Transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the Accurus vitrectomy platform to the Constellation Vision System for microincisional vitrectomy surgery

, , , , , & show all
Pages 367-377 | Published online: 19 Feb 2013

Abstract

Background

Microincisional vitrectomy surgery (MIVS) is the current standard surgical approach for pars plana vitrectomy. Historically, the most common surgical platform for vitrectomy surgery, since its introduction in 1997, has been the Accurus vitrectomy system. Recent introduction of the next generation of vitrectomy platforms has generated concerns associated with transitioning to new technology in the operating room environment. This study compared, in a matched fashion, surgical use of the Accurus vitrectomy system and the next generation Constellation Vision System to evaluate surgical efficiencies, complications, and user perceptions of this transition.

Methods

Electronic health records were abstracted as a hospital quality assurance activity and included all vitreoretinal surgical procedures at the Bascom Palmer Eye Institute, Anne Bates Leach Eye Hospital, during two discrete 12-month time periods. These two periods reflected dedicated usage of the Accurus (June 2008–May 2009) and Constellation Vision (July 2009–June 2010) systems. Data were limited to a single surgeon and evaluated for operating room (OR) total time usage/day, OR case time/case, and OR surgical time/case. Further analysis evaluated all patients undergoing combined MIVS and clear cornea phacoemulsification/intraocular lens (IOL) implantation during each individual time period to determine the impact of the instrumentation on these parameters. All records were evaluated for intraoperative complications.

Results

Five hundred and fourteen eligible patients underwent MIVS during the 2-year study windows, with 281 patients undergoing surgery with the Accurus system and 233 patients undergoing surgery with the Constellation system. Combined MIVS and phacoemulsification with IOL implantation was performed 141 times during this period with the Accurus and 158 times during the second study period with the Constellation. Total number of patients operated per day increased from 7.55 with Accurus to 8.53 with Constellation. Surgical room time decreased from 56 minutes with Accurus to 52 minutes with Constellation, and procedure time decreased from 35 minutes with Accurus to 31 minutes with Constellation (P < 0.004). Combined MIVS/phacoemulsification surgery saw similar declines in surgical room time and procedure time (P < 0.001). Subset analysis of procedures limited by case number per day (eg, four cases/day, five cases/day, six cases/day, and seven or more cases/day) showed similar outcomes with a decrease in surgical room time and procedure time. No increases in surgery-related complications were noted by quality assurance review during these time periods.

Discussion:

Transitioning to advanced surgical technology is a complex issue for the surgeon, the hospital team, and the hospital administration. This study documents improvement in three significant measures of surgical efficiency: operative number of patients per day, operative room time, and surgical procedure time that reflect the positive impact of the novel, combined, integrated, posterior and anterior, ophthalmologic surgical platform of the Constellation Vision System. These data are imperative to evaluate the impact of transition from one surgical platform to another. During this transition, hospital quality assurance review and surgeon evaluation of operative complications showed no increased concerns for the shift from the Accurus to the Constellation Vision System surgical platform. Further, both operative staff and surgeons felt that the transition to the Constellation was not associated with increases in difficulty with setup, turnover, or use and that the Constellation decreased safety concerns for surgical usage. Ultimately, in this case, new technology benefited the surgeon, the patient, and the hospital.

Rapid advances in surgical technology have occurred since the first automated vitrectomy was performed by Machemer in 1971.Citation1,Citation2 From 1971 until 2012, instruments have moved from separate systems for vitrectomy cutting, illumination, air/fluid exchange, silicone oil injection, automated scissors/forceps, and operative laser photocoagulation toward integrated platforms that include multiple technologies.Citation3Citation5

A major advance in technology occurred with the introduction of the Accurus-integrated platform in 1997 (Alcon Surgical, Fort Worth, TX, USA). This platform achieved enhanced vitrectomy cutter performance, improved surgical fluidics, automated silicone oil infusion, and incorporated a posterior segment fragmatome and an anterior segment phacoemulsification capability. The Accurus platform rapidly became the standard within the United States and internationally.

By 2008 the Accurus platform was present in over 90% of ophthalmologic operating rooms in the United States, including both hospital and ambulatory care surgical centers. The Accurus platform was characterized by an advanced cutter design, increased cutting rates to 2500 cuts per minute, gas-forced fluid infusion, dual halogen light sources, and software parameters that were established to maximize cutter efficiency at very high cut rates.

