Small coronary vessel angioplasty: outcomes and technical considerations

Pages 915-922 | Published online: 11 Oct 2010

Abstract

Small vessel (<3 mm) coronary artery disease is common and has been identified as independent predictor of restenosis after percutaneous coronary intervention. It remains controversial whether bare-metal stent (BMS) implantation in small vessels has an advantage over balloon angioplasty in terms of angiographic and clinical outcomes. Introduction of drug-eluting stent (DES) has resulted in significant reduction in restenosis and the need for repeat revascularization. Several DESs have been introduced resulting in varying reduction in outcomes as compared with BMS. However, their impact on outcomes in small vessels is not clearly known. It is expected that DES could substantially reduce restenosis in smaller vessels. Large, randomized studies are warranted to assess the impact of different DESs on outcomes in patients with small coronary arteries.

Introduction

Coronary artery disease (CAD) is the leading cause of mortality and morbidity in the Western world.Citation1 Percutaneous coronary intervention (PCI) utilizing balloon angioplasty and stenting is a predominant treatment strategy for patients with CAD. Lesions involving the small coronary arteries (with a diameter of <3 mm) are the most relevant in terms of prevalence, and they account for around 40%–50% of all coronary stenosis.Citation2 Atherosclerosis of small arteries remains a major challenge to revascularization procedures, as coronary artery bypass grafting is limited by high rates of technical failure,Citation3 and PCI is associated with an increased risk of restenosis and adverse outcome. Small coronary vessel angioplasty remains the independent predictor of repeat revascularization and adverse cardiac events.Citation2,Citation4,Citation5 Percutaneous treatment of CAD has evolved from percutaneous balloon angioplasty to bare-metal stent (BMS) and more recently to drug-eluting stent (DES) implantation. As compared with balloon angioplasty, BMS prevents both early elastic recoil and late vascular remodeling in large vessels (>3 mm).Citation6Citation10 However, this benefit is not shown in small coronary vessels (<3 mm).Citation11Citation14 Stent implantation results in arterial injury, initiating a vasculoproliferative cascade with smooth muscle cell proliferation and migration resulting in neointimal hyperplasia. The amount of neointimal hyperplasia is largely independent of vessel size, and thus, late luminal loss, an angiographic measure of neointimal hyperplasia, is similar across a wide range of vessel diameters.Citation15,Citation16 Also, small vessels are more prone to restenosis than larger vessels because they are less able to accommodate neointimal tissue without compromising blood flow.Citation17

In this review, we describe the impact of balloon angioplasty, BMSs, and DESs on the outcomes after small vessel coronary intervention.

BMS vs balloon angioplasty in small vessels

Several randomized trials have compared stenting and percutaneous transluminal coronary angioplasty (PTCA), in terms of clinical and angiographic outcomes, in coronary arteries with a reference vessel diameter (RVD) of <3 mm. Agostoni et alCitation18 evaluated these outcomes in a meta-analysis, and they included 13 studies involving 4,383 randomized patients: 2,097 to PTCA and 2,286 to stenting. The studies showing mortality, myocardial infarction (MI) rates, target lesion revascularization (TLR), and combined major adverse cardiac events (MACE) are shown in .

Table 1 Comparison of angiographic characteristics, death, MI, TLR, and MACEs, following BMS and balloon angioplasty

The mean age of patients in these studies was 62.1 ± 10 years, with 77.2% men and 28.9% with diabetes. BMSs were used in 6 trials (Park et alCitation12 ISAR-SMART, BESMART, SISA, RAP, CHIVAS), whereas a stent coated with heparin (SISCA, COAST, Kinsara et al)Citation21 or silicon carbide (SVS) or phosphorylcholine (ISAR-SMART, LASMAL, LASMAL 11) was used in the remainder. RVD was 2.33 ± 0.29 mm in the stent group and 2.31 ± 0.29 mm in PTCA group. The mean lesion length was 10.1 ± 5.2 mm, and crossover rate was 22.2% among PTCA group. Postprocedural mean diameter stenosis and minimal luminal diameter were significantly better after stenting as compared with PTCA.

