Abstract
An easy method to measure blood volume is clinically needed. We used carbonmonoxide (CO) and the OSM3 to measure circulating hemoglobin and blood volume with the indicator dilution principle. 50 mL of CO was administered into a closed rebreathing system and taken up via the lungs, and the amount of hemoglobin in the blood was calculated from the increase in carboxyhemoglobin fraction after 10 min. Blood volume was calculated by division with the concentration of hemoglobin. We observed that the absorption spectrum of carboxyhemoglobin (COHb) depends on pH and pCO2, which must be controlled when very accurate spectrophotometry is necessary. The bias is 3 % COHb per pH unit during calibration of the OSM3, which may be permissible for patients with CO poisoning, but not for the present purpose. With this in mind the method is very accurate, precise and simple.