Mathematical evaluation of the influence of multiple factors on implant stability quotient values in clinical practice: a retrospective study

, , &
Pages 1525-1532 | Published online: 11 Oct 2016

Abstract

Objectives

The objective of this study is to mathematically evaluate the influence of multiple factors on implant stability quotient values in clinical practice.

Patients and methods

Resonance frequency analysis was performed at T1 (measured immediately at the time of implant placement) and at T2 (measured before dental restoration) in 177 patients (329 implants). Using a multivariate linear regression model, we analyzed the influence of the following eleven candidate factors: sex, age, maxillary/mandibular location, bone type, immediate/delayed implantation, bone grafting (presence or absence), insertion torque, I-/II-stage healing pattern, implant diameter, implant length, and T1–T2 time interval.

Results

The following factors were identified to significantly influence the implant stability quotient (ISQ) values at T1: insertion torque, bone grafting, I-/II-stage healing pattern, immediate/delayed implantation, maxillary/mandibular location, implant diameter, and sex. In contrast, the ISQ values at T2 were significantly influenced only by three factors: implant diameter, T1–T2 time interval, and insertion torque.

Conclusion

Among the eleven candidate factors, seven key factors were found to influence the T1-ISQ values, while only three key factors influenced the T2-ISQ values. Both T1 and T2-ISQ values were found to be influenced by implant diameter and insertion torque. T1 was influenced specifically by the sex of the patient, the location (maxillary or mandibular), the implantation mode (immediate/delayed implantation), the healing stage, and the absence or presence of bone graft materials.

Introduction

Since the pioneering work of Branemark in 1952,Citation1 dental implants have become a widely used treatment option in the past few decades. Dental implants are used to provide mechanical support for various dental prostheses, such as crowns, bridges, dentures, and orthodontic apparatuses. The basis for such a desired support function by an implant is its mechanical stability. This is generally described, as a function of time, as a primary and a secondary stability. The primary stability is largely based on an immediate mechanical anchoring of the implant in the surrounding bone upon surgical implantation. The secondary stability is achieved by a biological healing process – called osseointegration – and it forms a direct structural and functional connection between the implant and the neoformed surrounding bone tissues, without any interpositioned connective tissue.Citation2 In clinical practice, the degree of implant stability is considered to be an important parameter to estimate the scope of mechanical loading capability and to provide baseline information as a tool to assess the clinical outcome and time course.Citation3

A large number of efforts have been made to identify and to develop novel techniques for the quantitative assessment of the implant stability. An ideal technique should be simple, noninvasive, and clinician friendly. One of the candidate techniques is Periotest.Citation4,Citation5 However, the Periotest readings do not always correspond precisely to a biomechanical parameter, since they are strongly related to the excitation direction and position.Citation4,Citation6 As an alternative, resonance frequency analysis (RFA) was proposed to estimate implant stability in 1996.Citation7,Citation8 RFA is performed using a small transducer that is fastened to implants/abutments. This transducer has a vertical beam that is attached using two piezoceramic elements. One piezoceramic element produces a vibration consisting of a small sinusoidal signal in the range of 5–15 kHz in steps of 25 Hz and the other element analyzes the response of the transducer to the vibration.Citation8 The peak amplitude of the response is coded into a parameter called the implant stability quotient (ISQ). ISQ values range from 0 (indicating a totally mobile implant) to 100 (indicating a perfectly stable implant–bone complex).Citation9 The ISQ value is positively correlated to the mechanical stability of an implant. RFA is a noninvasive technique and shows a high reproducibility of results.Citation10,Citation11 In recent years, RFA has become one of the most widely used techniques to assess stability on the spot in order to determine the possible loading occasion and to assess the long-term survival of dental implants.Citation12 ISQ values ranging from 60 to 80 are widely accepted as a standard for achieving primary stability.Citation13Citation15 In the clinic, ISQ values of at least 55 at the time of implant placement might be considered as representing clinically relevant stability and possible predictors of successful osseointegration.Citation13,Citation15 As for immediate implant loading, an ISQ measure of 60–65 will ensure a good prognosis.Citation16