At the Bascom Palmer Eye Institute, Anne Bates Leach Eye Hospital (BPEI/ABLEH), the Accurus platform became the standard vitrectomy console from 1997 to 2008. The Vitreoretinal Surgical Service at the BPEI had extensive interest in improving surgical platforms and techniques for vitreoretinal surgery and actively pursued evaluation and acquisition of the next generation vitrectomy platform.

The Constellation Vision System (Alcon Surgical) was defined by a marked advance in cutter probe design specifically for micro-incisional vitrectomy surgery (MIVS), incorporating 23- and 25-gauge vitrectomy, markedly increased cutting rates to 5000 cuts per minute (enabled by elimination of spring technology to re-open the cutter after closure); integrated quadruple xenon light sources; had radiofrequency identification (RFID) recognition technology for identifying the cutter, light pipe, and endolaser probes; improved cassette design to eliminate loss of infusion fluidics; integrated a 532 nm solid state disc laser; and had torsional anterior segment phacoemulsification.

This next generation platform addressed many concerns for instrument limitation and fluidic compromise associated with very high cut rates, while incorporating features that had previously required independent stand-alone systems for use in complex vitrectomy surgery. Further, specific attention was given to the enhanced safety features and targeted platform improvements that were designed to increase operating room efficiency while improving patient safety; these were attractive to our hospital teams and hospital administration.

Prior to evaluation in this study, a transition period was established to educate our hospital teams, place the Constellation, establish a disposable supply support, and assist our vitreoretinal surgeons. This transition period utilized heavy educational support from the Alcon surgical support team and from in-house vitreoretinal surgeons aimed at supporting our hospital teams, with strong focus on our OR scrub teams and our OR circulating nursing teams.

This study was a follow-up to a pre-implementation review document that hypothesized reduction in OR turnover times that would enable increased surgical volume per vitreoretinal surgical room per day. In this study we compared two time periods: one in which the Accurus platform was utilized exclusively and one in which the Constellation platform was utilized exclusively. We evaluated a 12-month time period to better minimize potential case mix bias or transition bias and to capture a significant case volume for analysis. Use of the University of Miami electronic health record enabled broad data capture for evaluation of case volume per day, surgical room time per case, and surgical procedure time per case. Data sets were evaluated blinded to the platform utilized for both the Accurus and Constellation Vision System. These data provide a foundation for evaluating the selection of novel surgical systems for the ophthalmic hospital or ambulatory surgical center and delineate the impact of transition for critical technology required for vitrectomy surgery in the 21st century.

Methods

A data extract from the BPEI/ABLEH electronic health record system, satisfying internal review board requirements, was obtained for all surgeries performed by a single surgeon (TGM) during two time periods established through the hospital quality assurance program. The first time period, representing usage of the Accurus platform, was from June 2008 through May 2009. The second time period, representing usage of the Constellation platform, was from July 2009 through June 2010.

The patient’s electronic health records were then matched to BPEI’s billing system to extract the current procedural terminology (CPT) codes for each patient encounter. The two databases were then combined to form one de-identified analytic dataset. The dataset included the following variables: claim number, date of service, CPT codes, attending physician, operating room number, unique patient identifier, patient operating room in-time, surgery start time, surgery end time, patient operating room out-time.

The operative time data was manually entered into the patient electronic health record by the nursing staff on the service date as part of their standard operating procedures. Total room time was calculated by subtracting the in-room time from the out-room time. Total surgery time was calculated by subtracting the surgery start time from the surgery end time. Total surgical day time was calculated from the first time in the OR to the last time in the OR for the entire surgical day.

We eliminated procedures that could not be performed on the two platforms, such as primary scleral buckle, enucleation, or examination under anesthesia. To enhance the evaluation, we evaluated all surgical dates and then surgical dates with four, five, six, or greater then/equal to seven cases per day. Finally, we eliminated cases that did not include MIVS surgery on the surgical day evaluated, such as primary scleral buckle, enucleation, or examination under anesthesia.

To evaluate the changes in efficiency between the two platforms, we analyzed three metrics: case volume by analyzing patient volume per day, patient throughput by analyzing the operating room time, and intra-operative time by analyzing procedure time.

