Effects of hemolysis and lipemia interference on kaolin-activated thromboelastography, and comparison with conventional coagulation tests

, , &
Pages 98-103 | Received 23 May 2016, Accepted 10 Dec 2016, Published online: 20 Feb 2017
 

Abstract

The effects of hemolysis and lipemia on thromboelastography (TEG) analysis have been scarcely evaluated in human samples, and neglected in clinical practice. We aimed to investigate the effects of in vitro mechanical hemolysis and lipemia on TEG analysis and conventional coagulation tests. Twenty-four healthy volunteers were enrolled in the study. Besides the controls, three groups with slight, moderate and severe mechanical hemolysis were constituted according to free hemoglobin (Hb) concentrations of 0.5–1.0, 2.0–6.0 and 7.0–13.0 g/L, respectively; and three groups with mild, moderate and high lipemia were established according to triglyceride concentrations of ∼6.0, ∼12.0, and ∼18.0 mmol/L, respectively. Four TEG parameters, reaction time (R), coagulation time (K), angle (α), and maximum amplitude (MA), were measured alongside conventional plasma tests including prothrombin time (PT), activated partial thromboplastin time (APTT) and fibrinogen (FIB) by mechanical method, and platelet count by optical method. Results showed that the median R and MA values at moderate and severe hemolysis and K at severe hemolysis exceeded respective reference intervals, and were considered unacceptable. Median values of TEG parameters in lipemic samples were all within reference intervals. Bias values of conventional plasma tests PT, APTT and FIB in hemolyzed or lipemic samples were all lower than the Clinical Laboratory Improvement Amendments (CLIA) allowable limits. Bias values of platelet count at moderate to severe hemolysis and lipemia exceeded the CLIA allowable limits. In conclusion, the detection of TEG was in general more affected by mechanical hemolysis than plasma coagulation tests. Pre-analytical variables should be taken into account when unexpected TEG results are obtained.

Disclosure statement

The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper. There was no external funding.

Additional information

Funding

This study was supported by grants from Natural Science Foundation of China (Grant No. 81200353).

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.