop devices suitable for use in this environment. Gathering the knowledge collected in in vitro, animal, and human studies, pacemaker manufacturers introduced a significant number of modifications to their devices.Citation55 The most important are specified below and summarized in .

Table 1 Main modifications introduced in MRI-conditional pacing systems

Minimization of ferromagnetic content

The reduction of ferromagnetic content is intended to decrease magnetic attraction and also minimize susceptibility artifacts produced by the pacemaker system (). This can be achieved through the use of nonferromagnetic materials, which have to be appropriately conductive, durable, and biocompatible. The leads are made of nonmagnetic materials.Citation10

Figure 1 Balanced steady-state free precession (b-SSFP) cine images of the heart in a patient with an implanted magnetic resonance imaging-conditional pacemaker.

Notes: Note the susceptibility artifacts from the pulse generator in left ventricular 2-chamber view (A) and from the right ventricular lead in short-axis view (B). Despite being clearly apparent, these artifacts don’t usually hinder diagnostic interpretation, except when the region of interest is in the proximity of the pacemaker generator.
Figure 1 Balanced steady-state free precession (b-SSFP) cine images of the heart in a patient with an implanted magnetic resonance imaging-conditional pacemaker.

Replacement of reed switches

Conventional pacemakers have reed switches devised to revert the device to asynchronous pacing mode (DOO or VOO) when a magnet is placed over the impulse generator. This feature is useful to avoid the effects of electromagnetic interactions such as the ones that occur when electrocautery is used during surgery. Reed switch behavior in the MRI environment may vary with the strength of the magnetic field and the orientation of the reed switch, and is essentially unpredictable.Citation14 To overcome this problem, reed switches have been replaced by solid-state Hall sensors, which have more predictable behavior when exposed to magnetic fields.Citation3,Citation10

Lead design

The goal is to minimize the amount of heat and electrical currents that are induced on the leads by the electromagnetic energy used in MRI. To achieve this, the resonant frequency should be avoided to prevent the lead from acting as a receiver of electromagnetic impulses. This was accomplished by modifications in lead geometry that included changing the winding pattern of the filaments that compose the inner lead coil, thus limiting the radiofrequencies that can conduct through the filaments.Citation10,Citation56,Citation57 Lead tip coating and bipolar lead designs were also used to minimize electromagnetic interference.Citation10 It should be emphasized that MRI-conditional pacemakers are designed to be used only with compatible MRI-conditional leads in accordance with the manufacturer’s conditions. So, except for some pacing leads that were proven to be “backwards MRI-compatible”, connecting an MRI-conditional generator to pre-existing conventional leads is insufficient to make the system MRI-conditional, even though this may theoretically minimize the chances of power-on reset during an MRI scan.

Special pacemaker circuitry

MRI-conditional pacemakers have been equipped with special filters that limit the transfer of certain frequencies and dissipate energy, reducing the risk of damage to the internal power supply and circuitry. Generator shielding has also been improved to minimize the transfer of electromagnetic energy.Citation10

Dedicated pacemaker programming

Dedicated programming is an essential feature of MRI-conditional pacemakers. An “MRI mode” must be switched on before entering the scanner, and switched off immediately afterwards. Specific programming pathways were developed to assist in the choice between asynchronous versus nonstimulation modes (a decision that still requires arrhythmia expertise). Programming also includes mandatory system-integrity checks, increased pacing output during MRI scanning, and effortless restoration of prescan program states and values.Citation51

Growing portfolio of MRI-conditional pacemakers and clinical experience so far

Medtronic® MRI-conditional pacemakers

Medtronic (Minneapolis, MN, USA) was the first company to introduce an MRI-conditional pacemaker in 2008 (named EnRhythm™ in Europe, Revo MRI™ in the USA). A second generation of MRI-conditional pacemakers (Ensura MRI™ and Advisa MRI™) was introduced in 2011, overcoming some of the limitations of the first models. These limitations include an upper rate limit of 150 beats per minute and the inability to be used in unipolar mode (sometimes useful to deal with sensing or threshold problems). All these dual chamber devices are approved for whole body scans and include comprehensive arrhythmia detection and software designed to minimize ventricular pacing. Medtronic provides MRI-conditional leads with active and passive fixation. The active fixation lead 5086 is based on the older 5076 CapSureFix Novus™ with a two-filar inner coil designed to obtain higher inductance and reduce lead tip heating. Recently, the 5076 lead received CE (Conformité Européenne) marking for backwards MRI compatibility, meaning that patients with older pacemaker systems using this lead can be upgraded to MRI-conditional systems with a simple box change. The passive fixation 5.3F isodiametric lead (CapSure Sense™) also received CE marking for backwards MRI compatibility.