Attempts to achieve early functionality of implants have been continuously pursued in the field of oral implantology. Immediate implantation is associated with several advantages, such as the reduction of surgical trauma, the shortening of the treatment time, and the improved preservation of surrounding bone and soft tissue. In cases with sufficient primary stability, evidence is presented in the literature that immediate implantation (or even immediate loading) yields equal efficacy in terms of long-term success and esthetic outcome compared to delayed implantation.Citation17 However, the technique of immediate implantation is still a challenge with respect to achieving sufficient primary stability of the implant that, if not achieved, may lead to a higher implant failure rate.Citation18 Careful case selection must be performed to avoid treatment failures and esthetic complications when deciding between immediate and delayed implant placements.Citation18 Therefore, it is also of great significance to estimate the case-specific ISQ values in order to create a detailed treatment plan. For this purpose, continuous efforts are made to elucidate the influence of various factors on ISQ values using the RFA technique. In previous studies, some of the factors that were investigated that possibly influence the ISQ values are implant design,Citation9 insertion torque,Citation19 immediate/delayed implantation,Citation20 drilling design,Citation21,Citation22 bone density,Citation23 bone grafting, and mechanical loading pattern.Citation24 Most of these studies demonstrated a significant influence of such factors on the basis of the assessment of the relationship between ISQ values and single and/or several factors. Albeit so, the weight coefficients of the various influencing factors for the ISQ values remained unrevealed, so that most of the decisions made by clinicians are still made largely based on practical experience. A mathematical model may play a critical role to thoroughly assess the individual contribution of the various factors on ISQ values in clinical situations by performing multivariate analyses. Hitherto, there is still a lack of such an adequate mathematical model.

In this study, we retrospectively analyzed both the demographic and clinical data of 329 implants from 177 patients by using a multivariate linear regression analysis. We wished to determine the contribution of each of the individual factors to the ISQ values in a clinical set up in order to provide baseline data for the creation of a mathematical model to estimate the likely ISQ value for an individual case.

Patients and methods

Patients and implants

The conduct of this study was approved by the Review Board of the Best & Easy Dental Clinic, People’s Republic of China. It is routine for all patients at the Best & Easy Dental Clinic to give an informed written consent for potential inclusion in clinical studies. In this retrospective study, we reviewed the data of all the patients who received implant treatment in the Best & Easy Dental Clinic, Hangzhou, People’s Republic of China, from 2012 to 2015. SICace implants (SIC Invent AG, Basel, Switzerland) with different diameters and lengths were used. All the implants were placed by the same surgeon. In total, 178 patients with 331 implants were included in the study. There were two implant failures (the failure rate was 0.6%) over this time period. The data of these two implants were not included in the following analysis.

General inclusion and exclusion criteria for implant treatments

In the Best & Easy Dental Clinic, we used the American Society of Anesthesiologist (ASA) classifications (ASA1, ASA2, and ASA3) to evaluate the systemic health status of patients for establishing the inclusion criteria for implant treatment.Citation25 Briefly, well-controlled status of the patient in case of systemic disease (to tolerate the surgery) was considered. Regarding the oral health, patients with only mild and/or moderate (but well controlled) periodontitis and patients with a good oral hygiene status were included. Patients were excluded from implant surgery if they were pregnant or would be unable to withstand the stress of dental implant surgery (ASA4–5). Patients were also excluded if they had severe/uncontrolled periodontitis.

Patient records

We retrospectively collected the following data from patients: potential candidate factors possibly influencing the ISQ values: (X1) sex, (X2) age, (X3) maxillar/mandibular location, (X4) immediate/delayed implantation, (X5) presence or absence of bone grafting, (X6) implant diameter, (X7) implant length, (X8) I/II-stage healing pattern, (X9) insertion torque, (X10) bone type, and (X11) T1–T2 time interval.

According to the record of implantation procedure, the II-stage healing pattern was used only if the insertion torque was <20 N or the ISQ value <65. The bone type of the implant sites was categorized into types I, II, III, and IV according to the classification of Lekholm and Zarb.Citation26

The ISQ values (measured with Osstell™ Mentor; Integration Diagnostic Ltd., Goteborg, Sweden) were recorded from the mesial, distal, lingual, and buccal sites of each implant at both T1 (immediately after implantation) and T2 (immediately before restoration and loading). Typically, after 6–12 weeks, patients received the restoration therapy. In a few cases, the patients received the restoration/loading therapy as late as 1 year.

