Aim: Objective of this study is to develop a robust multi-matrix LC–MS/MS for the quantitation of endogenous short-chain fatty acids (SCFA) biomarkers in human plasma and urine. Methods: Developed method utilizes stable isotope-labeled internal standards, high-throughput derivatization procedure for sample preparation and LC–MS/MS analysis using multiple reaction monitoring transitions in positive electrospray ionization mode. Results: Surrogate matrix method was used for quantitation. Accuracy, precision, parallelism, curve linearity, derivatization efficiency, stability and recovery were all evaluated, and the results were well within the acceptable criteria. Conclusion: SCFA levels in human plasma and urine of inflammatory bowel disease patients versus non-disease subjects were quantified and compared by LC–MS/MS.
Supplementary data
See online at: https://www.tandfonline.com/doi/10.4155/bio-2018-0241
Financial & competing interests disclosure
The authors have no relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript. This includes employment, consultancies, honoraria, stock ownership or options, expert testimony, grants or patents received or pending, or royalties.