An evaluation of the accuracy of flow measurements using magnetic resonance imaging (MRI)

, , , &
Pages 170-176 | Published online: 09 Jul 2009
 

Abstract

A study was undertaken to examine the accuracy and precision of the measurement of flow by magnetic resonance imaging (MRI) with consideration to the equipment and patient related parameters that might be encountered in vivo. For this purpose, test objects were devised consisting of PVC tubing, in which the internal diameter simulated the size of the arteries in the body. The design of the lest objects ensured that steady laminar flow was obtained in the sections being imaged. The calibration study suggests that, using MRI. flow can be measured in vitro with systematic error of better than 7·0 ± 5·0% and random error of better than 7·5%. In general, flow measurements obtained from MRI were found to correlate well with the known flows. However, the results indicated that there are prerequisite conditions for the validity of the measurements, such as the selection of appropriate flow pulse sequences and velocity limits.

Measurements taken at vertically 40 mm away from the isocentre of the magnetic field -were significantly different (p < 0·01)

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.