Corrosion and tribocorrosion study of 316L steel, Ti–6Al–4V and Ti–10Zr–10Nb–5Ta

, &
Pages 112-119 | Received 24 Jun 2018, Accepted 13 Mar 2019, Published online: 28 Mar 2019
 

ABSTRACT

In this paper, corrosion and tribocorrosion tests have been conducted on three biomaterials: Ti–6Al–4V and 316L steel that are well-known commercial biomaterials and Ti–10Zr–10Nb–5Ta, a new biomaterial elaborated in a levitation-melting furnace. Tests have been conducted in five different electrolytes: NaCl (9g/l), Ringer’s solution, phosphate buffered saline solution (PBS) with and without an addition of bovine serum albumin (BSA). Two amounts of BSA were used: 1 and 5 g/l. In NaCl solution, Ringer’s solution and PBS solution, Ti–10Zr–10Nb–5Ta presents the best corrosion resistance. Conversely, in the presence of proteins, 316L steel shows better resistance to corrosion in comparison with Ti–10Zr–10Nb–5Ta and Ti–6Al–4V. During tribocorrosion tests, the surfaces have been submitted to friction against an alumina ball at open circuit potential (OCP). Measurement of wear rate at the end of the tests indicates clearly that 316L steel shows better resistance to material removal in comparison with Ti–10Zr–10Nb–5Ta and Ti–6Al–4V.

Disclosure statement

No potential conflict of interest was reported by the authors.

Additional information

Funding

This work was supported by OSEO under contract No. A0907009I and UEFISCIDI under contract No. 026/2010.

Log in via your institution

Log in to Taylor & Francis Online

There are no offers available at the current time.

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.