The follow-up period ranged from 6–16 months. Death and MI rates did not differ significantly between groups (1.3% and 3.1% in the stent group vs 1.7% and 4.2% after PTCA, respectively). Stenting showed a significant reduction of the risk of TLR when compared with PTCA group (14.9% vs 18.7%, respectively). However, authors have mentioned that there was significant heterogeneity among the trials. Among the studies where optimal PTCA result (diameter stenosis <20%) was achieved, there was no difference in the repeat TLR rates (15% vs 16.7%). On the contrary, among trials in which a suboptimal PTCA result was achieved, stenting resulted in significant reduction of repeat TLR rates (14.8% vs 20.4%). MACE rates were higher in PTCA group (17.6% vs 22.7%) and were mainly driven by repeat TLR. Angiographic follow-up was performed in 81.9% of the patients, and the angiographic restenosis was seen in 27.8% patients in stent group compared with 35.8% in the PTCA group with significant heterogeneity. Optimal PTCA group has similar angiographic restenosis rates as stent group.

Therefore, stenting appears safe with small vessel CAD and significantly reduces angiographic restenosis and repeat revascularization rates. However, optimal PTCA in this group of patients have achieved comparable results to BMS implantation. The revascularization rates remain high with PTCA or BMS implantation in patients with small vessel CAD.

BMS vs DES in small vessels

Paclitaxel-eluting stents vs BMS in small vessels

There are no dedicated trials comparing paclitaxel-eluting stent (PES) and BMS in small coronary arteries. However, large PES trials have described substudy results in small vessels. In TAXUS IV trial,Citation28 there were 176 patients with RVD < 2.5 mm randomized to PES and BMS. Angiographic restenosis rate and 12-month TLR rate in the PES group were significantly lower than that in the BMS group, respectively (10.2% and 5.6% vs 38.5% and 20.6%; P < 0.001).

In the TAXUS V trial,Citation29 more patients with complex lesions were investigated. In the patient group treated with the 2.25-mm stent, both PES and BMS have similar acute outcomes. However, at the 9-month follow-up, the angiographic restenosis rates and TLR rates were significantly lower in PES group as compared with BMS group (31% and 10.4% vs 49.4% and 21.5%; P = 0.01 and P = 0.03, respectively). In this post hoc analysis, 9-month MACE rates were 18.9% vs 26.9%; P = 0.23, which did not reach statistical significance.

In the TAXUS VI trial,Citation30 angiographic and clinical outcomes were followed up to 9 months in complex subset of patients. In the subgroup with small vessels (RVD < 2.5 mm), in-stent late lumen loss was considerably smaller in the PES group than in the BMS group (0.23 ± 0.45 mm vs 0.95 ± 0.52 mm; P < 0.0001), explaining the significantly lower angiographic restenosis and TLR observed in the PES group (7.3% and 5.0% vs 40.4% and 29.7%, respectively; P < 0.001).

Taking these TAXUS subgroup results in consideration, PES seems to confer clinical benefits in patients with small vessels compared with BMS. This is mainly because of marked inhibition of neointimal hyperplasia by PES. These results could have been influenced by the release kinetics of drug, polymer used, and stent platform. TAXUS® Express® stent (Boston Scientific Inc, Natick, Massachusetts) was used in TAXUS IV trial, and TAXUS Express2 stent was used in TAXUS V and TAXUS VI trials. TAXUS Express stent consists of balloon-expandable stent with Translute™ polymer-coating containing paclitaxel (Boston Scientific Inc). TAXUS Express2 stent is composed of a balloon-expandable Express2 stent with a triblock copolymer coating with paclitaxel. The release kinetics of drug was slow in TAXUS IV and V trials, whereas this was moderate in TAXUS VI trial.

Sirolimus-eluting stent vs BMS in small vessels

Sirolimus-eluting stent (SES) has been tested in pivotal SIRIUS trial,Citation31 and the substudy involving small vessels (≤2.75 mm) showed significantly lower TLR rates with SES as compared with BMS (6.6% vs 22.3%; P < 0.0001). Similarly, angiographic substudyCitation32 has shown significantly lower restenosis in SES group (17.6% vs 42.7%; P < 0.001).

