Abstract
Kvarstein, B. The Use of an Electronic Particle Counter (Celloscope 101) for Counting Leucocytes. Scand. J. din. Lab. Invest. 19, 196-202 1967. A method for counting leucocytes with an electronic particle counter using a Cetrimide solution as hemolyzing diluent was examined.
The electronic particle counter gives reliable results with a coefficient of variation of 4.4 per cent.
The leucocyte counts can be done immediately, or at any time within 24 hours with the same result.
Rise in temperature led to a slight fall in white cell counts.
The procedure did not require siliconized glass tubes.