Simulation of airbag impact on eyes with different axial lengths after transsclerally fixated posterior chamber intraocular lens by using finite element analysis

, &
Pages 263-270 | Published online: 05 Feb 2015

Abstract

Purpose

To determine the biomechanical response of an impacting airbag on eyes with different axial lengths with transsclerally fixated posterior chamber intraocular lens (PC IOL).

Materials and methods

Simulations in a model human eye were performed with a computer using a finite element analysis program created by Nihon, ESI Group. The airbag was set to be deployed at five different velocities and to impact on eyes with three different axial lengths. These eyes were set to have transsclerally fixated PC IOL by a 10-0 polypropylene possessing a tensile force limit of 0.16 N according to the United States Pharmacopeia XXII.

Results

The corneoscleral opening was observed at a speed of 40 m/second or more in all model eyes. Eyes with the longest axial length of 25.85 mm had the greatest extent of deformity at any given impact velocity. The impact force exceeded the tensile force of 10-0 polypropylene at an impact velocity of 60 m/second in all eyes, causing breakage of the suture.

Conclusion

Eyes with transsclerally fixated PC IOL could rupture from airbag impact at high velocities. Eyes with long axial lengths experienced a greater deformity upon airbag impact due to a thinner eye wall. Further basic research on the biomechanical response for assessing eye injuries could help in developing a better airbag and in the further understanding of ocular traumas.

Introduction

Airbags have saved thousands of lives since their introduction in the early 1980s. Airbags protect the passengers against a crash by providing a padding device, which allows the impacting and impacted surface to deform, thereby extending the duration of the impact and reducing its severity.Citation1 Although airbags have substantially reduced the rate of mortality and morbidity, those who have survived may suffer from various fatal and nonfatal injuries to the head, eyes, neck, chest, or arms.Citation2,Citation3 Ocular traumas are among the most severe airbag-induced injuries due to a high risk for detrimental vision impairment after an impact.Citation4 Airbag-induced ocular trauma includes corneal abrasion, corneoscleral laceration, subluxated lens, endothelial cell loss, cyclodialysis, choroidal rupture, globe rupture, retinal detachment, and periorbital fracture.Citation5Citation11

Given the initiatives in cataract or corneal surgeries for better vision, an increase in the population achieving the visual acuity test criteria for operating a vehicle would be assumed.Citation4 Any driver could be at risk of airbag-induced ocular trauma. In a 1991–1998 review of 97 patients with airbag-associated eye trauma, 50% of traumas were limited to the anterior segment, 6% to the posterior segment, and 44% to both.Citation12 Postcataract surgery patients with implantation of an intraocular lens (IOL) may be at a higher risk with impact for wound rupture, subluxation of a posterior chamber (PC) IOL, anterior capsule rupture, and dislocation of the lens.Citation11,Citation13,Citation14

Unlike with human bones and ribcages, the injury biomechanics of soft organs, such as human eyes, are difficult to simulate due to limited available mechanical information. Cadavers and dummies have been used for research purposes; however, the physiological and biological properties of these eyes do not resemble living eyes, making the trauma research more difficult and results only marginally reliable. Therefore, creating a humanlike eye with raw data from the human eye for biomechanical simulations using finite element analysis (FEA) would help to investigate and better explain the physical and physiological responses to impact injuries.Citation1

We have previously developed a simulation model resembling a human eye based on the information obtained from cadaver eyes and applied three-dimensional FEA to determine the physical and mechanical conditions of impacting foreign bodies that cause an intraocular foreign body.Citation15 This model human eye was also used in our studies on airbag impact in a postradial keratotomy eye and on a post-transsclerally fixated PC IOL eye.Citation4,Citation16 In general, the capsular bag is considered the best position for IOL implantation. When the capsular support is insufficient or absent, ciliary sulcus fixation or, more recently, scleral fixation, is recommended.Citation4 In this study, we extended the simulation model after renovation to further determine the physical and mechanical response of an impacting airbag deploying at additional velocities on transsclerally fixated PC IOL eyes with different axial lengths, especially surveying the mechanical threshold in a highly myopic eye.