The Medtronic EnRhythm SureScan™ pacemaker was first tested outside the MRI environment, where it was compared in terms of efficacy and safety with a conventional dual-chamber pacemaker system with active fixation leads. In the first of two single-center studies, procedural and fluoroscopy times were longer for the MRI-conditional device, but implantation success rate and clinical performance did not differ in the 107 patients evaluated.Citation58 A subsequent study did not find significant differences in reoperation rates or short-term clinical performance between conventional pacemakers and the new pacemaker system with the newly designed leads (5086 CapSureFix).Citation59

The first prospective, randomized multicenter study assessing the efficacy and safety of an MRI-conditional pacemaker in the MRI environment was published in 2011. Wilkoff et al assessed 464 patients implanted with the Medtronic Revo SureScan pacemaker randomized to undergo (n=258) or not undergo (n=206) a nonmedically indicated brain and lumbar MRI in the 9–12 weeks post-implantation.Citation60 MRI scan restrictions were similar to those used in protocols for conventional pacemakers, ie, exclusion of thoracic scanning, static magnetic field strength limited to 1.5 T, maximum specific absorption rate of 2 W/kg, and maximum gradient slew rate of 200 mT/m per second. There were no complications (primary safety endpoint) and no MRI-attributed pacemaker sensing or threshold changes (primary efficacy endpoint) in patients who underwent scans. These included subjects with and without pacemaker dependency, using the asynchronous mode (n=158) and no pacing with a continuous intrinsic rhythm during scanning (n=67).

More recently, the second-generation Medtronic Advisa MRI™ SureScan Pacemaker and CapSureFix MRI SureScan lead were studied in a clinical trial where 263 patients were randomized in a 2:1 ratio to undergo 16 chest and head scans at 1.5 T between 9 and 12 weeks after pacemaker implantation or not to undergo MRI.Citation61 There were no MRI-related complications during or after MRI in the scanned subjects. Differences in pacing capture threshold values from pre-MRI up to one month post-MRI were minimal and similar between the MRI and control groups. This device has subsequently received CE marking in Europe and was also approved by the FDA without positioning restrictions for MRI scans or limitations of body parts scanned, overcoming previous limitations on chest imaging.

St Jude Medical® MRI-conditional pacemakers

The Accent MRI™ conditional pacemaker, available in single and dual chamber versions, was introduced in the European market in 2011 and is currently awaiting FDA approval. In Europe, this device is approved for full body scans (no zone restrictions) at 1.5 T when used in combination with Tendril™ MRI leads. St Jude (St Paul, MN, USA) also introduced a hand-held MRI Activator™ that allows quick enabling and disabling of preapproved MRI settings to facilitate the MRI scanning workflow (these can be preprogrammed by the physician at any follow-up visit). The Accent MRI-conditional pacemaker features full bradyarrhythmia therapy and is also compatible with the St Jude Merlin™ home monitoring system. The Tendril™ MRI lead has a coaxial design and 6.6 F body (requiring an 8F introducer). It is based on standard Tendril™ lead inner and outer coils with additional silicone inner tubing and Optim™ insulation, including a second filter to prevent tissue heating and unintended cardiac stimulation. A soft silicon tip was also introduced to reduce the chances of cardiac perforation, addressing previous concerns regarding increased stiffness of MRI-conditional leads compared with conventional leads.Citation62

Boston Scientific® MRI-conditional pacemakers

The Ingenio™ and Advantio™ Boston Scientific (Natick, MA, USA) MRI pacemakers are available in single and dual chamber with CE approval. The devices are based on the Ingenio conventional pacemaker and are compatible with the Latitude™ remote monitoring system. They include a programmable MRI timer designed to return pacemaker settings back to normal after the scan. FINELINE™ II 5.1F active fixation leads received backwards MRI-conditional approval. A multicenter, nonrandomized single-arm study (INFINITE MRI)Citation63 to collect data on the Boston Scientific MRI-conditional pacing system (consisting of an Ingenio MRI or Advantio MRI pacemaker with FINELINE™ II Sterox or FINELINE™ II Sterox EZ leads) is presently underway.