Statistical analysis

Initially, we used the Kruskal–Wallis test to compare the ISQ values from the mesial, distal, lingual, and buccal sites of implants at T1 and T2. We used paired t-tests to assess the difference in ISQ values at T1 and T2 for either immediately placed or delayed placed implants. We also applied paired t-tests to assess the influence of immediate/delayed implantation on ISQ values at T1 or T2. Thereafter, we performed multivariate linear regression analyses to determine the weight coefficients of the eleven candidate factors possibly influencing the ISQ values at both T1 and T2 time points. All the statistical analyses were performed using an SPSS® 21.0 software (SPSS Inc., Chicago, IL, USA). Level of significance was set at P<0.05, and the confidence level at 95%.

In the multivariate linear regression analysis, the categories of the influencing factors were transformed into numbers as follows: (X1) male =1, female =2; (X3) maxillary =1, mandible =2; (X4) immediate =1, delayed =2; (X5) bone grafting: no =1, yes =2; and (X8) I-stage =2, II-stage =1. Dummy variables were used for bone types (X10): type 1 =100, type 2 =010, type 3 =001, and type 4 =000. The numbers for the remaining factors were directly used for statistical analysis.

Results

A total of 329 implants from 177 patients were included in this study. The descriptive characteristics of all the patients and implants are listed in .

Table 1 Descriptive characteristics of the patients and implants

There were no significant differences among the ISQ values measured from the labial, lingual, distal, and mesial sites at the time points of either immediately after implantation (T1) or right before loading (T2; ). For both immediate and delayed implantation, the ISQ values at T1 were significantly lower than those at T2 (). At T1, the ISQ values of immediately placed implants were significantly lower than those of delayed placed implants. At T2, there was no significant difference between the ISQ values of immediately placed implants and those of the delayed implants ().

Table 2 Kruskal–Wallis analysis to compare the values of implant stability quotient that were measured from the labial, lingual, distal, and mesial sites using resonance frequency analysis technique immediately after implantation (T1) and right before loading (T2), respectively

Table 3 Paired t-test analysis to compare the values of implant stability quotient (ISQ) between immediately after implantation (T1) and right before loading (T2) for either immediate implantation or delayed implantation (horizontal)

At T1, the multivariate linear regression analysis showed that the ISQ values were significantly influenced by the following seven factors: (X1) sex, (X3) maxillary/mandibular location, (X4) immediate/delayed implantation, (X5) bone graft, (X6) implant diameter, (X8) I-stage/II-stage implantation, and (X9) insertion torque (). The relative weight coefficients (presented as standardized coefficients) of these factors were as follows: (X1) 0.111, (X3) 0.121, (X4) 0.148, (X5) −0.235, (X6) 0.119, (X8) 0.241, and (X9) 0.286. The equation used to calculate the ISQ values with the contribution of each factor is as follows: Y(T1)=57.263+1.317(X1)+1.471(X3)+1.836(X4)4.990(X5)+1.669(X6)+2.961(X8)+0.131(X9).(1)

Table 4 Multivariate linear regression analysis to analyze the weight coefficient of each influencing factor for the values of implant stability quotient that were measured immediately after implantation T1

At T2, the ISQ values were significantly influenced by three factors: (X6) implant diameter, (X9) insertion torque, and (X11) T1–T2 time interval (). The equation used to calculate the ISQ value with the contribution of each factor is as follows: Y(T2)=56.988+4.080(X6)+0.048(X9)+0.014(X11).(2)

Table 5 Multivariate linear regression analysis to analyze the weight coefficient of each influencing factor for the values of implant stability quotient that were measured right before loading T2

The relative weight coefficients of these factors were as follows: (X6) 0.414, (X9) 0.150, and (X11) 0.191.

