An in vitro characterization of a silicone tonometer system for synchronous measurement of tissue oxygen- and carbon dioxide tension

, &
Pages 317-325 | Received 24 Apr 1992, Accepted 09 Dec 1992, Published online: 08 Jul 2009
 

Abstract

A new tonometry system for continuous and synchronous measurement of tissue oxygen- and carbon dioxide tension is described and characterized in vitro. The tonometer system consists of an O2 and CO2 permeable silicone tube continuously flushed with isotonic saline by an injection pump. When the saline passes through the tonometer tube it equilibrates with O2 and CO2 outside the tube. The oxygen- and carbon dioxide tension of the flushing solution after passage of the tonometer tube are measured by a transcutaneous combined oxygen/carbon dioxide electrode (E5280 Radiometer A/S, Copenhagen, Denmark), connected to the tonometer tube via an airtight polycarbonate chamber. In order to characterize the tonometer system in vitro the tonometer tube was submerged in a test chamber containing isotonic saline, 33°C to 41°C, with varying partial pressures of O2 and CO2. For various lengths of the tonometer and flushing rates through the tonometer the partial pressures of oxygen and carbon dioxide in the flushing solution (pO2eq and pCO2eq), after passage through the tonometer were recorded and compared to the known partial pressures of oxygen and carbon dioxide in the test chamber solution (pO2 test and pCO2test). POzeq and pCO2eq approached pO2test and pCO2test, when the length of the tonometer was increased, and the flushing rate through the tonometer was decreased. The relative differences (D) between pO2eq and pCO2eq at the one hand and pO2test and pCO2test at the other hand were calculated, and equilibration curves were constructed. The time taken for the flushing solution to equilibrate with the oxygen and carbondioxide in the test chamber decreased with increasing flushing rate, but was independent of the length of the tonometer, when the length was less than 10cm. A flushing rate of 4mlh−1 typically used in the clinic, gave a 90% equilibration time for Oz and CO2 of 73 s and 130 s respectively.

When the tonometer system is used in vivo, the equilibration curves enables conversion of the measured pOzeq and pCO2eq values to pO2test and pCO2test values, which in vivo corresponds to the oxygen- and carbon dioxide tensions of the tissue surrounding the silicone tube.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.