Novel Method for Implantation of Balloon Expandable Transcatheter Aortic Valve Replacement to Reduce Pacemaker Rate—Line of Lucency Method

, MD, FACC, , MD, , MD, , DOORCID Icon, , MD, , MD, , MD, , MD, , PA, , RN, , MD, , PhDORCID Icon & , MD show all
Pages 427-432 | Received 25 Feb 2020, Accepted 23 Jul 2020, Published online: 17 Sep 2020
 

ABSTRACT

Background

The risk of permanent pacemaker implantation (PPM) following transcatheter aortic valve replacement (TAVR) remains high. Valve implantation depth is an independent predictor of PPM. This study describes the rate of PPM following a novel method for valve positioning during TAVR.

Methods

A retrospective cohort study involved patients undergoing TAVR at our institute between January 2019 and January 2020. Baseline characteristics, procedural data, and clinical outcomes were collected and analyzed. All patients were implanted with the Edwards SAPIEN 3 or SAPIEN 3 Ultra using the radiolucent line on the stented valve, rather than the central marker, to guide positioning. We describe this as the line of lucency method.

Results

In total, 50 patients (96% SAPIEN 3, and 4% SAPIEN 3 Ultra) were included. The median age was 81.5 years, and 30 (60%) patients had existing conduction abnormalities: first degree atrio-ventricular block (18%), right bundle branch block (12%), left bundle branch block (10.2%), and left anterior fascicular block (16%). The modified technique was successful in all cases. No patient required PPM within 30 days. Two patients (4%) had moderate aortic regurgitation on echocardiogram at 30 days; 3 (6%) developed new left bundle branch block. There was no instance of valve embolization, coronary obstruction, or postprocedural death.

Conclusion

A novel technique utilizing the radiolucent line on the bottom of the stented valve during TAVR provided reproducibly high positioning with excellent clinical results. Optimal position of the valve may prevent need for subsequent PPM.

Disclosure statement

The authors have no conflicts of interest to report.

Supplementary material

Supplemental data for this article can be accessed on the publisher’s website.

Additional information

Funding

The authors have no funding to report for this article.

Log in via your institution

Log in to Taylor & Francis Online

There are no offers available at the current time.

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.