Biotronik® MRI-conditional pacemakers

The Evia™ and Estella™ MRI-conditional pacemakers, available in single and dual chamber models, are approved in Europe for MRI scanning at 1.5 T with isocenter positioning above the eye level and below the level of the greater trochanters (in practice, this limitation usually precludes thoracic and upper abdominal imaging, but not head or lower extremity scans). Biotronik (Berlin, Germany) recommends limiting scan duration to 30 minutes and total device lifetime scan time to 10 hours. Both devices are compatible with a remote monitoring system.

A prospective feasibility pilot study assessing 30 patients with the Biotronik Evia SR-T and DR-T pacemaker with Safio™ S53/S60 screw-in leads showed encouraging results.Citation64 All patients underwent MRI scans of the head and lumbar spine and were evaluated before, immediately after, and at one and 3 months post-MRI. There were no MRI-related adverse events, and no significant differences in lead parameters were identified. These findings await confirmation from a multicenter, randomized clinical trial designed to evaluate the one-month rate of adverse events with the Biotronik ProMRI™ Entovis™ MRI-conditional pacing system, which has finished the recruitment phase (n=245) and is expected to reach completion in early 2014. A trial extension aiming to evaluate the safety of this device without exclusion zones has recently received FDA approval.Citation65

Sorin Group (Milano, Italy) MRI-conditional pacemaker

The KORA 100 MRI-conditional pacemaker is available in Europe in single and dual chamber models. This device, built on the REPLY™ pacemaker platform, includes a filter between the lead and device electronics and received CE mark approval for MRI scanning at 1.5 T with chest exclusion. The KORA 100 was designed for implantation with MRI-conditional BEFLEX™ leads. Sorin (Milan, Italy) developed an “auto MRI mode” that automatically switches to asynchronous pacing in the presence of a strong magnetic field, thus limiting the time in asynchronous mode to the MRI scan duration.

Safe scanning of patients with MRI-conditional pacemakers

The term “MRI-conditional” is applied to devices that pose no known hazards in a specific MRI environment under specific device and MRI scanner conditions.Citation66 So, as the name implies, scanning patients with “MRI-conditional” pacemakers is safe only if a number of conditions are met. Such conditions are related to the MRI scanner, the MRI-conditional pacemaker (generator and leads), and patient characteristics. These are summarized in .

Table 2 Requirements for scanning patients with MRI-conditional pacemakers

In practice, a number of steps should be followed when a patient with an MRI-conditional pacemaker is to undergo an MRI scan.Citation10,Citation67,Citation68 Safe and effective scanning starts at the moment of scheduling, where the presence and close cooperation of several health care professionals during the scan (radiologist, MRI technologist, electrophysiology physician, and electrophysiology technician) should be guaranteed in advance.

Prior to the MRI examination, this team should exclude any contraindications, review the specific scanning conditions for the implanted pacemaker, and plan the procedure. A pre-MRI chest roentgenogram is useful to confirm electrode position and integrity, and check for MRI-conditional radiopaque markers if there are doubts regarding the implanted material (it should be noted that leads for which MRI-conditional status has been established a posteriori will not have such markers). Selecting the appropriate pacing mode starts with clinical assessment, reviewing the original indication for implantation, underlying rhythm and pacemaker dependency. Device interrogation prior to the examination is mandatory, except in patients carrying the Accent MRI conditional pacemaker with the optional hand-held Activator (these patients can undergo MRI after activation of preprogrammed device settings). However, even in those circumstances, device interrogation should ideally be performed to ensure that conditions have not changed significantly between preprogramming and the MRI scan. Device interrogation provides information on system integrity, percentage of pacing, programmed pacing rate, and AV interval, aiding in correct pacemaker programming prior to entering the MRI environment. In pacemaker-dependent patients, asynchronous mode should be selected. In patients with an underlying rhythm who are not bradycardic, asynchronous pacing competing with a ventricular rate could result in pacing during ventricular repolarization (“R on T phenomenon”) potentially causing life-threatening arrhythmias.Citation10 After adequate programming, the MRI mode is then turned on before the patient enters the room.

