ABSTRACT
Purpose: To determine the optimal number of data points required for optimization of formulae for classical lens power calculation.
Methods: A large dataset of preoperative biometric values was used to assess the convergence of formula constants in a number of established intraocular lens power calculation formulae.
Results: In formulae with a single constant, 80–100 clinical data points are sufficient to obtain convergence. The Haigis formula (three constants) requires 200–300 data points although refractive error converges more rapidly.
Conclusions: In all formulae, 80–100 clinical data points are sufficient to achieve a stable mean refractive error.