Materials and methods

The model human eye was created and used in simulations with a computer using an FEA program, PAM-CRASH (Nihon ESI, Tokyo, Japan), described elsewhere.Citation15 The material properties and geometry of the model were obtained from past experiments with three pairs of human cadaver eyes.Citation15 Poisson ratios of the cornea at 0.420 kg/mm3 and the sclera at 0.470 kg/mm3 were used to determine the standard stress strain curves for the cornea and sclera.Citation17Citation19 The reference point for globe rupture was then calculated to be at a strain of 18.0% and stress of 9.45 MPa for the cornea, and at a strain of 6.8% and stress of 9.49 MPa for the sclera.Citation15 The cornea was assumed to be spherical, with a central thickness of 0.5 mm and a central radius of curvature of 7.8 mm. The anterior chamber was set at a depth of 5.1 mm. A transsclerally fixated polymethyl methacrylate PC IOL element was integrated at the original physiological position of the crystal lens. The vitreous length was assumed to be 18.6 mm, and the posterior curvature of the retina was assumed to be 12.0 mm. The mass densities of ocular tissues from past reports were applied as follows:Citation15,Citation20 cornea, 1.149 kg/mm3; sclera, 1.243 kg/mm3; and vitreous and aqueous humor, 1.002 kg/mm3 and 1.000 kg/mm3, respectively. A new approach in this study was the addition of model human eyes with different axial lengths, representing a normal eye with a normal axial length of 23.85 mm, a hyperopic eye with a shorter axial length of 21.85 mm, and a myopic eye with a longer axial length of 25.85 mm. Eyes with different axial lengths were created by setting the mass density of the cornea and sclera as constants, and the element types including the three layers of the model eye (outer, middle, and inner) as variables for meshing principles.Citation15 A threshold for the thickness reduction ratio was 0.5 (50%). The elastic properties and meshing principles of the model human eye were similar to those in previous reports.Citation15,Citation16 The vitreous model as a solid mass was also assigned with a hydrostatic pressure of 20 mmHg (2.7 KPa). The changes in the deformity of the eye and the strain induced were calculated by the Virtual Performance Solver (VPS) (Nihon ESI KK) and evaluated by color mapping. In this study, mapping properties were renovated for more finite displaying due to the development of computer technology following the previous study.Citation4

The simulation of PC IOL implantation was based on a corneoscleral incision size of 6.0 mm and wound closure with 10-0 polypropylene sutures in five places. A detailed description of a single-piece polymethyl methacrylate IOL and its elasticity and mass density of 2.85 GPa and 1.185 kg/mm3, respectively, are described elsewhere.Citation4 The suturing method was done the same way as reported, with a 10-0 polypropylene looped suture (by an Alcon Surgical PC 91/4 circular needle; Alcon, Inc., Hünenberg, Switzerland) and the loop of each suture was tied with the girth–hitch method.Citation4 Two tying needles were placed at the 4 and 10 o’clock positions, and the sutures exited the sclera 1.0 mm posterior to the limbus and were tied directly to the superficial sclera. The limit of tensile force (N) of a 10-0 polypropylene suture was specified to be 0.16 N by the United States Pharmacopeia XXII.Citation21,Citation22 Breakage of the fixating suture is assumed to occur in the middle of the suture when the strain becomes intolerable due to deformity of the eye caused by airbag impact. Element deletion was incorporated in the VPS, with rupture of the eye and the breakage of the suture defined as the strain exceeding the corneoscleral and fixation suture threshold, respectively.

A biomechanical head of a dummy was created, assuming that everything excluding the eye was a solid element, to reduce the computing time. The Hybrid III model was modifiedCitation4 by replacing the head of the dummy with a biomechanical model of the head in which the IOL-fixated model eye was inserted.Citation4,Citation23 For simplification of calculation, the angular momentum of the impact missile was ignored in this study. The impact missile for the airbag simulation was set up as described.Citation15 The impact missile was a decahedron that was composed of a cuspidal hexahedron and a rectangular parallelepiped, so as to be blunt in shape.Citation15 In addition to our previous study, the airbag was deployed at five different velocities (20 m/second, 30 m/second, 40 m/second,50 m/second, and 60 m/second) and it impacted eyes of three different axial lengths.

