Accuracy of PurePulse photoplethysmography technology of Fitbit Charge 2 for assessment of heart rate during sleep

, , &
Pages 927-933 | Received 30 Oct 2018, Accepted 16 Mar 2019, Published online: 16 Apr 2019
 

ABSTRACT

Elevated asleep heart rate (HR) is a risk factor for cardiovascular disease and other-cause morbidity and mortality. We assessed the accuracy of Fitbit Inc. PurePulse® photoplethysmography with reference to three-lead electrocardiography (ECG) in determining HR during sleep. HR of 35 (17 female) healthy adults 25.1 ± 10.6 years of age (mean ± SD) was continuously recorded throughout a single night of sleep. There was no significant difference in asleep HR mean (0.09 beats per minute [bpm], P = 0.426) between Fitbit photoplethysmography and ECG; plus, there was excellent intraclass correlation (0.998) and narrow Bland–Altman agreement range (2.67 bpm). The regression analysis of Bland–Altman plot of mean asleep HR indicates Fitbit tends to slightly overestimate reference values in the lower range of HR (HR < 50 bpm) by 0.51 bpm and slightly underestimate reference values in the higher range of HR (HR > 80 bpm) by 0.63 bpm. Mixed model analysis of epoch-by-epoch (5-min epochs) asleep HR showed significant “U” shape trend (P < 0.001) in amount of Fitbit error (absolute amount of difference between ECG and Fitbit values regardless of overestimation or underestimation) in regard to HR, i.e. smaller error in the medium range of HR (60–80 bpm) and slightly larger error for lower (<60 bpm) and higher (>80 bpm) ranges of HR. However, effect of age, body mass index, gender, and subjective sleep quality measured by Pittsburgh sleep quality index (good/poor sleepers) on error in estimating HR by the Fitbit method was not significant. It is concluded that Fitbit photoplethysmography suitably tracks HR during sleep in healthy young adults.

Declaration of interest

The authors have no conflicts of interest that influence the content of this study.

Additional information

Funding

This research was sponsored by the Robert and Prudie Leibrock Professorship in Engineering at the University of Texas at Austin.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 489.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.