8
Views
2
CrossRef citations to date
0
Altmetric
Original Article

Blood Compatibility of Biomaterials: Further Evaluation of the Lindholm Test

, , &
Pages 21-30 | Published online: 11 Jul 2009
 

Abstract

The Lindholm test, a modified whole blood clotting time test designed to permit estimation of the blood compatibility of biomaterials, has been analyzed statistically with regard to sources of variance and desired sample sizes. Of five materials tested for compatibility with blood, only Cuprophane had an among-donor component of variance significantly different from zero. Analysis indicated that, under certain conditions, performance of 11 Lindholm tests upon each of two materials can be expected to furnish sufficient data for an observed average difference of 5 min in clotting time to be significant at the 5% level; performance of 36 pairs of tests gives a 98% probability of detecting a true difference of 5 min when testing at the 5% level of significance.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.