During the MRI scan, electrocardiography, pulse oximetry, and noninvasive blood pressure measurements (at least one of these) must be used for continuous hemodynamic monitoring. An external defibrillator (preferably capable of external pacing) should be readily available outside the MRI suite.

Once the scan is completed, the pacemaker’s MRI mode should be turned off and the device interrogated and reprogrammed, preferably using the parameters stored prior to MRI scanning. Ideally, an electrophysiologist should supervise these steps. If these precautions are taken, scanning patients with MRI-conditional devices should be safe and relatively simple.

Who should receive an MRI-conditional pacemaker?

There is currently no consensus regarding who should receive an MRI-conditional pacemaker. Opinions range from universal adoption to almost complete dismissal, as some critics consider that the increased cost is unjustified and that conventional pacemakers can be safely scanned with appropriate precautions.Citation13 From a clinical standpoint, the option for MRI-conditional pacemakers seems logical and beneficial in patients who have no other contraindications for MRI. However, adoption of this technology has been slower than expected and, in many centers, conventional pacemakers still account for the majority of implantations. The possible explanations for this slow uptake include the higher cost of MRI-conditional pacemakers, the lack of clear guidance on who should receive one of these devices, and the absence of long-term data on safety and durability. This later issue may be particularly important for the newer leads, since the introduced modifications may (at least theoretically) affect their performance characteristics, and their rates of long-term failure and fracture are essentially unknown. Other factors such as the limited range of choices in device manufacturer and lead fixation options may also contribute to their slow adoption (particularly in the USA), but these will probably wane with the recent and planned additions to the MRI-conditional hardware portfolio.

Even if the universal adoption of MRI-conditional pacemakers seems desirable, the current financial constraints argue for selective implantation in patients who are more likely to need MRI in the future. Although some reasonable predictors can be put forwardCitation10,Citation62,Citation68,Citation69 (), foreseeing the need for MRI on an individual basis is a challenging task.

Table 3 Compelling reasons for choosing an MRI-compatible pacemaker

Another option to consider is the implantation of MRI-conditional leads together with a less expensive conventional pacemaker generator (that can be replaced by an MRI-conditional box if MRI is required). However, this strategy does not allow for emergency scanning and its cost-effectiveness is uncertain.

Unanswered questions and future directions

Besides patient selection, other important questions have been raised by the arrival of MRI-conditional pacemakers. While their safety and effectiveness has been reasonably demonstrated, cost-effectiveness versus conventional pacemakers remains largely unknown.

This appraisal will ultimately depend on the impact of MRI-conditional technology on patient outcomes, which may be difficult to demonstrate, since the advantage of these devices lies in the ability to perform diagnostic imaging, which may influence patient outcome, but only in an indirect and difficult to measure way. Other economic issues such as reimbursement for peri-scanning assessment and monitoring will also have to be considered. Legal implications may surface, especially in cases where, despite the availability of MRI-conditional pacemakers, a conventional device is implanted and the patient subsequently needs to undergo MRI.Citation62 The impact on image quality will also have to be assessed, especially (but not only) in thoracic imaging.

Regarding hardware improvements, manufacturers will face the task of overcoming current device limitations such as restricted scan areas and specific absorption rate limits. Also, as 3.0 T imaging becomes more prevalent, it will also be important to have devices tested and approved for scanning at this magnetic field strength.

Meanwhile, the medical and scientific community should try to reach consensus and establish clear criteria for the implantation of these devices. Other important steps include consolidating the safety profile of this technology with evidence from large multicenter registries, and promoting the education of health care professionals on this subject (with an emphasis on multidisciplinary algorithms for safe and efficient scanning).