Discussion

In the field of dental implantology, a consensus has been reached that sufficient primary stability is critical in order to provide a mechanically stable microenvironment for the proper establishment of implant osseointegration – the biological basis for the secondary stability and for full implant functionality. Consequently, the availability of numeric stability values is a prerequisite for the estimation of the loading time schedule and the assessment of the long-term success rate of implants. RFA is a tool for the rapid, easy, objective and noninvasive measurements of the stability of implants without causing any patient discomfort. In a recent well-controlled in vitro study, ISQ values measured with RFA were found to be proportional to the mechanical stability of implants.Citation19 On this basis, ISQ values are widely used as a basic parameter for clinical decision making. A precise and reliable estimation of the ISQ value in each case is thus a fundamental need to provide grounding for designing a realistic and accurate treatment plan. In this study, using a retrospective analysis, the possible role of eleven different candidate factors was considered. On these grounds, we formulated a mathematical model to estimate the weight coefficients of candidate factors for a more precise assessment of both the primary and secondary implant stabilities ( and ).

The design of an implant is one of the most fundamental elements to affect the implant primary and secondary stability.Citation9 The design features consist of two major categories: 1) the macrodesign, such as thread design and body shape and 2) the microdesign, such as the implant topography.Citation9 Gehrke et al recently indicated that the conical implants with a wide pitch were associated with significantly greater primary stability values than the semiconical implants with narrow pitch bores. In our study, we used only one implant type (SICace) with an identical macro- and microdesign, which thus may exclude the potential influence of implant design factors. Therefore, we did not include the implant design as a candidate factor in our analysis. Similarly, the preparation technique of the surgical site may also potentially influence implant stability.Citation27 This parameter was also excluded in this study, since the surgical site preparation was performed by the same experienced implantologist using one single implant system.

Bone type was not found to be a determining parameter influencing either T1 or T2 in our study. This finding was consistent with a recent 1-year follow-up study with 101 implants.Citation28 In that study, it was concluded that the baseline microstructural bone characteristics that were assessed by histomorphometric and microtomographic analyses did not significantly influence implant stability. Furthermore, using a similar classification method as in this study, the bone type was found not to be a significant influencing parameter either.Citation29

Apart from the implant design, the diameter and length of implants were other implant-related factors that might influence implant stabilities. In a recent in vitro biomechanical study, the primary stability of wider implants was found to be significantly higher in hard bone than the narrower implants using insertion torque as a parameter.Citation23 However, such differences have not been confirmed when using ISQ values as the estimator. These conflicting data might originate from a much smaller correlation (than generally assumed) between micromotion and insertion torque values than those obtained with ISQ measurements.Citation19 In fact, in a small-scale prospective clinical trial, Han et alCitation13 showed that ISQ values were not correlated with implant diameter values over a 12-week postoperative monitoring time period. However, in our retrospective study with 329 implants, the implant diameter was found to be a significant parameter influencing ISQ values both at T1 () and T2 (). At T1, using Equationequation 1, the 1.5 mm diameter difference between the 3.5 mm and 5 mm implants could be transformed into a difference of 2.503±1.131 (calculated by multiplying 1.5 by 1.669) in ISQ values. However, its weight coefficient was 0.119, which was quite similar with X1, X3, and X4, but much lower than X5, X8, and X9. These data suggested that the implant diameter was a significant but relatively mild influencing factor to estimate ISQ values at T1. In contrast, such 1.5 mm difference in implant diameter could be transformed into a difference of 6.120±1.047 at T2. The weight coefficient of implant diameter (0.414) was also much higher than X9 (0.150) and X11 (0.191) at this time point. These results thus indicated that the implant diameter was a major influencing factor on ISQ values at T2 (). Previous studies also showed that implant diameters could significantly influence ISQ values.Citation30,Citation31 In contrast to this, the implant length was not found to be a significant influencing parameter at either T1 or T2 time points in our study. This finding was consistent with a previous study showing that implant length did not significantly influence primary stabilities of implants.Citation29 However, the implant length still might play a role in influencing implant stability provided that singly calculated correlations between implant length and implant stabilities were performed.Citation32,Citation33 Furthermore, in particular cases, such as in patients with low bone quality, the optimization of the implant length and diameter should be considered in order to achieve higher primary implant stability values.Citation34

