321
Views
0
CrossRef citations to date
0
Altmetric
Research Article

New Measurements to Resolve Discrepancies in Evaluated Model Parameters of 181Ta

ORCID Icon, , , , , , , & show all
Pages 1155-1165 | Received 30 Mar 2023, Accepted 14 Aug 2023, Published online: 28 Sep 2023

Figures & data

Fig. 1. Discrepancies exist between major evaluations on the mean values of cross sections and where specific cross-section models should be applied. As an example, total cross section σt is shown here. None of the evaluations include uncertainty information.

Fig. 1. Discrepancies exist between major evaluations on the mean values of cross sections and where specific cross-section models should be applied. As an example, total cross section σt is shown here. None of the evaluations include uncertainty information.

TABLE I Experimental Parameters for All Three Experiments*

Fig. 2. Sample-in and sample-out count rates and corresponding backgrounds. The solid lines represent the measured open beam count rate and calculated open background count rate. The dashed lines represent the sample count rate and calculated sample background count rate.

Fig. 2. Sample-in and sample-out count rates and corresponding backgrounds. The solid lines represent the measured open beam count rate and calculated open background count rate. The dashed lines represent the sample count rate and calculated sample background count rate.

Fig. 3. All three transmission measurements. Note the large variance at 5.9 and 35 keV caused by structural Al.

Fig. 3. All three transmission measurements. Note the large variance at 5.9 and 35 keV caused by structural Al.

Fig. 4. ENDF/B-VIII.0 resonance parameters are used in a SAMMY calculation to model the experimental transmission. The end of the ENDF/B-VIII.0 RRR is shown here, where a poor fit to the 330-eV resonance can be seen. Note that sample “1 mm - a” is nearly identical but physically different from sample “1 mm - b” used in the capture measurements.

Fig. 4. ENDF/B-VIII.0 resonance parameters are used in a SAMMY calculation to model the experimental transmission. The end of the ENDF/B-VIII.0 RRR is shown here, where a poor fit to the 330-eV resonance can be seen. Note that sample “1 mm - a” is nearly identical but physically different from sample “1 mm - b” used in the capture measurements.

Fig. 5. Covariance for transmission as a function of energy, linearly propagated from the systematic and statistical uncertainty in the variables of EquationEq. (2).

Fig. 5. Covariance for transmission as a function of energy, linearly propagated from the systematic and statistical uncertainty in the variables of EquationEq. (2)(2) T(ti)=α1C˙Ta(ti)−α2kTaB˙(ti)−B˙0Taα3C˙O(ti)−α4kOB˙(ti)−B˙0O,(2) .

Fig. 6. Capture yield for the 1- and 2-mm samples. Though capture yield should never be greater than unity, the uncertainty band in this case includes values that are >1. This is expected because the normalization introduces 3% uncertainty. Note that the 1-mm sample includes the identifier “b,” which simply indicates that it is a different 1-mm sample from that of the 100-m transmission measurement (which was “1 mm – a”). The difference between “a” and “b” is mostly negligible, but each has unique documentation.

Fig. 6. Capture yield for the 1- and 2-mm samples. Though capture yield should never be greater than unity, the uncertainty band in this case includes values that are >1. This is expected because the normalization introduces ≈3% uncertainty. Note that the 1-mm sample includes the identifier “b,” which simply indicates that it is a different 1-mm sample from that of the 100-m transmission measurement (which was “1 mm – a”). The difference between “a” and “b” is mostly negligible, but each has unique documentation.

Fig. 7. The count rate for the thick Ta transmission validation measurement. The solid lines represent the measured open beam count rate and calculated open background count rate. The dashed lines represent the sample count rate and calculated sample background count rate.

Fig. 7. The count rate for the thick Ta transmission validation measurement. The solid lines represent the measured open beam count rate and calculated open background count rate. The dashed lines represent the sample count rate and calculated sample background count rate.

Fig. 8. (Top) 12-mm Ta and (bottom) thick U sample transmissions as measured at the 35-m detector. Notice the matching structure in each measurement at 132 eV from the Co fixed notch and 5.9, 35, and 88 keV from the Al fixed notch.

Fig. 8. (Top) 12-mm Ta and (bottom) thick U sample transmissions as measured at the 35-m detector. Notice the matching structure in each measurement at 132 eV from the Co fixed notch and 5.9, 35, and 88 keV from the Al fixed notch.

Fig. 9. Correlation, transmission, and standard deviation for the 12-mm Ta sample measured at 35 m. Spikes in uncertainty correspond to low count rates at resonances of Co and Al.

Fig. 9. Correlation, transmission, and standard deviation for the 12-mm Ta sample measured at 35 m. Spikes in uncertainty correspond to low count rates at resonances of Co and Al.

Fig. 10. ENDF/B-VIII.0 resonance parameters were used as input to SAMMY to calculate theoretical neutron transmission and capture yield as a comparison to the measurements. The fit becomes worse as energy increases toward 330 eV.

Fig. 10. ENDF/B-VIII.0 resonance parameters were used as input to SAMMY to calculate theoretical neutron transmission and capture yield as a comparison to the measurements. The fit becomes worse as energy increases toward 330 eV.

TABLE II Resulting χ2 Values from the Fits for Each Sample Measurement and Each Library*

Fig. 11. The experimental energy resolution based on SAMMY 8.1 fit to 238U. The 100-m transmission and 45-m capture yield measurements are capable of resolving resonances beyond 2.5 keV (upper limit of JEFF-3.3), and the 35-m validation transmission is capable of resolving resonances up to approximately 1.5 keV. The width plotted here is taken as the full-width at half-maximum of the resolution function at a given neutron energy. The 100-m transmission with the 3-mm sample is shown for comparison to the resolution function on the inset plot.

Fig. 11. The experimental energy resolution based on SAMMY 8.1 fit to 238U. The 100-m transmission and 45-m capture yield measurements are capable of resolving resonances beyond 2.5 keV (upper limit of JEFF-3.3), and the 35-m validation transmission is capable of resolving resonances up to approximately 1.5 keV. The width plotted here is taken as the full-width at half-maximum of the resolution function at a given neutron energy. The 100-m transmission with the 3-mm sample is shown for comparison to the resolution function on the inset plot.