The SES-SMART trial,Citation33 which enrolled patients with small vessels (mean RVD = 2.2 mm), indicated that the incidence of TLR and MACE in the SES arm was significantly lower compared with BMS arm (7.0% and 9.3% vs 21.1% and 31.3%; P = 0.002 and P < 0.001, respectively). Angiographic restenosis in the SES arm was also significantly lower compared with the BMS arm (9.8% vs 53.1%, P < 0.001).

Lee et alCitation34 evaluated the predictive factors for restenosis following implantation of SES in small coronary arteries (<2.8 mm) in an observational study. They identified lesion length and restenotic lesions as independent predictors of angiographic restenosis in small vessels.

Zotarolimus-eluting stent vs BMS in small vessels

ENDEAVOR program has assessed the efficacy of zotarolimus- eluting stent (ZES) in patients with CAD. ENDEAVOR IICitation35 trial has randomly assessed outcomes after ZES and BMS in patients with native CAD. In a subgroup analysis of patients who had angiographic follow-up at 8 months, results were reported for 371 patients with RVD < 3 mm. There were significantly less angiographic restenosis rates with the use of ZES in both <2.5-mm and 2.5–3 mm group as compared with BMS (18.2% vs 38.6%; odds ratio [OR] = 0.47; 95% confidence interval [CI]: 0.28–0.79 in <2.5-mm group and 4.6% vs 35.1%; OR = 0.13; 95% CI: 0.05–0.32 in 2.5–3.0 mm group, respectively). They have also reported significantly less TLR rate with the use of ZES as compared with BMS (7.2% vs 16.5% in <2.5-mm group and 3.0% vs 11.5% in 2.5–3.0 mm group).

PES vs SES in small vessels

There is limited information available comparing SES and PES in small vessels. Some trials comparing the outcomes in small vessels are shown in . ISAR-SMART 3Citation36 was the first head-to-head trial comparing SES and PES. This trial showed significantly less angiographic restenosis and TLR rates with the use of SES as compared with PES (11.4% and 6.6% vs 19.0% and 14.7%, respectively).

Table 2 Comparison of angiographic characteristics, death, MI, TLR, and MACEs, following SES and PES

More recently, subgroup analysis of SIRTAX trial,Citation37 which included patients with RVD < 2.75 mm was reported. In patients with small vessel stents, SES reduced MACE by 55% (10.4% vs 21.4%; P = 0.004), mainly driven by a 69% reduction of TLR rate (6.0% vs 17.7%; P = 0.001). However, there were no significant differences with respect to death and MI.

There are two nonrandomized trials that have been performed comparing SES and PES in small coronary vessels. Park et alCitation38 have performed retrospective study involving 197 patients with RVD of nearly 2.45 mm and reported lower angiographic restenosis and TLR rates with the use of SESs (6.7% and 3.3% vs 27.7% and 14.4%; P < 0.01). Another RESEARCH and T-SEARCHCitation39,Citation40 registry adopted a nonrandomized design and evaluated outcomes in patients treated with SES and PES in small coronary vessels (RVD = 2.25 mm). The incidence of 12-month TLR and MACE was numerically more frequent with the use of PES, but they did not reach statistical significance (11.1% and 18.9% vs 6.5% and 9.3%; P = 0.31) and P = 0.06, respectively.

ZES vs PES in small vessels

ENDEAVOR IVCitation41 compared outcomes following implantation of ZES and PES in patients with CAD. In a subgroup analysis involving small vessels (<2.5 mm), there was significantly less target vessel failure at 12 months with the use of ZES as compared with PES (8.3% vs 13.4%; OR = 0.62; 95% CI: 0.37–1.03). However, there was no difference observed between PES and ZES in 2.5–3 mm group (8.8% vs 7.5%; OR = 1.18; 95% CI: 0.69–2.04).

Everolimus-eluting stent vs PES in small vessels

SPRIT IVCitation42 trial has compared the efficacy of everolimuseluting stent (EES) and PES in patients with CAD. In a whole cohort, 1,352 patients were treated for small vessel disease (≤2.75 mm). In this subgroup analysis, there was significantly less occurrence of target vessel failure with the use of EES as compared with PES (3.9% vs 6.8%; OR = 0.57; 95% CI: 0.35–0.91).

