Computerized ellipse method for measuring acetabular version after total hip replacement – a precision study using synthetic and real radiographs

, , , , &
Pages 195-200 | Received 12 Jan 2012, Accepted 19 Feb 2013, Published online: 25 Mar 2013
 

Abstract

Background: Previous work by our group to address the problem of acetabular positioning based on 2D methods resulted in the development of a measurement method with better precision – Liaw's version. This method may help the early diagnosis of acetabular loosening. In the present study, we hypothesized that our computerized ellipse method could improve the precision of measuring acetabular version.

Methods: We developed our Elliversion software to measure acetabular version. Using total hip replacement (THR) Simulator, 96 radiographs were synthesized with random femoral inclination and 5° to 52° version, half with the femoral head included and half without. These synthetic radiographs and 28 real radiographs were measured with both Elliversion and the trigonometric method twice by one of the authors with a one-week interval between measurements. We then calculated the difference in the repeated measurements. Student's t-test was used for statistical analysis of the measuring error and inter-measurement difference.

Results: In the precision study, for synthetic radiographs including the femoral head, the ellipse method was significantly better than the trigonometric method (p < 0.01). For synthetic radiographs without the femoral head, there was no significant difference between the ellipse method and the trigonometric method (p = 0.19). As for the repeated measurements, for synthetic radiographs including the femoral head, the ellipse method was significantly better than the trigonometric method (p = 0.001), whereas for synthetic radiographs without the femoral head, there was no significant difference between the two methods (p = 0.17). For real radiographs, there was no significant difference between the two measuring methods (p = 0.12). However, if we excluded the four poor-quality radiographs, there was a significant difference between the two measuring methods (p = 0.04).

Discussion: We developed a computerized ellipse method for measuring acetabular version on synthetic radiographs and good-quality real radiographs. This method is characterized by its superior precision as compared to the trigonometric method. With the 2D standardized method (Liaw's version), improving the precision of measurement will help earlier diagnosis of acetabular loosening.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.