Reliability and validity of a new ankle electrogoniometer

, &
Pages 350-355 | Received 18 Nov 2009, Accepted 13 May 2010, Published online: 29 Jun 2010
 

Abstract

Objectives: To establish reliability, accuracy and concurrent validity of a new electrogoniometer (SG110A, Biometrics) for measuring ankle movement.

Methods: The new electrogoniometer, placed at the lateral malleolus (A-perp), was compared to a customary electrogoniometer (SG110, Biometrics), placed along the Achilles tendon (A-para). Concurrent recordings were made with motion analysis. Common dance movements were performed by 17 dancers on two days. Intraclass correlation coefficients (ICC) and standard error of measurement (SEM) were calculated for sagittal angular displacements.

Results: Instrument and intra-tester reliability of A-perp was high (r ≥ 0.985, r ≥ 0.979, respectively). Criterion and concurrent validity were also high (r ≥ 0.954, r ≥ 0.937). SEM ranged from 1.43° to 6.99°.

Conclusions: The high ICC values establish acceptable reliability and validity. Subjects found A-perp more comfortable. Although both electrogoniometers are acceptable (comparable SEM) to measure extreme ankle motions, the greater comfort and durability of the new A-perp sensor is advantageous.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.