Phantoms for Cross-Calibration of Dual Energy X-ray Absorptiometry Measurements in Infants

, MD, , MD, PhD, , MD, , PhD, , MD & , MBBS, FACN
Pages 328-332 | Received 19 Nov 2001, Accepted 24 May 2002, Published online: 26 Jun 2013
 

Abstract

Objective: To test the suitability of phantoms to cross-calibrate body composition measurements in small subjects among different dual energy X-ray absorptiometry (DXA) instruments.

Methods: A set of four phantoms with total weights 1520g, 3140g, 4650g and 7490g were made with low cost and easily available materials. Each phantom was made from assembling polyethylene bottles (100 to 1000 mL) filled with either pure olive oil or electrolyte solution in different combinations, and borosilicate tubes (3 and 5 mL) and flexible polypropylene tubing filled with calcium carbonate. Triplicate measurements of each of the four phantoms were performed with three pencil beam densitometers made by the same manufacturer (Hologic Inc., Waltham, MA): two QDR 2000 (University of Liege, Liege, Belgium, and Wayne State University, Detroit, Michigan) and a QDR1500 (University Children’s Hospital, Greifswald, Germany) using infant whole body-scanning mode and analyzed with software V5.73P.

Results: DXA measured total weight, or bone, lean and fat masses, from one center were highly predictive of DXA measurements from the other centers with an adjusted r2 of 0.94 to 1.00, p < 0.001. This was the case whether the measurements from single scan or from average of triplicate scans were used in the analysis.

Conclusions: Systematic corrections, in the form of linear transformations, are possible to allow comparison of clinical data generated from different centers. Different size phantoms can be made to accommodate the varying range of weights and body composition of study subjects.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 53.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 139.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.