Comparison of conventional versus three-dimensional ultrasound in fetal renal pelvis measurement and their potential prediction of neonatal uropathies

, , , &
Pages 2494-2499 | Received 30 Mar 2015, Accepted 02 Sep 2015, Published online: 02 Oct 2015
 

Abstract

Objective: To establish a threshold value for fetal renal pelvis dilatation measured by automatic volume calculation (SonoAVC) in the third trimester of pregnancy to predict neonatal uropathies, and to compare these results with conventional antero–posterior (AP) measurement, fetal kidney 3D volume and renal parenchymal thickness.

Methods: In a prospective cohort study, 125 fetuses with renal pelvis AP diameter of ≥5 mm both at 20 weeks of gestation and in the third trimester, underwent an additional 3D volume measurement of the fetal kidney in the third trimester. Receiver operating characteristic (ROC) curves for establishing threshold values for fetal renal pelvis volume, AP measurement, fetal kidney volume and renal parenchymal thickness to predict neonatal uropathies were analyzed. Also, sensitivity, specificity, area under the curve (AUC) and likelihood ratios were calculated.

Results: A cut-off point of 1.58 cm³ was identified in the third trimester of pregnancy (AUC 0.865 (95% CI 0.789–0.940), sensitivity 76.3%, specificity 87.4%, LR+ 6.06, LR− 0.27) for measurements with SonoAVC. A cut-off value of 11.5 mm was established in the third trimester of pregnancy (AUC 0.828 (95% CI 0.737–0.918), sensitivity 71.1%, specificity 85.1%, LR+ 4.77, LR− 0.34) for the conventional AP measurement. A cut-off point for fetal kidney volume was calculated at 13.29 cm³ (AUC 0.769 (95% CI 0.657–0.881), sensitivity 71%, specificity 66%, LR+ 2.09, LR− 0.44). For renal parenchymal thickness, a cut-off point of 8.4 mm was established (AUC 0.216 (95% CI 0.117–0.315), sensitivity 31.6%, specificity 32.6%, LR+ 0.47, LR− 2.10).

Conclusion: This study demonstrates that 3D fetal renal pelvis volume measurements and AP measurements both have a good and comparable diagnostic performance, fetal renal volume a fair accuracy and renal parenchymal thickness a poor accuracy in predicting postnatal renal outcome.

Declaration of interest

The authors report no conflict of interest.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.