64-slice Computed Tomography Assessment of Coronary Artery Stents: a Phantom Study

, , , , , , & show all
Pages 36-42 | Accepted 15 Sep 2005, Published online: 09 Jul 2009
 

Abstract

Purpose: To compare the use of a new 64-slice computed tomography (CT) scanner with 16-slice CT in the visualization of coronary artery stent lumen.

Material and Methods: Eight different coronary artery stents, each with a diameter of 3 mm, were placed in a static chest phantom. The phantom was positioned in the CT gantry at an angle of 0° and 45° towards the z-axis and examined with both a 64-slice and a 16-slice CT scanner. Effective slice thickness was 0.6 mm with 64-slice CT and 1 mm with 16-slice CT. A reconstruction increment of 0.3 mm was applied in both scanners. Image quality was assessed visually using a 5-point grading scale. Stent diameters were measured and compared using paired Wilcoxon tests.

Results: Artificial lumen reduction was significantly less with 64-slice than with 16-slice CT. Average visible stent lumen was 53.4% using 64-slice CT and 47.5% with 16-slice MSCT. Most severe artifacts were seen in stents with radiopaque markers. Using 64-slice CT, image noise increased by approximately 30% due to thinner slice thickness.

Conclusion: Improved spatial resolution of 64-slice CT resulted in superior assessment of coronary artery stent lumen compared to 16-slice CT. However, a relevant part of the stent lumen is still not assessable with multi-slice CT.

Log in via your institution

Log in to Taylor & Francis Online

There are no offers available at the current time.

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.