Conclusion

While the risk of performing MRI on patients with conventional pacemakers is probably lower than previously thought, patient safety can only be assured with the use of MRI-conditional devices engineered to minimize the interactions between the pacemaker system and the MRI environment. Several MRI-conditional pacemakers are now available on the market and more will be introduced in the future. Currently available devices differ from each other in several aspects, including conditions of use and robustness of clinical trial data. Although this leap in pacing technology has opened the doors for this versatile imaging modality to a growing number of patients, widespread adoption will likely depend on pragmatic issues such as cost, definition of clear criteria for implanting an MRI-conditional device, and continuous education of health care professionals.

Disclosure

The authors report no conflicts of interest in this work.

References

  • IMV Medical Information Division, Inc.IMV Benchmark Report MR2008
  • IMV Medical Information Division, Inc.IMV Benchmark Report MR 2012 Available from: http://www.imvinfo.com/index.asp?sec=mkt&sub=omkt&pag=def&pid=90631Accessed April 17, 2014
  • SuttonRKanalEWilkoffBLSafety of magnetic resonance imaging of patients with a new Medtronic EnRhythm MRI SureScan pacing system: clinical study designTrials200896819055703
  • GreensponAJPatelJDLauETrends in permanent pacemaker implantation in the United States from 1993 to 2009: increasing complexity of patients and proceduresJ Am Coll Cardiol2012601540154522999727
  • MondHGProclemerAThe 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009 – a World Society of Arrhythmia’s projectPacing Clin Electrophysiol2011341013102721707667
  • KalinRStantonMSCurrent clinical issues for MRI scanning of pacemaker and defibrillator patientsPacing Clin Electrophysiol20052832632815826268
  • LevineGNGomesASAraiAESafety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic ResonanceCirculation20071162878289118025533
  • RoguinASchwitterJVahlhausCMagnetic resonance imaging in individuals with cardiovascular implantable electronic devicesEuropace20081033634618308754
  • NazarianSBeinartRHalperinHRMagnetic resonance imaging and implantable devicesCirc Arrhythm Electrophysiol2013641942823592868
  • ShinbaneJSCollettiPMShellockFGMagnetic resonance imaging in patients with cardiac pacemakers: era of “MR Conditional” designsJ Cardiovasc Magn Reson2011136322032338
  • BeinartRNazarianSEffects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practiceCirculation20131282799280924366589
  • LuechingerRDuruFScheideggerMBBoesigerPCandinasRForce and torque effects of a 1.5-Tesla MRI scanner on cardiac pacemakers and ICDsPacing Clin Electrophysiol200124199120511270700
  • IrnichWIrnichBBartschCStertmannWAGuflerHWeilerGDo we need pacemakers resistant to magnetic resonance imaging?Europace2005735336515944094
  • LuechingerRDuruFZeijlemakerVAScheideggerMBBoesigerPCandinasRPacemaker reed switch behavior in 0.5, 1.5, and 3.0 Tesla magnetic resonance imaging units: are reed switches always closed in strong magnetic fields?Pacing Clin Electrophysiol2002251419142312418737
  • NijmGMSwirynSLarsonACSahakianAVCharacterization of the magnetohydrodynamic effect as a signal from the surface electrocardiogram during cardiac magnetic resonance imagingComput Cardiol200633269272
  • BirkholzTSchmidMNimskyCSchuttlerJSchmitzBECG artifacts during intraoperative high-field MRI scanningJ Neurosurg Anesthesiol20041627127615557829
  • GimbelJRUnexpected pacing inhibition upon exposure to the 3T static magnetic field prior to imaging acquisition: what is the mechanism?Heart Rhythm2011894494521199688
  • GimbelJRUnexpected asystole during 3T magnetic resonance imaging of a pacemaker-dependent patient with a ‘modern’ pacemakerEuropace2009111241124219556252
  • TandriHZvimanMMWedanSRLloydTBergerRDHalperinHDeterminants of gradient field-induced current in a pacemaker lead system in a magnetic resonance imaging environmentHeart Rhythm2008546246818313607