The simulation program was similar to that of a previous report by Digital Equipment Corporation AlphaStationCitation4 using the FEA program, Pam-Crash version 1998 (Nihon ESI KK).Citation4 Similarly, suture strain was recorded sequentially in all velocities and the deformity of the eye was displayed sequentially in milliseconds in slow motion. Breakage of the suture was defined as the point when the strain exceeded the tensile tolerance.

Results

The postoperative PC IOL model human eye with an axial length of 23.85 mm created by the VPS is shown in . The skull surrounding the eye was assumed to be a solid element to minimize the computing time for impact simulation (). The thickness of the eye wall depicted by the color maps showed correlation with the axial lengths of the eyes (). The eye with a longer axial length is shown in orange, representing a thinner eye wall compared to eyes with normal and shorter axial lengths in red. The deformity of the three eyes with transsclerally sutured IOL at five impact velocities (20 m/second, 30 m/second, 40 m/second, 50 m/second, and 60 m/second) is captured in slow motion (shown in , , and ). Corneoscleral deformity was observed after 0.1~0.2 ms of impact in all velocities. The deformity of the eye was small at 20 m/second and 30 m/second. Deformity reached the threshold of corneoscleral strain (0.068) at 40 m/second, 50 m/second, and 60 m/second, causing the model eye to rupture. The tensile force on the 10-0 polypropylene suture of the three eyes is shown in , , and . At 0.3 ms after the impact, the eye with the longest axial length had the greatest extent of deformity at any given impact velocity ().

Figure 1 Frontal view of the model eye with transsclerally fixated PC IOL.

Notes: (A) Four and 10 o’clock positions; Cornealscleral incision site for IOL insertion at the 12 o’clock position (B) sagittal view of the model eye.
Abbreviations: IOL, intraocular lens; PC, posterior chamber.
Figure 1 Frontal view of the model eye with transsclerally fixated PC IOL.

Figure 2 A biomechanical model of the head.

Notes: Parts other than the eye were assumed to be solid elements, and the impact object was placed adjacent to the eyeball to reduce the computing time of the airbag impact simulation.
Figure 2 A biomechanical model of the head.

Figure 3 The thickness of the eye wall (mm) is depicted from cold to warm colors.

Notes: A warmer color of red represents a thicker eye wall; a myopic eye with a longer axis shows a thinner eye wall.
Figure 3 The thickness of the eye wall (mm) is depicted from cold to warm colors.

Figure 4 Sequential deformity of a normal axial length model eye upon airbag impact at five different velocities, shown in ms intervals.

Notes: (A) The deformity became irreversible, suggesting a corneoscleral rupture at 0.3 ms after impact velocities over 40 m/second, shown in hot pink. (B) The tensile force of the 10-0 polypropylene suture was exceeded only at an impact velocity of 60 m/second after 0.3 ms, suggesting fixation suture breakage (lower: 4 o’clock fixation suture; upper: 10 o’clock fixation suture).
Figure 4 Sequential deformity of a normal axial length model eye upon airbag impact at five different velocities, shown in ms intervals.

Figure 5 Sequential deformity of a long axial length (myopic) model eye upon airbag impact at five different velocities, shown in ms intervals.

Notes: (A) Deformity became irreversible at 0.3 ms after impact velocities over 40 m/second, shown in hot pink. (B) The tensile force of the 10-0 polypropylene suture was exceeded only at an impact velocity of 60 m/second after 0.3 ms, suggesting fixation suture breakage (lower: 4 o’clock fixation suture; upper: 10 o’clock fixation suture).
Figure 5 Sequential deformity of a long axial length (myopic) model eye upon airbag impact at five different velocities, shown in ms intervals.

Figure 6 Sequential deformity of a short axial length (hyperopic) model eye upon airbag impact at five different velocities, shown in ms intervals.