A subset analysis was performed on combination MIVS and phacoemulsification surgeries. Combination surgeries were defined as surgeries with both an anterior segment and a posterior segment procedure coded on the same claim.

We evaluated risk management reporting to detect any increase in operative complications, instrument concerns, or reported surgical delays. We did not measure the profitability or profit margin between the two time periods. Based on the staffing model at BPEI, it was determined that staffing levels and staff hours remained consistent between the two time periods and did not affect any change in throughput efficiency.

Statistical analysis utilized a paired t-test (Student’s paired t-test, SAS v9.3; SAS, Cary, NC, USA). Statistical significance was established as a P-value less than 0.05 for the comparative analysis.

Results

A total of 514 eligible patients identified by evaluation of the surgical electronic health records were included in this analysis. In the first 12-month study period (Accurus surgical system), 281 patients underwent vitrectomy surgery with 141 patients undergoing combined pars plana vitrectomy and phacoemulsification/IOL implantation. In the second 12-month study period (Constellation Vision System), 233 patients underwent vitrectomy surgery with 158 patients undergoing combined pars plana vitrectomy and torsional phacoemulsification/IOL implantation ().

Table 1 Overall efficiency review comparing the Accurus to the Constellation surgical platform

Evaluation of surgical efficiencies documented an increase from an average of 7.5 cases per day during the Accurus time period to 8.5 cases per day during the Constellation Vision System time period. During these study windows, overall surgical time per day decreased during the Constellation Vision System time period ().

Operating room case times averaged 58 minutes during the Accurus time period and decreased to 52 minutes during the Constellation Vision System time period. Operating room surgical times averaged 36 minutes during the Accurus time period and decreased to 31 minutes during the Constellation Vision System time period (P < 0.004) ().

Table 2 Comparison of Accurus and Constellation surgical platforms by procedure time and room time (in minutes) for all cases performed with at least seven cases per day

Table 3 Comparison of Accurus and Constellation surgical platforms by procedure time and room time (in minutes) for all cases performed regardless of number of cases per day

Table 4 Comparison of Accurus and Constellation surgical platforms by procedure time and room time (in minutes) for all cases performed with at least four cases per day

Table 5 Comparison of Accurus and Constellation surgical platforms by procedure time and room time (in minutes) for all cases

Finally, operating room surgical times for combined pars plana vitrectomy and phacoemulsification with IOL implantation averaged 43 minutes during the Accurus time period and decreased to 37 minutes during the Constellation Vision System time period (P < 0.001) (, ).

Table 6 Comparison of Accurus and Constellation surgical platforms for combined microincisional vitrectomy surgery with phacoemulsification and IOL implantation for all cases performed with at least seven cases per day

Table 7 Comparison of Accurus and Constellation surgical platforms for combined microincisional vitrectomy surgery with phacoemulsification and IOL implantation for all cases performed

Table 8 Comparison of Accurus and Constellation surgical platforms for combined microincisional vitrectomy surgery with phacoemulsification and IOL implantation for all cases performed with at least four cases per day

Table 9 Comparison of Accurus and Constellation surgical platforms by procedure time and room time (in minutes) for combined microincisional vitrectomy and phacoemulsification with IOL implantation

Ongoing surgical documentation of intra-operative and postoperative complications noted stable complication profiles as previously reported.

To determine the potential impact of case volume on efficiency evaluation, we correlated all cases and then independently evaluated datasets with cutoffs of at least four, five, six, or greater than/equal to seven cases per day. This analysis showed no statistically significant increased efficiency but clearly suggested a trend to increased efficiency with increasing case volume. Clear positive impacts were seen for each case volume noting a benefit even for surgical volumes as low as four cases per day (unreported analysis of a second data set documented improved efficiencies with case volumes averaging approximately two cases per day) ().

Statistical analysis noted statistically significant improvements in efficiencies for operative time with both decreased procedure and room time associated with transition to the Constellation Vision System (P < 0.0004). Additionally, for combined MIVS pars plana vitrectomy and torsional phacoemulsification/IOL implantation, a marked decrease in both procedure and room time were documented (P < 0.0001) ().

