835
Views
19
CrossRef citations to date
0
Altmetric
Original Article

Automated Fiducial Marker Detection for Patient Registration in Image-Guided Neurosurgery

, , , , &
Pages 17-23 | Received 02 Feb 2002, Accepted 17 Mar 2003, Published online: 06 Jan 2010
 

Abstract

Objective: The registration of applied fiducial markers within the preoperative data is often left to the surgeon, who has to identify and tag the center of each marker. This is both time-consuming and a potential source of error. For this reason, the development of an automated procedure was desirable. In this study, we have investigated the accuracy of a software algorithm for detecting fiducial markers within the navigation data set. The influence of adjustable values for accuracy and threshold on the sensitivity and specificity of the detection process, as well as the time gain, was investigated.

Patients and Methods: One hundred MP-RAGE MRI data sets of patients with different pathologies who were scheduled for image-guided surgery were used in this study. A total of 591 applied fiducial markers were to be detected using the algorithm of the software VVPlanning 1.3 (BrainLAB, Heimstetten, Germany) on a Pentium II standard PC. The size value of a marker in the y-direction is called “accuracy” and depends on the slice thickness. “Threshold” describes the gray level above which the algorithm starts searching for pixel clusters. The threshold value was changed stepwise on the basis of a constant “accuracy” value. The “ccuracy” value was changed on the basis of that threshold value at which all markers were detected correctly.

Results: The time needed for automatic detection varied between 12 s and 25 s. An optimum value for adjustable marker size was found to be 1.1 mm, with 8 undetected markers (1.35%) and 7 additionally detected structures (1.18%) out of 591. The mean gray level (Threshold) for all data sets above which marker detection was correct was 248.9. The automatic detection of markers was good for higher gray levels, with 11 missed markers (1.86%). Starting the algorithm at lower gray levels led to a decreased incidence of missed markers (0.17%), but increased the incidence of additionally detected structures to 27.92%.

Conclusion: The automatic marker-detection algorithm is a robust, fast and objective instrument for reliable fiducial marker registration when used with optimum settings for both threshold and accuracy.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.