Biolimus-eluting stent vs SES in small vessels

LEADERS trialCitation43 compared biolimus-eluting stent (BES) with biodegradable polymer and SES with durable polymer in patients with CAD. In a substudy, investigators have assessed the impact of vessel size on outcomes with these two different stent strategies. All-comer patients (1,707 patients) were included in the study, and comparison was done between vessel size <2.75 mm (50.0% of the total cohort) and >2.75 mm. There was no significant difference between TLR rate (9.6% vs 7.4%; P = 0.26) and MACE (12.1% vs 11.8%; P = 0.89) in both BES and SES arms. However, the TLR rate (9.6% vs 2.6%) and MACE (12.7% vs 7.1%) were significantly higher in small vessels as compared with large vessels in the BES arm.

Summary of impact of DES in small vessels

As evident from the aforementioned studies, the outcomes in small coronary arteries have improved following the introduction of DESs. TLR rate has improved from 20%–30% in the BMS era to around 10% with the usage of various DESs. This still remains higher as compared with other subgroups of CAD. Among the various DESs available, the TLR rate is relatively low (3%–7%) with the use of EES and SES as compared with PES and ZES (8%–13%). These results are biologically plausible because a reduction in luminal diameter by a constant amount of neointimal hyperplasia results in proportionally higher-grade diameter stenosis in small vessels compared with large vessels. Moreover, EES and SES have been invariably shown to afford lower late luminal loss in the trials (0.14 ± 0.41 mm for EES in SPIRIT IIICitation44 trial and 0.16 ± 0.30 mm for SES in pivotal trials) as compared with higher late luminal loss with the use of ZES and PES (0.67 ± 0.49 mm for ZES and 0.42 ± 0.50 mm for PES in ENDEAVOR IV and 0.61 ± 0.46 mm in ENDEAVOR III),Citation45 and a late luminal loss is an established marker to discriminate between different stent types.Citation46

Influence of stent strut thickness on outcomes in small vessels

The mechanical differences of the BMS and the DESs may affect the outcomes in small coronary arteries. Briguori et alCitation47 have shown that strut thickness was an independent predictor of angiographic restenosis in small coronary arteries (RVD = 2.75–2.99 mm); thinner-strutted stents were associated with lower incidence of restenosis than thicker- strutted stents.

Brambilla et alCitation48 have assessed that in a prospective, multicentre registry, the impact of thin-strut chrome–cobalt stent (Mini VISION-strut size 0.081 mm) angiographic and clinical outcomes in small coronary vessels. The average RVD was 2.41 ± 0.41 mm, and the 6 months MACE was 11.6%, death 2.9%, MI 2.9%, and TLR 5.8%. These outcomes are comparable with DES outcomes.

TAXUS ATLAS Small Vessel (SV) programCitation49 is a multicentre study comparing the performance of the thin-strut (0.095 mm) TAXUS Liberte® 2.25-mm stent (Boston Scientific Inc) in small vessels. This study compared the outcomes following implantation of TAXUS Liberte and TAXUS Express (0.132-mm strut size) stents in small vessels. TAXUS Liberte significantly reduced the rate of 9-month angiographic restenosis (18.5% vs 32.7%; P = 0.02) and 12-month TLR (6.1% vs 16.9%; P = 0.0003) as compared with TAXUS Express stents. This TAXUS liberte (Atom™; Boston Scientific Inc) stent was specially designed for small vessels (<2.25 mm) and was approved by the US Food and Drug Administration for the treatment. This stent has 27-inch thinner struts, and the unique, uniform repeating cell geometry supports even drug distribution and delivery. This stent also has lowest stent crossing and tip profile making it a more trackable stent delivery system.

Safety issue following small vessel DES implantation

Since DESs were approved, these devices have been implanted in a large number of patients with CAD and with off-label indications including small vessels. Their use seems to be safe. However, issue of stent thrombosis has been one of the concerns. Although BMS implantation in small vessels had been previously cited as a risk factor for stent thrombosis,Citation50Citation52 improved techniques of optimal stent deployment and dual antiplatelet therapy appear to have largely resolved this problem so that the risk of stent thrombosis of BMS in small vessel stenting now seems to be similar to that in large vessel stenting.Citation53,Citation54