Discussion

Vitreoretinal surgical advances have been rapid since the first automated vitrectomy surgical units were developed four decades ago.Citation6Citation12 During this period, marked improvements in instrument design contributed to significant increases in patient safety and improved surgical outcomes. Initially, Machemer and others focused on multifunction single-port instrumentation, but instrument design continuously evolved toward smaller instrument sizes and integrated multiport pars plana vitrectomy surgery.Citation1,Citation13Citation15

Further modifications, from Machemer’s initial vitrectomy systemCitation1 continued to decrease instrument size, ultimately achieving a standard instrument approach with 20-gauge instruments placed through the sclera in a three-port pars plana vitrectomy approach whereby one port was utilized for infusion, one port for illumination, and one port for the cutter/forceps/scissors. Seeking smaller wounds, more rapid wound healing, and elimination of transscleral repetitive instrument passage led to smaller gauge instruments that incorporated transconjunctival/transscleral trocars focused on 23-gauge and 25-gauge surgical instruments.Citation16Citation20 This MIVS approach has rapidly become the current standard with a transition from 20-gauge sutured sclerotomies to 23-gauge and 25-gauge trocared instrument approaches not requiring suture closure. Currently, 27-gauge (and smaller) instruments are available and in design.

A major impetus for the development of a novel, advanced, integrated platform design has been the surgical requirements of increased cutting rates, improved intraocular fluidics, enhanced lighting, and deliverable endolaser. This shift to small gauge surgery, along with the interest in an integrated platform, necessitated the design and development of a novel next-generation surgical platform. This ideal platform would incorporate very high speed cutting, stable real-time evaluation of intraocular fluidics, markedly improved illumination sources, capacity for delivery of high centistoke liquids, microvolume deliveries, and integrated laser technology.Citation21Citation37 These characteristics define the minimal surgical requirements for an integrated platform designed for the 21st century.

Additionally, this study took place during the transition from 20-gauge pars plana vitrectomy to 23/25-gauge MIVS. This surgical platform transition recognized the initial concerns for increased risk of endophthalmitis, choroidal detachment, iatrogenic retinal tear and/or detachment, or postoperative hypotony associated with microincisional vitrectomy.Citation38Citation42 Focused investigation and training on wound construction and surgical technique were instrumental in the use of transconjunctival, trocared, nonsutured pars plana vitrectomy in the surgical care of our patients. Fortunately, these concerns have been alleviated by clinical reviews that have not noted increased complication profiles with microincisional vitrectomy.Citation43

In this study, we evaluated the “real world” surgical performance of the platform that has been the “gold standard” in vitreoretinal surgical systems, the Accurus platform, and contrasted that performance with the next generation vitreoretinal platform, the Constellation Vision System.Citation44Citation48 This comparison utilized standard metrics incorporated within the BPEI/ABLEH surgical electronic medical record to determine case volume per room per day, operating room time per case, and surgical time per case along with total operating time per room per day. These metrics allow a standardized comparison of technologies but require evaluation after a transition window when each technology has achieved a steady implementation state of usage. Further, evaluation of a large time frame coupled with high surgical numbers, as in this study, eliminates many potential biases to evaluation of the utility of new technology, such as the Constellation Vision System.

This study documents the increased efficiency of the Constellation platform relative to the prior standard Accurus system. The Constellation achieved increased patient surgical cases per day by decreasing both operative case time and room time. This increase is related to an integrated design that facilitates case turnover, vitrectomy instrument and cassette setup (and particularly combined vitrectomy/phacoemulsification instrument setup), and enhancements with integrated instrument recognition technology, prepopulated user settings, surgeon-controlled endolaser parameters, and rapid priming associated with improved fluidics. These platform design changes clearly target improvements for the surgeon, the OR team, and the patient.

Ultimately, the decision to transition from existing technology to new technology should focus first on enhanced patient care, including patient safety, improved anatomic outcomes, improved visual outcomes, and translation to a better quality of life.

Disclosure

The authors report no relevant financial conflicts of interest in this work. Dr Murray consults for Thrombogenics and Alcon.

