Case Report

“Interferent Detect” on the Intoxilyzer® 8000C in an individual with an elevated blood acetone concentration due to ketoacidosis

&
Pages 157-163 | Received 01 Nov 2016, Accepted 10 Mar 2017, Published online: 22 May 2017
 

ABSTRACT

A 72-year-old male was arrested for impaired driving. He had a history of chronic alcoholFootnote1 consumption and limited food intake on the day of his arrest. The Intoxilyzer® 8000C provided the following communication messages during testing: “Interferent Detect”, “Invalid Sample” and then an additional “Interferent Detect.” This information from the Intoxilyzer® 8000C prompted a medical investigation. The individual was taken to the hospital where ketoacidosis was part of the medical diagnosis. Blood samples were collected and analyzed at two hospital laboratories and the Centre of Forensic Sciences. The results from the hospital analyses were serum ethanol concentrations of 172 and 161 mg/100 mL and an acetone concentration of 18 mg/100 mL; the results from the forensic analyses were blood ethanol, acetone, isopropanol and beta-hydroxybutyrate concentrations of 139 mg/100 mL, 21 mg/100 mL, <5 mg/100 mL and 468 mg/L, respectively.

RÉSUMÉ

Un homme de 72 ans a été arrêté pour conduite avec facultés affaiblies. Il avait des antécédents de consommation chronique d'alcoolFootnote2 et s’était peu alimenté le jour de son arrestation. Au cours du test, l'Intoxilyzer® 8000C a délivré les messages suivants: « Interferent Detect », « Invalid Sample », puis de nouveau « Interferent Detect ». Ces informations fournies par l'Intoxilyzer® 8000C ont déclenché une enquête médicale. L'homme a été emmené à l'hôpital, où le diagnostic médical comportait, entre autres, une acidocétose. Des échantillons sanguins ont été prélevés et analysés dans deux laboratoires hospitaliers et au Centre des sciences judiciaires. Les résultats des analyses hospitalières indiquaient des concentrations sériques d’éthanol à 172 et 161 mg/100 mL et une concentration d'acétone à 18 mg/100 mL; les résultats des analyses médicolégales indiquaient des concentrations sanguines d’éthanol, d'acétone, d'isopropanol et de bêta-hydroxybutyrate à respectivement 139 mg/100 mL, 21 mg/100 mL, moins de 5 mg/100 mL, et 468 mg/L.

Disclosure statement

No potential conflict of interest was reported by the authors.

Notes

1. Alcohol and ethanol are used interchangeably throughout the document.

2. Dans ce document, les termes « alcool » et « éthanol » sont indifféremment utilisés.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 102.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.