Incorporation of Intraoperative Computerized Tomography in a Newly Developed Spinal Navigation Technique

, , , &
Pages 18-27 | Received 10 Aug 1999, Accepted 02 Dec 1999, Published online: 06 Jan 2010
 

Abstract

Objective: We report on the first successful incorporation of intraoperative computerized tomography in spinal navigation procedures.

Materials and Methods: All operations were performed with the aid of a Tomoscan M mobile CT system (Philips Medical Systems, Eindhoven, The Netherlands). The system comprises a mobile gantry, a mobile patient examination table, and a mobile workstation. Three different navigation systems were used: the EasyGuide (Philips Medical Systems, Eindhoven, The Netherlands), the Surgical Tool Navigator (Zeiss, Oberkochen, Germany), and a prototype of an ultrasound navigation system developed by ourselves (TVS GbR, Chemnitz, Germany). All surgical operations were performed with the patient positioned on the mobile table of the CT system. Following dorsal preparation of the vertebral region, the surgeon implanted small titanium screws in the vertebrae to serve as fiducial markers. Image data acquisition and image-to-patient registration were performed after implantation of the marker screws. The pedicle screws were inserted using the navigation system, and the position of each implant was confirmed by intraoperative CT scans.

Results: To date, 35 patients with various spine disorders have been operated upon using the technique described, and 161 pedicle screws have been inserted. There were no misplacements of pedicle screws. In three (1.9%) cases we observed lateral perforation (maximum 2 mm) of the lateral pedicle wall. We achieved a registration error of 0.85 mm (RMSE) ±0.42 mm (SD).

Conclusions: The combination of intraoperative computerized tomography and spinal navigation allows easy navigation with a high application accuracy of 0.8 mm ± 0.4 mm (SD) at the target point (measured in experiments with a plastic spine model). The possibility of performing an intraoperative quality check on demand markedly improves the safety of these procedures.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.