Determining optimal cochlear implant electrode array with OTOPLAN

ORCID Icon, , , & ORCID Icon
Pages 748-752 | Received 21 Jul 2023, Accepted 25 Aug 2023, Published online: 22 Sep 2023

Figures & data

Table 1. Patient characteristics evaluated in this study (n = 105).

Figure 1. Histogram with a normal distribution curve (n = 105) of each CDL measured using OTOPLAN ver. 3.0. The average, maximum, and minimum CDLs were 35.1, 39.1 and 31.7 mm, respectively. CDL: cochlear duct length.

Figure 1. Histogram with a normal distribution curve (n = 105) of each CDL measured using OTOPLAN ver. 3.0. The average, maximum, and minimum CDLs were 35.1, 39.1 and 31.7 mm, respectively. CDL: cochlear duct length.

Figure 2. Comparison of CDL by sex (Male: n = 40, female: n = 65), laterality (right: n = 55, left: n = 50) and implantation age. Unpaired t-test (A,B) and linear regression (C) analysis were perfomed. CDL: cochlear duct length.

Figure 2. Comparison of CDL by sex (Male: n = 40, female: n = 65), laterality (right: n = 55, left: n = 50) and implantation age. Unpaired t-test (A,B) and linear regression (C) analysis were perfomed. CDL: cochlear duct length.

Figure 3. Correlation between AID via OTOPLAN preoperatively and that measured by postoperative radiographs. Linear regression analysis was performed. AID: angular insertion depth.

Figure 3. Correlation between AID via OTOPLAN preoperatively and that measured by postoperative radiographs. Linear regression analysis was performed. AID: angular insertion depth.

Figure 4. Image modified from the OTOPLAN results and postoperative radiographs. Each degree means the position of the tip channel (C1) of the electrode according to each selected electrode array. (A) In case #99, FLEX 28 was selected. (B) In case #103, FLEX SOFT was selected. CDL: cochlear duct length; AID: angular insertion depth.

Figure 4. Image modified from the OTOPLAN results and postoperative radiographs. Each degree means the position of the tip channel (C1) of the electrode according to each selected electrode array. (A) In case #99, FLEX 28 was selected. (B) In case #103, FLEX SOFT was selected. CDL: cochlear duct length; AID: angular insertion depth.