Notes: (A) Deformity became irreversible at 0.3 ms after impact velocities over 40 m/second, shown in hot pink. (B) The tensile force of the 10-0 polypropylene suture was exceeded only at an impact velocity of 60 m/second after 0.3 ms, suggesting fixation suture breakage (lower: 4 o’clock fixation suture; upper: 10 o’clock fixation suture).
Figure 6 Sequential deformity of a short axial length (hyperopic) model eye upon airbag impact at five different velocities, shown in ms intervals.

Figure 7 Sagittal view of the model eye at 0.3 ms after the airbag impact.

Notes: The myopic eye, the eye with the longest axial length of 25.85 mm, showed the most corneoscleral deformity, with multiple color spots on the anterior portion of the model eye.
Figure 7 Sagittal view of the model eye at 0.3 ms after the airbag impact.

The tensile force on the suture exceeded the critical value at only the velocity of 60 m/second in all eyes. This breakage of the suture occurred in all eyes after 0.3 ms impact, with the longest axial length exceeding the breaking force initially between 0.2 ms and 0.3 ms. There was no breakage of the scleral fixation suture at 40 m/second and 50 m/second, even when the force on the corneoscleral incision suture was continuously greater with increasing velocity. At 60 m/second impact, the suture tensile force was exceeded at the sites of the scleral fixation suture, leading to both ocular rupture and breakage of the suture (, , and ).

Discussion

In a frontal crash automobile accident, drivers move forward rapidly and may be struck forcefully by a rearward-expanding bag at high velocities, causing the impacted eye to deform.Citation1,Citation12 Despite its overall protective effects, this forceful inflation of the airbag itself could be the major contributor of ocular injuries, even in minor accidents such as bumper-to-bumper contact or even a flat tire.Citation14 According to the US Federal Motor Vehicle Safety Standards, airbags take only 20 m/second to fully inflate, which is about four times faster than blinking.Citation25 Airbag deployment occurs 15 ms after an impact with complete expansion at 50 ms and deflation by 100 ms.Citation8 Automobile airbags are reported to deploy at a speed ranging from 81–322 kph (22–89 m/seconds).Citation8,Citation25Citation27 Considering that some situations exceeded the impact velocities in the previous report,Citation4 we added 50 m/second and 60 m/second velocities to this study. From our results, ocular rupture and breakage of the sutures occurred at 60 m/second. Therefore, no further velocities were performed for the simulation. Since eyeballs are viscoelastic, their response to impact injury is rate sensitive.Citation1 Therefore, we simulated the impacts of various velocities within the range of speed of deployment in this study to see the biomechanical response for different axial lengths. Our present study demonstrated that eyes remained intact up to a 30 m/second impact speed. Corneoscleral opening occurred at 40 m/second, 50 m/second, and 60 m/second, and eventually the ocular and fixation suture ruptured at 60 m/second. The results suggested that ocular rupture and fixating suture breakage could happen even at mid-speed of airbag deployment. Furthermore, at the fastest velocity of 60 m/second, corneoscleral rupture was observed as early as 0.1~0.2 ms after the airbag impact, suggesting that all eyes regardless of axial length would be vulnerable to a higher speed of impact.

In the era of refractive surgeries, the effect on the ocular integrity of cataract surgical techniques may depend on the physical properties of the eye. To our knowledge, no published data on the mechanical properties, especially on the tensile strength, of the zonule of Zinn, which is essential for the simulation of airbag impact on a phakic or a pseudophakic eye, are available. Transsclerally fixating the PC IOL with a 10-0 polypropylene suture is still the most widely performed method for complicated eyes without capsular support. In contrast, since the tensile strain of the polypropylene suture is available from the manufacturer, the simulation of airbag impact on the transsclerally-IOL-sutured eye can be established by modification of our previous simulation model.Citation4 We have already demonstrated and assessed the mechanical properties of an eye with a transsclerally fixated PC IOL in a past report showing IOL subluxation with an impact velocity of 40 m/second.Citation4 However, individual variations regarding axial lengths, age, or ethnicity were not considered in previous studies.Citation4 Then, we carried out the current simulated study using the PC IOL transsclerally fixing model with several modifications.