  • CalcagniniGTriventiMCensiFIn vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: role of implant geometryJ Magn Reson Imaging20082887988618821629
  • MollerusMAlbinGLipinskiMLuccaJCardiac biomarkers in patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing an MRI scanPacing Clin Electrophysiol2008311241124518811802
  • NordbeckPWeissIEhsesPMeasuring RF-induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leadsMagn Reson Med20096157057819132759
  • PictetJMeuliRWickySvan der KlinkJJRadiofrequency heating effects around resonant lengths of wire in MRIPhys Med Biol2002472973298512222860
  • FiekMRempTReithmannCSteinbeckGComplete loss of ICD programmability after magnetic resonance imagingPacing Clin Electrophysiol2004271002100415271024
  • BaserKGurayUDurukanMDemirkanBHigh ventricular lead impedance of a DDD pacemaker after cranial magnetic resonance imagingPacing Clin Electrophysiol201235e251e25321303387
  • LuechingerRZeijlemakerVAPedersenEMIn vivo heating of pacemaker leads during magnetic resonance imagingEur Heart J20052637638315618060
  • LangmanDAGoldbergIBFinnJPEnnisDBPacemaker lead tip heating in abandoned and pacemaker-attached leads at 1.5 Tesla MRIJ Magn Reson Imaging20113342643121274985
  • MartinETSandlerDAMRI in patients with cardiac devicesCurr Cardiol Rep20079637117362687
  • KanalEGimbelJRMR-conditional pacemakersAJR Am J Roentgenol2012198W502W50322528933
  • ZikriaJFMachnickiSRhimEBhattiTGrahamREMRI of patients with cardiac pacemakers: a review of the medical literatureAJR Am J Roentgenol201119639040121257892
  • MartinETComanJAShellockFGPullingCCFairRJenkinsKMagnetic resonance imaging and cardiac pacemaker safety at 1.5-TeslaJ Am Coll Cardiol2004431315132415063447
  • SommerTVahlhausCLauckGMR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 TRadiology200021586987910831713
  • NazarianSRoguinAZvimanMMClinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 teslaCirculation20061141277128416966586
  • MollerusMAlbinGLipinskiMLuccaJEctopy in patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing an MRI scanPacing Clin Electrophysiol20093277277819545340
  • NaehleCPMeyerCThomasDSafety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinationsRadiology2008249991100119011193
  • PulverAFPuchalskiMDBradleyDJSafety and imaging quality of MRI in pediatric and adult congenital heart disease patients with pacemakersPacing Clin Electrophysiol20093245045619335853
  • CohenJDCostaHSRussoRJDetermining the risks of magnetic resonance imaging at 1.5 tesla for patients with pacemakers and implantable cardioverter defibrillatorsAm J Cardiol20121101631163622921995
  • NazarianSHansfordRRoguinAA prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devicesAnn Intern Med20111554152421969340
  • BuendiaFCanoOSanchez-GomezJMCardiac magnetic resonance imaging at 1.5 T in patients with cardiac rhythm devicesEuropace20111353353821227955
  • SommerTNaehleCPYangAStrategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinationsCirculation20061141285129216966587
  • RoguinAZvimanMMMeiningerGRModern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe: in vitro and in vivo assessment of safety and function at 1.5 TCirculation200411047548215277324
  • MollerusMAlbinGLipinskiMLuccaJMagnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictionsEuropace20101294795120353963
  • StrachKNaehleCPMuhlsteffenALow-field magnetic resonance imaging: increased safety for pacemaker patients?Europace20101295296020356915
  • NaehleCPKreuzJStrachKSafety, feasibility, and diagnostic value of cardiac magnetic resonance imaging in patients with cardiac pacemakers and implantable cardioverters/defibrillators at 1.5 TAm Heart J20111611096110521641356
  • BoilsonBAWokhluAAckerNGSafety of magnetic resonance imaging in patients with permanent pacemakers: a collaborative clinical approachJ Interv Card Electrophysiol201233596721938517
  • RussoRJDetermining the risks of clinically indicated nonthoracic magnetic resonance imaging at 1.5 T for patients with pacemakers and implantable cardioverter-defibrillators: rationale and design of the MagnaSafe RegistryAm Heart J201316526627223453091
