A method of testing the compliance of 15 and 22 mm conical connectors on heat and moisture exchangers to ISO 5356–1: 1987

, &
Pages 147-150 | Published online: 09 Jul 2009
 

Abstract

Heat and moisture exchangers (HMEs) are used to warm and humidify the gas delivered to patients during anaesthesia and in intensive care. Typically, HMEs are connected between the tracheal tube or catheter mount and the breathing system. This point of connection has been identified as having a high risk of disconnection. To reduce the possibility of disconnection, the International Standard for HMEs specifies that the connectors on an HME shall comply with ISO 5356–1:1987, the International Standard for cones and sockets. A simple method is described to test whether or not the connectors comply with this standard. This method is used when evaluating HMEs for the Medical Devices Agency (MDA). Results are quoted from the MDA series Evaluation. The majority of connectors tested were found to comply with the standard. The test method is also suitable for testing conical connectors on other devices.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.