References

  • MachemerRBuettnerHNortonEWParelJMVitrectomy: a pars plana approachTrans Am Acad Ophthalmol Otolaryngol19717548138205566980
  • NortonEWMachemerRNew approach to the treatment of selected retinal detachments secondary to vitreous loss at cataract surgeryAm J Ophthalmol19717247057075110860
  • ChangSIntraocular gasesRyanSGlaserBMRetina2nd edSt LouisMosby1994
  • AbramsGWAzenSPMcCuenBWIIVitrectomy with silicone oil or long-acting gas in eyes with severe proliferative vitreoretinopathy: results of additional and long term follow-up. Silicone Study Report #11Arch Ophthalmol19971153353449076205
  • ChangSVitrectomyDukerJYanoffMAOphthalmologyLondonMosby2008
  • KoenigSBMielerWFHanDPCombined phacoemulsification, pars plana vitrectomy, and posterior chamber intraocular lens insertionArch Ophthalmol1992110110111041497524
  • ThompsonJTde BustrosSMichelsRGResults and prognostic factors in vitrectomy for diabetic vitreous hemorrhageArch Ophthalmol19871051911953813948
  • ChangSLincoffHZimmermanNJGiant retinal tears: surgical techniques and results using perfluorocarbon liquidsArch Ophthalmol19891077617662719587
  • GardnerTBlankenshipGWProliferative diabetic retinopathy: principles and techniques of surgical treatmentRyanSGlaserBMRetina3rd edSt LouisMosby2012
  • CollGEChangSSunJPerfluorocarbon liquid in the management of retinal detachment with proliferative vitreoretinopathyOphthalmology19941026306387724180
  • MelbergNSThomasMADickinsonJDSurgical removal of subfovealchoroidal neovascularization: ingrowth site as a predictor of visual outcomeRetina1996161901958789856
  • SjaardaRNMichelsRGMacular PuckerRyanSGlaserBMRetina3rd edSt LouisMosby2012
  • FreemanWMacular Hole Study GroupVitrectomy for the treatment of full-thickness stage 3 or stage 4 macular holes: results of a multicentered randomized clinical trialArch Ophthalmol199711511219006420
  • BorneMJTasmanWRegilloCOutcomes of vitrectomy for retained lens fragmentsOphthalmology19961039719768643257
  • ScottIUFlynnHWLaiMChangSAzenSPFirst operation anatomic success and other predictors of postoperative vision after complex retinal detachment repair with vitrectomy and silicone oil tamponadeAm J Ophthalmol200013074575011124293
  • IbarraMSHermelMPrennerJLHassanTSLonger-term outcomes of transconjunctival sutureless 25-gauge vitrectomyAm J Ophthalmol200513983183615860288
  • LakhanpalRRHumayunMSde JuanEJrOutcomes of 140 consecutive cases of 25-gauge transconjunctival surgery for posterior segment diseaseOphthalmology200511281782415878061
  • OshimaYOhjiMTanoYSurgical outcomes of 25-gauge transconjunctivalvitrectomy combined with cataract surgery for vitreoretinal diseasesAnn Acad Med Singapore20063517518016625266
  • KadonosonoKYamakawaTUchinoEComparison of visual function after epiretinal membrane removal by 20- gauge and 25-gauge vitrectomyAm J Ophthalmol200614251351516935609
  • FineHFIranmaneshRIturraldeDSpaideRFOutcomes of 77 consecutive cases of 23-gauge transconjunctival vitrectomy surgery for posterior segment diseaseOphthalmology20071141197120017544779
  • ShimadaHNakashizukaHHattoriTMoriRIncidence of endophthalmitis after 20- and 25-gauge vitrectomy: Causes and preventionOphthalmology2008152215222018930557
  • ShinodaHShinodaKSatofukaSVisual recovery after vitrectomy for macular hole using 25-gauge instrumentsActa Ophthalmol20088615115517725613
  • BamonteGMuraMTanHSHypotony after 25-gauge vitrectomyAm J Ophthalmol2010151115616020970109
  • FujiiGYde JuanEJrHumayunMSInitial experience using the transconjunctival sutureless vitrectomy system for vitreoretinal surgeryOphthalmology20021091814182012359600
  • FujiiGYDe JuanEJrHumayunMSA new 25-gauge instrument system for transconjunctival sutureless vitrectomy surgeryOphthalmology20021091807181212359598
