Thrombotic Obstruction of the Björk-Shiley Aortic Valve Prosthesis. Report of Four Cases

, , , , &
Pages 255-257 | Received 30 Oct 1978, Published online: 12 Jul 2009
 

Abstract

We present four cases of late thrombotic obstruction of the Björk-Shiley aortic valve prosthesis. The incidence of this complication reported in the literature to vary from 0.7% to 5%. In our experience of a total of 623 Björk-Shiley aortic valve replacements, we observed this complication in only 4 patients (0.6%). Two of them had poorly controlled anticoagulation therapy. The clinical presentation was subacute in all four patients. Surgical treatment, thrombectomy and debridement, was performed in all of them. The diagnosis was made upon abrupt and progressive onset of dyspnoea, physical examination data and echocardiographic and radioscopic findings. Angiocardiographic studies were needed in two patients.