Lactate determination with the lactate analyser LA 640: a critical study

, , &
Pages 145-149 | Received 24 May 1984, Accepted 17 Sep 1984, Published online: 08 Jul 2009
 

Abstract

Using the new method of sample preparation developed by the manufacturer in 1979, the lactate analyser LA 640 now provides reliable measurements of blood lactic acid from blood micro-samples of 20 μ1 within 2 min. Reproducibility is comparable to that obtained by spectrophotometric assay with ±0.2 mmol.l-1 for 95% confidence limits. High accuracy was also obtained with 95% linked confidence limits within ±0.19 mmol.l-1 and ±0.46 mmol.l-1 for blood lactate concentrations of from 2–4 and 7–9 mmol.l-1 respectively. Due to a lactate concentration increase of 0.5 mmol.l-1.h-1 in whole blood kept at 2 0 d`C, the blood must be immediately diluted in the buffer haemolysing solution after withdrawal. This diluted sample can then be stored at room temperature without alteration, provided the concentration in the dilution is lower than 0.8 mmol.l-1. For serial successive measurements, a new sample can be tested before the electrode current reaches its background value.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.