Abstract
The main results in Poulsen & Oakley (2009) are given as the equivalent threshold sound pressure level, ETSPL, measured in an acoustic coupler specified in IEC 60318-3. These results are all correct. The ETSPL values for the ear simulator specified in IEC 60318-1 were calculated from acoustic laboratory measurements and unfortunately a simple sign error has been found that changes two of the columns in Table 3 of Poulsen & Oakley (2009).
Acknowledgements
The author is grateful to Volker Schirkonyer, Technische Universität München, Germany, who made me aware of this mistake.
Declaration of interest: The author reports no conflicts of interest. The author alone is responsible for the content and writing of the paper.