The Spectralab-M quadrupole medical mass spectrometer

Pages 265-272 | Published online: 09 Jul 2009
 

Abstract

A practical universal gas concentration measuring instrument would have great advantages as the basis of many physiological measurements. The mass spectrometer offers such an instrument but difficulty in calibration has made it inaccessible to all but the most technically minded department. This paper reports measurements on the Spectralab-M, a medical mass spectrometer that offers easy use and calibration by using microprocessor control.

The performance of the machine configured for respiratory measurements gave an accuracy better than 0.03 ± 0.1% vol/vol. Configured for anaesthetic gas measurements the instrument gave an accuracy better than 0.07 ± 0.36% vollvol for all gases. In both cases no systematic alinearities were obserued. The speed of response of the machine was measured with a standard heated stainless-steel capillary inlet of 1.5 m. Response times (10–90./) of less than 100 ms were obtained on all gases measured with typical lag times of 500 ms at a flow rate of 15 ml min-1.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.