102
Views
15
CrossRef citations to date
0
Altmetric
Original Research

Implant survivorship analysis after minimally invasive sacroiliac joint fusion using the iFuse Implant System®

, &
Pages 485-492 | Published online: 23 Nov 2015
 

Abstract

Introduction

Surgical revision rate is a key outcome with all permanent implants. The iFuse Implant System® is a permanent implant used to perform minimally invasive sacroiliac joint fusion. The purpose of this study is to determine the surgical revision rate after sacroiliac joint fusion surgery with this system.

Methods

Using two internal sources of information, revision surgeries were identified and linked to index surgeries. The likelihood of revision surgery was calculated using the Kaplan–Meier life table approach. Predictors of revision were explored.

Results

Four-year survivorship free from implant revision was 96.46%. Revision rate did not differ by sex and was lower for age >65. In all, 24% of revisions occurred within the first 30 days after surgery; 63.5% occurred within year 1. Implant survivorship has improved annually since the device was introduced in 2009.

Conclusion

The survivorship rate with this implant is high and improving; the rate is somewhat higher than total hip replacement but lower than that of lumbar spine procedures.

Author contributions

All authors contributed toward data analysis, drafting and revising the paper and agree to be accountable for all aspects of the work.

Disclosure

All authors are SI-BONE employees. The authors report no other conflicts of interest in this work.