A Disposable Uroflowmeter for Recording of Maximum Flow Rate

Accuracy and Clinical Acceptability

, , &
Pages 303-306 | Received 13 May 1982, Published online: 15 Feb 2010
 

Abstract

A disposable urinary flowmeter (Peakometer®) was evaluated in regard to accuracy of measurement and acceptability by patients. Laboratory tests of measurement accuracy were made with five Peakometers as recordings of maximum flow rate and voided volume. The accuracy was found to be acceptable. A minimum voided volume of 200 ml is recommended for a reasonably exact recording of maximum flow rate. Ten patients aged 28–86 years tested the acceptability of the Peakometer. All were able to use the apparatus correctly and to register maximum flow rate and voided volume of urine in accordance with written standardized instructions. The disposable flowmeter can be recommended for patients who live far from their hospital, thus saving time and transport costs. Clinics with relatively few patients requiring urinary flow studies may use the disposable flowmeters, instead of investing in relatively expensive electronic apparatus.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.