It is known that there has been a global increase in the prevalence of myopia over the past half-century, particularly in affluent, industrialized areas of East Asia.Citation24 Higher myopia, which means a longer axial length, is associated with comorbidities that increase the risks of severe and irreversible loss of vision, such as retinal detachment, subretinal neovascularization, dense cataract, and glaucoma.Citation24 Therefore, longer eyes with thinner walls would be much more vulnerable to any kind of impact trauma. This study demonstrated that these eyes with a longer axial length are prone to a more extensive ocular deformity, causing irreversible strain on the eye, leading to ocular and fixation suture rupture if present. Longer eyes showed breakage of the fixation suture as early as 0.1 ms after the airbag impact.

Other possible factors affecting the integrity of the driver’s eye upon airbag impact would be age, sex, weight, usage of spectacles, and ethnicity.Citation2 Some data indicated that women of small stature and shallower eye sockets, common in Asians, are more likely to be injured by airbag impacts because they are likely to sit too close to the wheel, which could cause fatal or nonfatal injuries when the airbag is deployed.Citation2,Citation3 Elderly women with smaller statures would be physiologically characterized with normal or smaller eye axial lengths.Citation3 This physiological factor would make this population at high risk for blunt ocular injuries. Our data showed that even at the medium velocity of 30 m/second, corneoscleral incisions and IOL would be at risk in normal to shorter axial length eyes. Sitting closer to the steering wheel would decrease the distance to the driver’s face and, therefore, increase the impact velocity and its force on these individuals.

Studies of the wound upon blunt airbag impact are of interest to ophthalmologists. The study of trauma could provide insights on how the eye responds to impact injury, and would thus help to establish preventative surgical measures such as revising the surgical wound, performing IOL fixation methods, improving suture material, or inserting IOL implants. At the same time, blunt airbag impact studies could also help automobile industries make more adaptable airbags customized for different ethnic backgrounds and physical characteristics.

Despite our careful calculation of the simulation model based on cadaver eyes, there are still limitations to our study.Citation15 First, since studies are based on only three pairs of cadaver eyes, the results do not completely resemble a living humanlike eye response, or represent the normal population. Second, in order to reduce the time for computer calculation, the impact object was placed almost adjacent to the surface of the eyeball for the impact simulation. Therefore, further study on the distance of the deploying airbag to the driver in regard to the change in velocity would be necessary to determine the extent of impact that is likely to happen in real situations. Finally, the vitreous model was a solid mass with a physiological intraocular pressure of 20 mmHg.Citation15 The age factor for vitreous viscosity or other physical characteristics were not taken into account as variables in our simulation. However, we believe that our goal to assess the mechanical properties of eyes with different axial lengths with a fixated IOL suture was achieved by demonstrating that the corneoscleral incision would open at a medium velocity followed by IOL subluxation, with the eye with the longest axial length being more susceptible to impact injuries. Furthermore, at a maximum velocity of 60 m/second, the impact force exceeded the fixated suture force, causing suture breakage.

Conclusion

In conclusion, even with improvements in automobile industry technology, there is still need for the conscientious reporting of vision-threatening airbag impact injuries. The lack of basic research data on the biomechanical response has retarded the development of more sophisticated models or dummies. Therefore, this current method of injury biomechanics research for assessing eye injuries would help advance our knowledge on the mechanism of ocular traumas and reduce the need for animal experiments and cadavers, which are difficult to obtain and could pose ethical issues. Finally, an ability to predict anatomical or structural injury upon airbag impact would be critical to safety. Continuous injury biomechanics research would be helpful not only in developing an adequate evaluation system for airbag safety, but also in further understanding ocular preventative measures. These may include the use of protective goggles, the revision of PC IOL fixations sutures, minimal surgical incisions, or placing the PC IOL without sutures.

Acknowledgments

We thank Mr Tony Yang for editing this manuscript.

Disclosure

The authors report no conflicts of interest in this work.