The maxillary/mandibular location was expected to represent a determining parameter influencing ISQ values, and indeed most implants in the maxilla had an ISQ of <60, and those in the mandible had an ISQ of >60.Citation35 It was also found that the ISQ values were generally higher in the mandible (59.8) than in the maxilla (55.0), but when using cylindrical implants, then they were not associated with a significant difference.Citation29 Furthermore, a similar phenomenon was also observed by Gehrke and Neto.Citation30 In contrast to this, mandibular implants were found to show statistically higher ISQ values than maxillary implants.Citation31 In our study, we showed that the maxillary/mandible implant location was clearly a significant influencing factor at T1, but not at T2. According to the Equationequation 1, the mandibular location might confer implants with 1.471±0.652 (mean ± standard error [SE]) higher values than those of the maxillary location. The weight coefficient of this factor was 0.121, which indicated its mild influence. This finding may also explain why a significant difference was not always detectable, even though a higher value was always found in the mandibular implants.

Immediate implantation is able to significantly shorten the clinical treatment time. Therefore, immediate implantation has been extensively evaluated (provided favorable conditions are given) in the last 2 decades, and it has been reported to yield success rates ranging from 92.7% to 98%.Citation36 The 7-year cumulative survival rate for immediately placed implants with an immediate loading scheme could also reach 94.6% success rate.Citation37 In a long-term follow-up study, no significant differences in the success rates and in the esthetic outcomes between immediately and delayed placed implantsCitation17 were reported. Gehrke et alCitation20 recently showed that delayed placed implants bore insignificantly higher ISQ values than the immediately placed implants. In our study, we showed that immediate/delayed implantation was a significant influencing factor on ISQ values at T1, at which a delayed implantation might confer implants with 1.836±0.664 (mean ± SE) higher ISQ values than immediate implants do (). However, at T2, this parameter is not significantly different anymore between the two groups (). These data from multivariate linear regression analyses were consistent with those from paired t-test (). These findings showed that, with a careful selection of cases, an immediate implantation exhibited no significant difference in secondary stabilities when comparing with delayed implantation. However, immediate/delayed implantation can result in significantly different ISQ values when considering maxillary locations.Citation38

Similarly, for some other candidate factors, conflicting findings were found regarding the relationship between sex and ISQ values. Previous studies showed that males were associated with either significantly higher,Citation39 or significantly lowerCitation40 or similarCitation41 ISQ values when comparisons were done with females. Guler et alCitation42 showed that the sex parameter indeed influenced the ISQ values significantly only if a second measurement was done. This inconsistency may be due to a large variation of the experimental conditions, such as the choice of the measurement time point, special implant locations and inclusion of different types of populations/ethnic groups. In our study, the female patients showed 1.317±0.622 (mean ± SE) higher ISQ values than the males (which was a significant difference at T1, but not at T2). We did not identify a significant influence of the age of the patient on the ISQ values at either T1 or T2.

In our study, the need of bone grafting indeed negatively influenced ISQ values. A 4.990±0.622 (mean ± SE) lower ISQ value could be expected when there was such a need. This sounded reasonable since such a need was indeed associated with significantly smaller bone coverage of the implants. The II-stage healing pattern showed significantly higher ISQ values (2.961±0.622 [mean ± SE]) than the I-stage healing pattern at T1. This was also not unexpected since the II-stage healing pattern was performed with insertion torques <20 N or the ISQ value <65 in this study. At T2, this factor became insignificant, which suggested that I/II-stage implantation might not influence the osseointegration process. Consistently, I/II-stage implantation was previously shown not to result in different degrees of osseointegration.Citation43

One limitation of this study is that the equations might be specific for the implantologist, this implant system and/or this dental clinic. Careful interpretation is thus needed if extrapolation of the current data is planned to estimate ISQ values for patients/implants of other implantologists. However, with this study, we would like to provide a mathematical basis to analyze the weight coefficients of potential influencing factors. Every implantologist can establish his or her own equation to more precisely estimate ISQ values for the future cases. In future studies, we will further investigate the reliability and accuracy of this mathematical model for other types of implants.

Conclusion

Among the eleven candidate parameters, seven key factors influencing the ISQ values at T1 were identified, and only three key factors at T2. Within the limitations of this study, the mathematical model used enabled us to evaluate not only the significance but also the weight coefficients of various influencing parameters, which thus provides a viable novel method to more accurately estimate the ISQ values of implants.

