Inadequate algorithm: a cause for ‘incorrect pH 7.40 correction’ in ionized calcium analysers

, , , &
Pages 561-565 | Received 10 Oct 1988, Accepted 04 Feb 1989, Published online: 08 Jul 2009
 

Abstract

As a preliminary step in a study of the effects of calcium ligands on the pH standardization of ionized calcium (Ca2+) measurements in blood, the slope of logCa2+=f(pH) linear relationship characterizing the pH-sensitive calcium buffer capacity of the specimen was investigated in 12 serum pools on three different instruments. The pH 7.40 correction line should be horizontal. This was the case for the ICA-2 but not for the ICA-1 and the NOVA-8. The discrepancy was caused by an incorrect setting of the built-in slope correction factor in the ICA-2; coincidentally, its value was close to the effective slopes of the serum pools used in the study. Thus, the ‘abnormal’ behaviour of the ICA-1 and the NOVA-8 was caused by an inadequacy of the built-in algorithm to the characteristics of our serum pools.

These findings lead us to reconsider the use of a fixed and constant correction factor to normalize actual ionized calcium values.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.