DES implantation in small vessels may increase the risk of stent thrombosis. The incidence of stent thrombosis in small vessel DES implantation has not been shown to differ between PES and BMS or SES. In a subanalysis conducted in the TAXUS VCitation29 clinical trial, both acute and late stent thrombosis rates were similar between PES and BMS (0.9% vs 1.1% and 1.0% vs 1.1%; P = 1.00 and P = 1.00, respectively). In ISAR- SMART 3Citation36 trial and a study by Park et alCitation38 no acute stent thrombosis was reported in both SES and PES arms, whereas there was no information about late stent thrombosis in either trial. In a subanalysis of the RESEARCH and T-SEARCHCitation39,Citation40 registries, 2.2% of patients had acute stent thrombosis in the PES arm; no thrombosis was observed in the SES arm (P = 0.21). No late stent thrombosis occurred in either arm. In SIRTAXCitation37 trial there was similar stent thrombosis seen in SES and PES arms in small vessels (2.2% vs 2.7%; P = 0.75, respectively). This was similar to the stent thrombosis rates seen in large vessels (1.9% vs 3.3%; P = 0.35) with the use of SES and PES, respectively. Lee et alCitation34 have reported the late stent thrombosis of 0.4% up to 20 months following implantation of SES in small vessels.

However, late stent thrombosis is multifactorial, and the limited data available so far have shown similar rates with the use of DES.

Factors predictive of adverse outcome after DES in small vessels

Revascularization of small coronary arteries has been problematic because of high risk of restenosis. Before the introduction of DES, the restenosis rate in these lesions was very high, ranging from 30%–50%, and there was only marginal benefit from stent implantation. However, DESs have markedly reduced the risk of restenosis, and their benefits are evident in small vessels. Small vessel disease still remains because of DES failure as compared with other groups. Lee et alCitation34 have evaluated the predictive factors of adverse outcome following implantation of SES in a consecutive series of 1,092 patients with reference vessel size <2.8 mm. They have reported significant correlation between restenosis rate and lesion length, and restenosis rate was highest (29.4%) in patients with very long lesions (>60 mm). Multivariate analysis showed that lesion length (OR = 1.04; 95% CI: 1.02–1.05; P < 0.001) and in-stent restenotic lesions (OR = 3.38; 95% CI: 1.80–6.35; P < 0.001) were significant predictors of restenosis, but diabetes was not shown to be significant predictor of restenosis (OR = 0.79, 95% CI: 0.47–1.33, P = 0.378).

Small vessel disease has become more common, and the proportion of patients requiring coronary intervention is likely to further increase. However, these patients have higher clinical and angiographic restenosis rates following small vessel angioplasty as shown in the aforementioned data. The risk of complications and restenosis in many of these patients are increased by other factors, particularly diffuse disease and diabetes mellitus. Although BMS have an advantage in acute gain compared with balloon angioplasty, the former results in more late loss due to neointimal overgrowth. This tissue encroachment presents a greater problem in small than in large vessels because it leaves less room for the lumen in the former. Therefore, using BMS, the long-term results of small vessel stenting were disappointing, and provisional stenting was considered a better option. The use of DESs in small coronary arteries, however, has reduced the rate of restenosis in comparison to BMSs. Various types of DESs are widely used in clinical practice. In head-to-head studies, the SES and EES have shown lower rate of late lumen loss as compared with PES and ZES, suggesting that the former group may be more effective in preventing restenosis in high-risk patients.

Future directions in small vessel coronary angioplasty

A number of stent-related properties, including stent configuration, strut thickness, and stent coating, can affect the long-term clinical outcome. Many trials have shown that the mesh-wire and coil-related stent designs suffer from a significantly higher risk of restenosis compared with the tubular or multicellular stent design. Studies have shown that stent with less strut–strut intersections (Multilink stent) is associated with the most favorable angiographic and clinical outcomes. Citation55 The design, material composition, surface features of the stent, and stent deployment technique affect strongly the acute performance of the stent, risk of stent thrombosis, degree of vascular response, and subsequent risk of in-stent restenosis.

The availability of new, highly biocompatible, and more radiovisible alloys with the same if not superior tensile strength than stainless steel will enable the production of low metal density stents that may further improve the anatomical and clinical outcomes of current stainless steel stents. Drug elution from these stent platforms could further improve the outcomes in small vessels. In future, different stent designs and variable strut thickness with different drug elutions need to be tested in small coronary vessels.

