Abstract
This paper proposes a segmented average method that combines the derived non-linear response (DNLR) and linear-averaging methods to increase the reproducibility of transiently evoked otoacoustic emissions (TEOAEs). In this method, the DNLR method is used at the 2.5-P-ms segment of each TEOAE signal, and the linear-averaging method is used from P to 20 ms. This exploits the reduced-artefact advantage of the DNLR method and the high-reproducibility advantage of the linear-averaging method. To evaluate the performance of the proposed method, 26 normal ears were tested, with P selected as 8 ms. Each TEOAE signal obtained was decomposed into three scales using the discrete wavelet transform. Then the similarities and differences between the signals obtained with this method and with the DNLR method were assessed. The results show that the average reproducibility of the TEOAE signals increases from 78.1% to 88.7% when applying the proposed method.
Sumario
Este artículo propone un método segmentado promedio que combina la respuesta derivada no lineal (DNLR) y un método de promediación lineal para aumentar la reproducibilidad de las emisiones otoacústicas evocadas por transientes (TEOAE). En este método, el método DNLR se utiliza en el segmento 2.5-P-ms de cada señal de TEOAE, y el método de promediación lineal se usa de P a 20 ms. Esto aprovecha la ventaja de reducción de artefactos del método DNLR y la ventaja de alta reproducibilidad del método de promediación lineal. Para evaluar el rendimiento del método propuesto, se evaluaron 26 oídos normales, con una P seleccionada de 8 ms. Cada señal obtenida de TEOAE fue descompuesta en tres escalas utilizando la transformación discreta de onda pequeña. Luego se evaluaron las similitudes y diferencias entre las señales obtenidas con este método y con el método DNLR. Los resultados muestran que la reproducibilidad promedio de las señales de TEOAE se incrementan del 78.1% al 88.7% cuando se aplica el método propuesto.