Comparison of Javal-Schiøtz keratometer, Orbscan IIz and Pentacam topographers in evaluating anterior corneal topography

, & ORCID Icon
Pages 476-483 | Received 06 Oct 2020, Accepted 08 Apr 2022, Published online: 25 May 2022
 

ABSTRACT

Clinical relevance

Inter-instrument variation in anterior corneal shape (ACS) measurement has a consequence for ocular clinical practice.

Background

To consider inter-instrument variability in keratometry measurements across the ACS and to explore instrument protocols for determining ACS keratometric analogues (KAs).

Methods

  Mean keratometry/KAs of the right eye were recorded using Javal-Schiøtz keratometer (J-S), Orbscan IIz and Pentacam from 124 subjects (78 females; mean ± SD age: 24.71 ± 6.61 years). Mean radii of curvature were obtained for 1-mm wide annular zones extending up to 6 mm (horizontally) and 4 mm (vertically) from the apex for Orbscan and Pentacam. Zonal mean radius of curvature was calculated by averaging keratometry values for all measured points within the zone.

Results

KA (mean ± SD): Horizontal: Orbscan (7.80 ± 0.31 mm) and J-S (7.82 ± 0.29 mm) were not significantly different (p = 0.072). Pentacam (7.86 ± 0.29 mm) was significantly flatter than J-S (p < 0.001) and Orbscan (p < 0.001). Vertical: Orbscan (7.64 ± 0.31 mm) was significantly steeper than J-S (7.67 ± 0.29 mm, p < 0.005) and Pentacam (7.70 ± 0.29 mm, p < 0.001). Pentacam was significantly flatter than J-S (p < 0.001) and significant flatter than Orbscan across the entire profile (1-4 mm zones horizontal and vertical, p < 0.001).                        

LoAs (CI): J-S/Orbscan: ±0.75 mm (0.05–0.18); J-S/Pentacam: ±0.72 mm (0.01–0.12); Pentacam/Orbscan: ±0.16 mm (0.04–0.08). There was a +0.03 mm positive bias for Orbscan compared to J-S, +0.06 mm positive bias for Orbscan compared to Pentacam and −0.03 mm negative bias for Pentacam compared to J-S.

Conclusions

Algorithms used by Orbscan and Pentacam to solve the peripheral paraxial ray problem produce significantly different KAs. Instrument-specific KAs cannot be used inter-changeably between instruments. Differences in KA between instruments are not significant for ocular surgery, but may influence rigid contact lens fitting. Pentacam measures flatter than Orbscan and J-S.

Disclosure statement

No potential conflict of interest was reported by the author(s).

Additional information

Funding

This study was supported by a collaborative partnership between Cardiff University and Menicon (Japan) Ltd.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Purchase Issue

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 84.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.