Conclusion

We have reviewed balloon angioplasty and the placement of BMS and different DESs in small vessels with respect to clinical outcomes. We could conclude that (1) small vessel coronary angioplasty is common; (2) DES considerably reduces the incidence of angiographic restenosis and TLR as compared with BMS; and (3) a trend is observed with regard to better angiographic and clinical outcomes of EES and SES over PES and ZES but with similar safety profile. Stent design and strut thickness along with DES influence the outcomes in small coronary vessels similar to other coronary lesions.

Large size, randomized, controlled, double-blinded multi-center trials with long-term follow-up are needed to evaluate different DESs and designs in small vessel CAD.

Disclosure

The author reports no conflicts of interest in this work.

References

  • MackayJMenashGThe Atlas of Heart Disease and StrokeGeneva, SwitzerlandWorld Health Organisation, Centers for disease control and prevention2004
  • WongPLauKWLimYLOestereleSNStent placement for non-STRESS/BENESTENT lesions: a critical reviewCatheter Cardiovasc Interv20005122323311025583
  • O’ConnorNJMortonJRBirmeyerJDOlmsteadEMO’ConnorGTNorthern New England Cardiovascular Disease Study GroupEffect of coronary artery diameter in patients undergoing coronary bypass surgeryCirculation1996936526558640991
  • BautersCHubertEPratAPredictors of restenosis after coronary stent implantationJ Am Coll Cardiol199831129112989581723
  • SerruysPWKayIPDiscoCPeriprocedural quantitative coronary angiography after Palmaz-Schatz stent implantation predicts the restenosis rate at six months: results of a meta analysis of the Belgian Netherlands Stent study (BENESTENT) I, BENESTENT II Pilot, BENESTENT II and MUSIC trials. Multicenter ultrasound stent in coronariesJ Am Coll Cardiol1999341067107410520792
  • SerruysPWde JaegerePKiemeneijFA comparison of balloon- expandable stent implantation with balloon angioplasty in patients with coronary artery disease. Benestent Study GroupN Engl J Med19943314894958041413
  • FischmanDLLeonMBBaimDSA randomised comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. Stent Restenosis Study InvestigatorsN Eng J Med1994331496501
  • MacayaCSerruysPWRuygrokPContinued benefit of coronary stenting versus balloon angioplasty: one year clinical follow up of Benestent Study GroupJ Am Coll Cardiol1996272552618557891
  • BetriuAMasottiMSeraARandomised comparison of coronary stent implantation and balloon angioplasty in the treatment of de novo coronary artery lesions (START): a four-year follow-upJ Am Coll Cardiol1999341498150610551699
  • KeimeneijFSerruysPWMacayaCContinued benefit of coronary stenting versus balloon angioplasty: five-year clinical follow up of Benestent-I trialJ Am Coll Cardiol2001371598160311345371
  • KastratiASchomigADirschingerJA randomised trial comparing stenting with balloon angioplasty in small vessels in patients with symptomatic coronary artery disease. ISAR-SMART Study Investigators. Intracoronary stenting or angioplasty for restenosis reduction in small arteriesCirculation20001022593259811085962
  • ParkSWLeeCWHongMKRandomized comparison of coronary stenting with optimal balloon angioplasty for treatment of lesions in small coronary arteriesEur Heart J2000211785178911052843
  • KoningREltchaninoffHCommeauPStent placement compared with balloon angioplasty for small coronary arteries: in-hospital and 6-month clinical and angiographic resultsCirculation20011041604160811581136
  • MoerRMyrengYMolstadPStenting in small coronary arteries (SISCA) trialA randomised comparison between balloon angioplasty and the heparin-coated stentJ Am Coll Cardiol2001381598160311704369
  • MauriLOravEJKuntzRELate loss in lumen diameter and binary restenosis for drug-eluting stent comparisonCirculation20051113435344215967844
  • MeirBSousaEGuagliumiGSirolimus-eluting coronary stents in small vesselsAm Heart J20061511017el1019e1