  • NazarianSHalperinHRHow to perform magnetic resonance imaging on patients with implantable cardiac arrhythmia devicesHeart Rhythm2009613814319121814
  • FarisOPSheinMJGovernment viewpoint: US Food and Drug Administration: Pacemakers, ICDs and MRIPacing Clin Electrophysiol20052826826915826256
  • ShinbaneJSCollettiPMShellockFGMR in patients with pacemakers and ICDs: defining the issuesJ Cardiovasc Magn Reson2007951317178675
  • GimbelJRThe safety of MRI scanning of pacemakers and ICDs: what are the critical elements of safe scanning? Ask me again at 10,000Europace20101291591720543196
  • JungWZverevaVHajrediniBJackleSSafe magnetic resonance image scanning of the pacemaker patient: current technologies and future directionsEuropace20121463163722237585
  • PooleJEGlevaMJMelaTComplication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registryCirculation20101221553156120921437
  • NeuzilPTaborskyMRezekZPacemaker and ICD lead extraction with electrosurgical dissection sheaths and standard transvenous extraction systems: results of a randomized trialEuropace200799810417272329
  • Expert Panel on MRSKanalEBarkovichAJBellCACR guidance document on MR safe practices: 2013J Magn Reson Imaging20133750153023345200
  • ShinbaneJSCollettiPMShellockFGMR imaging in patients with pacemakers and other devices: engineering the futureJACC Cardiovasc Imaging2012533233322421181
  • GrayRWBibensWTShellockFGSimple design changes to wires to substantially reduce MRI-induced heating at 1.5 T: implications for implanted leadsMagn Reson Imaging20052388789116275428
  • BottomleyPAKumarAEdelsteinWAAllenJMKarmarkarPVDesigning passive MRI-safe implantable conducting leads with electrodesMed Phys2010373828384320831091
  • ForleoGBSantiniLDella RoccaDGSafety and efficacy of a new magnetic resonance imaging-compatible pacing system: early results of a prospective comparison with conventional dual-chamber implant outcomesHeart Rhythm2010775075420167289
  • WollmannCGThudtKVockPGlobitsSMayrHClinical routine implantation of a dual chamber pacemaker system designed for safe use with MRI: a single center, retrospective study on lead performance of Medtronic lead 5086MRI in comparison to Medtronic leads 4592-53 and 4092-58Herzschrittmacherther Elektrophysiol201122233236239242 German22127540
  • WilkoffBLBelloDTaborskyMMagnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environmentHeart Rhythm20118657320933098
  • GimbelJRBelloDSchmittMRandomized trial of pacemaker and lead system for safe scanning at 1.5 TeslaHeart Rhythm20131068569123333721
  • SantiniLForleoGBSantiniMWorld Society of ArrhythmiasEvaluating MRI-compatible pacemakers: patient data now paves the way to widespread clinical application?Pacing Clin Electrophysiol20133627027823240924
  • GuidantINGENIO MRI/FINELINE II pacing system data collection in patients undergoing MRI (INFINITE-MRI) Available from: http://clinicaltrials.gov/ct2/show/NCT01890512. NLM identifier: NCT01890512Accessed January 13, 2014
  • WollmannCGSteinerEVockPNdikungBMayrHMonocenter feasibility study of the MRI compatibility of the Evia pacemaker in combination with Safio S pacemaker leadJ Cardiovasc Magn Reson2012146723009683
  • BiotronikProMRI study of the Entovis pacemaker system Available from: http://clinicaltrials.gov/show/NCT01761162. NLM identifier: NCT01761162Accessed January 13, 2014
  • ShellockFGWoodsTOCruesJV3rdMR labeling information for implants and devices: explanation of terminologyRadiology2009253263019789253
  • LobodzinskiSSRecent innovations in the development of magnetic resonance imaging conditional pacemakers and implantable cardioverter-defibrillatorsCardiol J2012199810422298178
  • AhmedFZMorrisGMAllenSKhattarRMamasMZaidiANot all pacemakers are created equal: MRI-conditional pacemaker and lead technologyJ Cardiovasc Electrophysiol2013241059106524016320
  • CroninEMMahonNWilkoffBLMRI in patients with cardiac implantable electronic devicesExpert Rev Med Devices2012913914622404775