Abstract

The SAR patterns were determined for four commercially available microwave (915 MHz) interstitial applicators. Values of SAR were determined using a miniature (3 mm diameter) implantable isotropic electric field probe or a custom 0.25 mm diameter fluoroptic temperature probe. These are the smallest such probes that are currently available. Similar radial variation of SAR was found at the axial position of the gap in the outer conductor for each applicator. Electric field probe measurements are much faster and avoid some of the errors caused by the rapid spatial variation of SAR with interstitial applicators. The major limitation on the electric field probe is its size; it is larger than the applicators being tested.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.