  • EllisSGPopmaJJLasalaJMRelationship between angiographic late loss and target lesion revascularization after coronary stent implantation: analysis from the TAXUS-IV trialJ Am Coll Cardiol2005451193120015837248
  • AgostoniPBiondi-ZoccaiGLGasparaniGLIs bare-metal stenting superior to balloon angioplasty for small vessel coronary artery disease? Evidence from a meta-analysis of randomised trialsEur Heart J20052688188915681573
  • DoucetCSchalijMJVrolixMCStent placement to prevent restenosis after angioplasty in small coronary arteriesCirculation20011042029203311673341
  • HaudeMKonorzaTFKalninsUErglisAHeparin-coated stent placement for the treatment of stenosis in small coronary arteries of symptomatic patientsCirculation20031071265127012628946
  • KinsaraAJNiaziKPatelIAmoudiOEffectiveness of stents in small coronary arteriesAm J Cardiol20039258458712943880
  • HanekampCKoolenJBonnierHRandomised comparison of balloon angioplasty versus silicon carbon-coated stent implantation for de novo lesions in small coronary arteriesAm J Cardiol2004931233123715135695
  • GarciaEGomez-ReicoMMorenoRStent reduces restenosis in small vessels. Results of the RAP study [abstract]J Am Coll Cardiol20013717A
  • MuramatsuTIwasakiKInouNEfficacy of coronary stenting versus balloon angioplasty in vessels of diameter less than 3.0 mm and less than 2.5 mm: CHIVAS investigators [abstract]Am J Cardiol20029096H
  • HausleiterJKastratiAMehelliJA randomised trial comparing phosphorylcholine-coated stents with balloon angioplasty in small coronary arteries in patients with symptomatic coronary artery disease. The ISAR-SMART 2 Trial [abstract]Circulation2003108IV-569
  • RodriguezAERodriguez-AlemparteMJPereiraCFFinal angiographic and clinical results of the Latin America small vessel randomised study (LASMAL trial) [abstract]J AM Coll Cardiol20034126A12570940
  • RodriguezALASMAL-II: a prospective randomised trial of phosphorylcholine- coated stenting vs balloon angioplasty in small vessels in the diabetic patient (featured lecture)TCT2003 Available from: http://www.tctmd.com/expert-presentationsAccessed 2004 Jun 27
  • StoneGWEllisSGCoxDAOne-year clinical results with the slow-release, polymer based, paclitaxel-eluting TAXUS stent: the TAXUS-IV trialCirculation20041091942194715078803
  • StoneGWEllisSGCanonLComparison of a polymer-based paclitaxel-eluting stent with a bare metal stent in patients with complex coronary artery disease: a randomised controlled trialJAMA20052941215122316160130
  • DawkinsKDGrubeEGuagliumiGClinical efficacy of polymer- based paclitaxel-eluting stents in the treatment of complex, long coronary artery lesions from a multicenter, randomised trial: support for the use of drug-eluting stents in contemporary clinical practiceCirculation20051123306331316286586
  • HolmesDRJrLeonMBMosesJWAnalysis of 1-year clinical outcomes in the SIRIUS trial: a randomised trial of a sirolimus-eluting stent versus a standard stent in patients at high risk for coronary restenosisCirculation200410963464014769686
  • PopmaJJLeonMBMosesJWQuantitative assessment of angiographic restenosis after sirolimus-eluting stent implantation in native coronary arteriesCirculation20041103773378015596568
  • ArdissinoDCavalliniCBramucciESirolimus-eluting vs uncoated stents for prevention of restenosis in small coronary arteries: a randomised trialJAMA20042922727273415585732
  • LeeCWSuhJLeeSWFactors predictive of cardiac events and restenosis after sirolimus-eluting stent implantation in small coronary arteriesCatheter Cardiovasc Interv20076982182517191211
  • FajadetJWijnsWLaarmanGJRandomised, double-blind, multicenter study of the Endeavor zotralimus-eluting phosphorylcholine- encapsulated stent for the treatment of native coronary artery lesions: clinical and angiographic results of the ENDEAVOR II TrialCirculation200611479880616908773
  • MehilliJDibraAKastratiARandomised trial of paclitaxeland sirolimus-eluting stents in small coronary vesselsEur Heart J20062726026616401670
  • TogniMEberSWidmerJImpact of vessel size on outcome after implantation of sirolimus-eluting and paclitaxel-eluting stents: a sub-group analysis of the SIRTAX TrialJ Am Coll Cardiol2007501123113117868802