References

  • VianoDCKingAIMelvinJWWeberKInjury biomechanics research: an essential element in the prevention of traumaJ Biomech19892254034172674145
  • Segui-GomezMLevyJRomanHThompsonKMMcCabeKGrahamJDDriver distance from the steering wheel: perception and objective measurementAm J Public Health19998971109111110394328
  • MancheEEGoldbergRAMondinoBJAir bag-related ocular injuriesOphthalmic Surg Lasers19972832462509076803
  • UchioEKadonosonoKMatsuokaYGotoSSimulation of air-bag impact on an eye with transsclerally fixated posterior chamber intraocular lens using finite element analysisJ Cataract Refract Surg200430248349015030846
  • de VriesSGeerardsAJLong-term sequelae of isolated chemical “airbag” keratitisCornea200726899899917721304
  • BakerRSFlowersCWSinghPSmithACaseyRCorneoscleral laceration caused by air-bag traumaAm J Ophthalmol199612167097118644816
  • ScottIUGreenfieldDSParrishRKAirbag-associated injury producing cyclodialysis cleft and ocular hypotonyOphthalmic Surg Lasers199627119559578938806
  • FukagawaKTsubotaKKimuraCCorneal endothelial cell loss induced by air bagsOphthalmology199310012181918238259280
  • HanDPRetinal detachment caused by air bag injuryArch Ophthalmol199311110131713188216003
  • ScottIUJohnGRStarkWJAirbag-associated ocular injury and periorbital fracturesArch Ophthalmol19931111258280191
  • VichninMCJaegerEAGaultJAJeffersJBOcular injuries related to air bag inflationOphthalmic Surg Lasers19952665425488746576
  • AlmahmoudTBarssPVehicle occupant restraint systems impact on eye injuries: a reviewSurv Ophthalmol201459333434424359757
  • OnwuzuruigboCJFuldaGJLarnedDHailstoneDTraumatic blindness after airbag deployment: bilateral lenticular dislocationJ Trauma19964023143168637090
  • GhafouriABurgessSKHrdlickaZKZagelbaumBMAir bag-related ocular traumaAm J Emerg Med19971543893929217535
  • UchioEOhnoSKudohJAokiKKisielewiczLTSimulation model of an eyeball based on finite element analysis on a supercomputerBr J Ophthalmol199983101106111110502567
  • UchioEOhnoSKudohKKadonosonoKAndohKKisielewiczLTSimulation of air-bag impact on post-radial keratotomy eye using finite element analysisJ Cataract Refract Surg200127111847185311709260
  • BuzardKAIntroduction to biomechanics of the corneaRefract Corneal Surg1992821271381591207
  • GreenePRClosed-form ametropic pressure-volume and ocular rigidity solutionsAm J Optom Physiol Opt198562128708784083331
  • HoeltzelDAAltmanPBuzardKChoeKStrip extensiometry for comparison of the mechanical response of bovine, rabbit, and human corneasJ Biomech Eng199211422022151602763
  • ReichelEMillerDBlancoEMastandunoRThe elastic modulus of central and perilimbal bovine corneaAnn Ophthalmol19892162052082764431
  • United States Pharmacopeial Convention. Webpage on Internet. The United States Pharmacopeia. The National Formulary, 22nd RevisionRockville, MD, USA1990
  • WhitleyJQPrewittMJKusyRPRelationship of the diameter and tensile strength of nylon sutures to the USP specification and the effect of preconditioningJ Appl Biomater19901431532010148999
  • RuanJSPrasadPCoupling of a finite element human head model with a lumped parameter Hybrid III dummy model: preliminary resultsJ Neurotrauma19951247257348683624
  • FosterPJJiangYEpidemiology of myopiaEye (Lond)201428220220824406412
  • National Highway Traffic Safety AdministrationAir Bag Deployment CharacteristicsSpringfield, VANational Technical Information Service1992
  • SchreckRMRouhanaSWSantrockJPhysical and chemical characterization of airbag effluentsJ Trauma19953845285327723091
  • BallDCBouchardCSOcular morbidity associated with airbag deployment: a report of seven cases and a review of the literatureCornea200120215916311248820