Abstract
The development of a coaxial TEM (transverse electromagnetic) deep-heating, non-contacting applicator employing two axially spaced concentric sleeves is described which has electrostatic characteristics and has been named the ESA. Thermal data obtained with the FDA/CDRH elliptic-shaped human torso phantom (with fat overlay) showed nearly uniform heating (±10%) throughout the inner cross-section. Saline tank measurements on a torso cross-section confirmed similar SAR uniformity. Animal experiments with a pig, both with and without blood flow, verified deep-heating and suggested that some preferential central heating occurred. The absence of excessive surface heating indicated that the major portion of the E-field excitation is axially aligned. The non-contacting applicator does not require a water bolus, and experiments showed that moderate patient movement had minor effect on performance.