A Method for Dose Calculation for High Energy Photon Beams Based on Measurements Performed at Reference Depth

&
Pages 425-433 | Received 29 Apr 1992, Accepted 09 Dec 1992, Published online: 08 Jul 2009
 

Abstract

An algorithm is proposed to calculate the dose per monitor unit at any point along the beam axis for blocked or unblocked fields. The proposed formalism takes into consideration the beam measurements performed at the reference depth as recommended by most dosimetry protocols (5 cm or 10 cm depending on the beam quality). The only parameters which cannot be measured at the reference depth are the peak scatter factors, but they appear only as a ratio of two peak scatter factors for two slightly different field sizes. A correction factor is proposed when the distance of one shielding block to the beam axis is smaller than 5 cm. An agreement better than 1% has been obtained between calculations and measurements in the range of beam qualities, distances to the source and field sizes defined by typical collimators or shielding blocks, usually encountered in clinical practice.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.