Acknowledgments

This study was supported by the funds of China Scholarship Council, National Natural Science Foundation of China (grant nos 81400475 and 81470724), and Zhejiang Provincial Natural Science Foundation of China (grant no LY13H140006).

Disclosure

The authors report no conflicts of interest in this work.

References

  • BranemarkPIOsseointegration and its experimental backgroundJ Prosthet Dent1983503994106352924
  • GuoCYMatinlinnaJPTangATEffects of surface charges on dental implants: past, present, and futureInt J Biomater2012201238153523093962
  • DottoreAMKawakamiPYBecharaKStability of implants placed in augmented posterior mandible after alveolar osteotomy using resorbable nonceramic hydroxyapatite or intraoral autogenous bone: 12-month follow-upClin Implant Dent Relat Res201416333033623148779
  • CaulierHNaertIKalkWJansenJAThe relationship of some histologic parameters, radiographic evaluations, and Periotest measurements of oral implants: an experimental animal studyInt J Oral Maxillofac Implants19971233803869197104
  • TeerlinckJQuirynenMDariusPvan SteenbergheDPeriotest: an objective clinical diagnosis of bone apposition toward implantsInt J Oral Maxillofac Implants19916155611843494
  • DerhamiKWolfaardtJFFaulknerGGraceMAssessment of the periotest device in baseline mobility measurements of craniofacial implantsInt J Oral Maxillofac Implants19951022212297744442
  • HuangHMChiuCLYehCYLinCTLinLHLeeSYEarly detection of implant healing process using resonance frequency analysisClin Oral Implants Res200314443744312869006
  • MeredithNAlleyneDCawleyPQuantitative determination of the stability of the implant-tissue interface using resonance frequency analysisClin Oral Implants Res1996732612679151590
  • GehrkeSANetoUTDDel FabbroMDoes implant design affect implant primary stability? A resonance frequency analysis-based randomized split-mouth clinical trialJ Oral Implantol2015416E281E28625438099
  • MeredithNAssessment of implant stability as a prognostic determinantInt J Prosthodont19981154915019922740
  • Al-JetailySAl-DosariAAAssessment of Osstell and Periotest(R) systems in measuring dental implant stability (in vitro study)Saudi Dent J201123172123960497
  • Lozano-CarrascalNSalomó-CollOGilabert-CerdàMFarré-PagésNGargallo-AlbiolJHernández-AlfaroFEffect of implant macro-design on primary stability: a prospective clinical studyMed Oral Patol Oral Cir Bucal2016212e214e22126827067
  • HanJLulicMLangNPFactors influencing resonance frequency analysis assessed by Osstell mentor during implant tissue integration: II. Implant surface modifications and implant diameterClin Oral Implants Res201021660561120666787
  • HuwilerMAPjeturssonBEBosshardtDDSalviGELangNPResonance frequency analysis in relation to jawbone characteristics and during early healing of implant installationClin Oral Implants Res200718327528017355357
  • SimCPLangNPFactors influencing resonance frequency analysis assessed by Osstell mentor during implant tissue integration: I. Instrument positioning, bone structure, implant lengthClin Oral Implants Res201021659860420666786
  • SennerbyLMeredithNResonance frequency analysis: measuring implant stability and osseointegrationCompend Contin Educ Dent1998195493498 quiz 5049693511
  • ManganoFGManganoCRicciMSammonsRLShibliJAPiattelliAEsthetic evaluation of single-tooth Morse taper connection implants placed in fresh extraction sockets or healed sitesJ Oral Implantol201339217218122126700
  • KohRURudekIWangHLImmediate implant placement: positives and negativesImplant Dent20101929810820386212
  • Brizuela-VelascoAAlvarez-ArenalAGil-MurFJRelationship between insertion torque and resonance frequency measurements, performed by resonance frequency analysis, in micromobility of dental implants: an in vitro studyImplant Dent201524560761126288162
  • GehrkeSAda Silva NetoUTRossettiPHWatinagaSEGiroGShibliJAStability of implants placed in fresh sockets versus healed alveolar sites: early findingsClin Oral Implants Res201627557758226037595
