Comparative study of two laser Doppler blood flowmeters

, &
Pages 243-249 | Published online: 09 Jul 2009
 

Abstract

A new infrared laser Doppler blood flow instrument (Moor MBF3D) was evaluated using an in vitro model allowing measurements over a range of flow velocities and concentrations. The responses correlated well (r = 0·96, p < 0·01) with those obtained simultaneously using a Perimed PF3 laser Doppler instrument. The different processing bandwidths of the instruments were investigated and the wideband mode of operation is recommended for flow measurements where there may be fast moving red blood cells (rbcs). The infrared instrument is capable of dual-channel operation, and the two channels are shown to respond almost identically for similar changes in blood flow through the in vitro model (r = 0·999, p < 0·01). The main advantage of the dual-channel instrument is that continuous measurements may be made simultaneously at two different skin sites allowing dynamic flow responses to be compared.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.