  • ParkKHParkSWHongMKComparison of the effectiveness of sirolimus- and paclitaxel-eluting stents for small coronary artery lesionsCatheter Cardiovasc Interv20066758959416547937
  • Rodriguez-GranilloGAValgimigliMGarcia-GarciaHMOne-year clinical outcome after coronary stenting of very small vessels using 2.25 mm sirolimus- and paclitaxel-eluting stents: a comparison between the RESEARCH and T-SEARCH registriesJ Invasive Cardiol20051740941216079445
  • TanimotoSDaemenJTsuchidaKTwo-year clinical outcome after coronary stenting of small vessels using 2.25 mm sirolimus- and paclitaxel-eluting stents: insight into the RESEARCH and T-SEARCH registriesCatheter Cardiovasc Interv2007699410317139687
  • LeonMBMauriLPopmaJJA randomised comparison of the Endeavor zotrolimus-eluting stent versus the TAXUS paclitaxel-eluting stent in de novo native coronary lesions: 12 month outcomes from the ENDEAVOR IV trialJ Am Coll Cardiol20105554355420152559
  • StoneGWRizveANewmanWEverolimus-eluting versus paclitaxel-eluting stents in coronary artery diseaseN Engl J Med20101621663166420445180
  • WykrzykowskaJJSerruysPWOnumaYImpact of vessel size on angiographic and clinical outcomes of revascularization with biolimus-eluting stent with biodegradable polymer and sirolimus-eluting stent with durable polymer the LEADERS trial substudyJACC Cardiovasc Interv2009286187019778775
  • StoneGWMideiMNewmanWRandomised comparison of everolimus-eluting and paclitaxel-eluting stents: two-year clinical follow-up from the clinical evaluation of the XIENCE V everolimuseluting coronary stent system in the treatment of patients with de novo native coronary artery lesions SPIRIT III trialCirculation200911968068619171853
  • KandzariDLeonMBPopmaJJComparison of zotralimus- eluting and sirolimus-eluting stents in patients with native coronary artery disease: a randomised controlled trial. ENDEAVOR III trialJ Am Coll Cardiol2006482440244717174180
  • MauriLOravEJCandiaSCRobustness of late lumen loss in discriminating drug-eluting stents across variable observational and randomised trialsCirculation20051122833283916267256
  • BriguoriCSaraisCPagnottaPIn-stent restenosis in small coronary arteries: impact of strut thicknessJ Am Coll Cardiol20024040340912142103
  • BrambillaNMoriciNBedogniFThin strut chrome-cobalt stent implantation for treatment of de-novo lesions in small coronary vessels: results of the RISCO Italian registry utilizing the Mini VISION coronary stent platformJ Cardiovasc Med (Hagerstown)20091085285819550353
  • TurcoMAOrmistonJAPomaJJReduced risk of restenosis in small vessels and reduced risk of myocardial infarction in long lesions with the new thin-strut TAXUS Liberte stent: 1-year results from the TAXUS ATLAS programJACC Cardiovasc Interv2008169970919463387
  • KarrillonGJMoriceMCBenvenisteEIntracoronary stent implantation without ultrasound guidance and with replacement of conventional anticoagulation by antiplatelet therapy. 30-day clinical outcome of the French multicenter registryCirculation199694151915278840839
  • MakKHBelliGEllisSGSubacute stent thrombosis: evolving issues and current conceptsJ Am Coll Cardiol1996274945038557927
  • MoussaIDi MarioCReimersBSubacute stent thrombosis in the era of intravascular ultrasound-guided coronary stenting without antocoagulation: frequency, predictors and clinical outcomeJ Am Coll Cardiol1997296128996288
  • AkiyamaTMoussaIReimersBAngiographic and clinical outcome following coronary stenting of small vessels: a comparison with coronary stenting of large vesselsJ Am Coll Cardiol199832161016189822086
  • LauKWDingZPSimLLClinical and angiographic outcome after angiography-guided stent placement in small coronary vesselsAm Heart J200013983083910783217
  • KastratiADirschingerJBoekstegersPInfluence of stent design on 1-year outcome after coronary stent placement: a randomised comparison of five stent types in 1,147 unselected patientsCatheter Cardiovasc Interv20005029029710878624