ABSTRACT

We sought to determine effect of signal strength on mean retinal nerve fibre layer (RNFL) using Spectralis optical coherence tomography (S-OCT). Thirty normal subjects (18 female, mean 37.9 years, range 24-61) were imaged with S-OCT using variably dense Bangerter foils to alter Q value (1 unit signal strength = 4 units Q). We found a statistically significant (p < 0.01) linear relationship (R = 0.8643) between Q and RNFL (1 unit decrease Q = 0.181 um mean RNFL increase). Unlike previous observations of Cirrus and Stratus OCT, we found RNFL thickness does not decrease with decreasing signal strength in S-OCT.

Declaration of interest

The authors declare that there are no conflicts of interest. The authors alone are responsible for the writing and content of the article.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 455.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.