  • NamYChungHLeeJYKimJGYoonYHComparison of 25- and 23-gauge sutureless microincision vitrectomy surgery in the treatment of various vitreoretinal diseasesEye20102419119942940
  • EckardtCTransconjunctival sutureless 23-gauge vitrectomyRetina20052520821115689813
  • OkamotoFOkamotoCSakataNChanges in corneal topography after 25-gauge transconjunctival sutureless vitrectomy versus after 20-gauge standard vitrectomyOphthalmology20071142138214118054632
  • KadonosonoKYamakawaTUchinoEComparison of visual function after epiretinal membrane removal by 20- gauge and 25-gauge vitrectomyAm J Ophthalmol200614251351516935609
  • ShinodaHShinodaKSatofukaSVisual recovery after vitrectomy for macular hole using 25-gauge instrumentsActa Ophthalmol20088615115517725613
  • ChangSEKimKHKangSWRetinal breaks associated with the induction of posterior vitreous detachmentAm J Ophthalmol200914761012101619327743
  • TanHSMuraMde SmetMDIatrogenic retinal breaks in 25-gauge macular surgeryAm J Ophthalmol2009148342743019477712
  • OshimaYShimaCWakabayashiTMicroincision vitrectomy surgery and intravitreal bevacizumab as a surgical adjunct to treat diabetic traction retinal detachmentOphthalmology2009116592793819269033
  • RizzoSComparative study between standard 25-gauge vitrectomy systems and new ultra-high speed 25-gauge system with duty cycle control in the treatment of various vitreoretinal diseasesPaper presented at: Vail Vitrectomy 2010March 13–17, 2010Vail, CO
  • RizzoSGenovesi–EbertFMurriS25-gauge, sutureless vitrectomy and standard 20-gauge pars plana vitrectomy in idiopathic epiretinal membrane surgery: a comparative pilot studyGraefes Arch Clin Exp Ophthalmol2006244447247916421743
  • ChenE25-Gauge transconjunctival sutureless vitrectomyCurr Opin Ophthalmol200718318819317435424
  • KellnerLWimpissingerBStolbaU25-gauge vs 20-gauge system for pars plana vitrectomy: a prospective randomized clinical trialBr J Ophthalmol200791794594817202200
  • KunimotoDYKaiserRSIncidence of endophthalmitis after 20- and 25-gauge vitrectomyOphthalmology2007114122133213717916378
  • ScottIUFlynnHWJrDevSEndophthalmitis after 25-gauge and 20-gauge pars plana vitrectomy: incidence and outcomesRetina200828113814218185150
  • TabanMSharmaSVenturaAAKaiserPKEvaluation of wound closure in oblique 23- gauge sutureless sclerotomies with Visante optical coherence tomographyAm J Ophthalmol2009147110110718835474
  • TabanMVenturaAASharmaSKaiserPKDynamic evaluation of sutureless vitrectomy wounds: an optical coherence tomography and histopathology studyOphthalmology2008115122221222819041476
  • MartidisAChangTSSutureless 25-gauge vitrectomy: risky or rewarding?Ophthalmology2007114122131213218054631
  • KaiserRSPrennerJScottIUThe Microsurgical Safety Task Force: evolving guidelines for minimizing the risk of endophthalmitis associated with microincisional vitrectomy surgeryRetina201030469269920386097
  • ParkeDW3rdSiskRAHoustonSKMurrayTGOcular hypertension after intravitreal triamcinolone with vitrectomy and phacoemulsificationClin Ophthalmol2012692563122791974
  • MurrayTGTornambePDugelPTongKBEvaluation of economic efficiencies in clinical retina practice: activity-based cost analysis and modeling to determine impacts of changes in patient managementClin Ophthalmol20115913925 Epub July 12, 201121792278
  • SiskRABerrocalAMFeuerWJMurrayTGVisual and anatomic outcomes with or without surgery in persistent fetal vasculatureOphthalmology2010117112178218320619897
  • SiskRAMurrayTGCombined phacoemulsification and sutureless 23-gauge pars plana vitrectomy for complex vitreoretinal diseasesBr J Ophthalmol20109481028103220519234
  • HartleyKLSmiddyWEFlynnHWJrMurrayTGPars plana vitrectomy with internal limiting membrane peeling for diabetic macular edemaRetina200828341041918327132