Some aspects of the clinical use of thermodilution in measuring cardiac output: With particular reference to the Swan-Ganz thermodilution catheters

Pages 383-388 | Received 23 Mar 1977, Accepted 25 Nov 1977, Published online: 08 Jul 2009
 

Abstract

Accuracy of the thermodilution method in measuring cardiac output was tested in a model and against Fick in ten patients. Reproducibility was determined from 573 thermodilution curves. The influence of catheter position, catheter irrigation, post-injection aspiration, and the substitution of rectal temperature for central blood temperature was also studied. The thermodilution method showed very good correlation to direct measurements in the model (corr. coeff. 1.00) and the Fick method in patients (corr. coeff. 0.97). Mean standard deviation for cardiac outputs calculated from individual thermodilution curves was 5%. Post-injection aspiration and variation in catheter tip location did not influence the results. Catheter irrigation during recordings and the use of rectal temperature for TB resulted in underestimation of cardiac output.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.