  • GehrkeSAGuiradoJLBettachREvaluation of the insertion torque, implant stability quotient and drilled hole quality for different drill design: an in vitro InvestigationClin Oral Implants Res Epub201638
  • DeliGPetroneVDe RisiVTadicDZafiropoulosGGLongitudinal implant stability measurements based on resonance frequency analysis after placement in healed or regenerated boneJ Oral Implantol201440443844725106008
  • RomanosGEDelgado-RuizRASacksDCalvo-GuiradoJLInfluence of the implant diameter and bone quality on the primary stability of porous tantalum trabecular metal dental implants: an in vitro biomechanical studyClin Oral Implants Res Epub2016224
  • WentaschekSSchellerHSchmidtmannISensitivity and specificity of stability criteria for immediately loaded splinted maxillary implantsClin Implant Dent Relat Res201517suppl 2e542e54925536516
  • RinaldiMGanzSDMottolaAComputer-Guided Applications for Dental Implants, Bone Grafting, and Reconstructive Surgery (adapted translation)ChinaElsevier2015
  • LekholmUZarbGAPatient selection and preparationBranemarkP-IZarbGAAlbrektssonTTissue Integrated Prostheses: Osseointegration in Clinical DentistryChicagoQuintessence1985199209
  • RastelliCFalisiGGattoRImplant stability in different techniques of surgical sites preparation: an in vitro studyOral Implantol2014723339
  • DiasDRLelesCRLindhCRibeiro-RottaRFMarginal bone level changes and implant stability after loading are not influenced by baseline microstructural bone characteristics: 1-year follow-up. Clin Oral Implants ResEpub201512919
  • BischofMNedirRSzmukler-MonclerSBernardJPSamsonJImplant stability measurement of delayed and immediately loaded implants during healingClin Oral Implants Res200415552953915355394
  • GehrkeSANetoUTDDoes the time of osseointegration in the maxilla and mandible differ?J Craniofac Surg20142562117212025329842
  • LiajeAOzkanYKOzkanYVanliogluBStability and marginal bone loss with three types of early loaded implants during the first year after loadingInt J Oral Maxillofac Implants201227116217222299093
  • TsolakiINNajafiBTonsekarPPDrewHJSullivanAJPetrovSDComparison of osteotome and conventional drilling techniques for primary implant stability: an in vitro studyJ Oral Implantol Epub201633
  • TozumTFTurkyilmazIBalBTInitial stability of two dental implant systems: influence of buccolingual width and probe orientation on resonance frequency measurementsClin Implant Dent Relat Res201012319420119438965
  • BarikaniHRashtakSAkbariSBadriSDaneshparvarNRoknAThe effect of implant length and diameter on the primary stability in different bone typesJ Dent (Tehran)201310544945524910653
  • NedirRBischofMSzmukler-MonclerSBernardJPSamsonJPredicting osseointegration by means of implant primary stabilityClin Oral Implants Res200415552052815355393
  • PenarrochaMUribeRBalaguerJImmediate implants after extraction. A review of the current situationMed Oral2004923424215122126
  • BaroneAMarconciniSGiammarinaroEMijiritskyEGelpiFCovaniUClinical outcomes of implants placed in extraction sockets and immediately restored: a 7-year single-cohort prospective studyClin Implant Dent Relat Res Epub2016216
  • GranicMKatanecDVucicevic BorasVImplant stability comparison of immediate and delayed maxillary implant placement by use of resonance frequency analysis – a clinical studyActa Clin Croat2015543826058235
  • ZixJKessler-LiechtiGMericske-SternRStability measurements of 1-stage implants in the maxilla by means of resonance frequency analysis: a pilot studyInt J Oral Maxillofac Implants200520574775216274149
  • BrochuJFAndersonJDZarbGAThe influence of early loading on bony crest height and stability: a pilot studyInt J Prosthodont200518650651216335170
  • OstmanPOHellmanMWendelhagISennerbyLResonance frequency analysis measurements of implants at placement surgeryInt J Prosthodont20061917783 discussion 8416479765
  • GulerAUSumerMDuranISandikciEOTelciogluNTResonance frequency analysis of 208 Straumann dental implants during the healing periodJ Oral Implantol201339216116722103915
  • DegidiMDaprileGPiattelliAPrimary stability determination of implants inserted in sinus augmented sites: 1-step versus 2-step procedureImplant Dent201322553053324013393