REVERSED-PHASE SEQUENTIAL INJECTION LIQUID CHROMATOGRAPHIC METHOD FOR SILDENAFIL ASSAY

, , , &
Pages 2256-2270 | Published online: 18 Nov 2011
 

Abstract

Sequential injection chromatography (SIC) has been recently proposed as an alternative separation technique with potential benefits and some limitations. In the current study, a more pressure resistant selection valve (SV) with additional ports was installed in an SIC assembly. The current SV allowed using a buffer in a Mobile phase (MP) without leakage and analyzing ten solutions with a one-shot run. In addition, a new method for sildenafil assay in tablets, bulk, and synthetic pharmaceutical samples was proposed utilizing the newly developed SIC assembly. Response surface approach was adopted to screen the effect of MP composition on tR, PH, and B and, accordingly, to optimize the method. Sample volume and flow rate were optimized using the univariate approach. The optimum chromatographic conditions were: C18 monolithic column (4.6 × 25 mm), an MP composition of 0.3 M ammonium acetate:acetonitrile (50:50, v/v) at pH 6.8, sample volume 40 µL and flow rate 40 µL/s. UV detection at 240 nm was carried out using miniaturized fiber optic spectrometric devices. A comparative study on some analytical features of the SIC method with those of previous HPLC methods was conducted. The SIC method is more rapid and more reagent-saving and thus safer to the environment. The sample frequency was 40 samples/hr. The total volume of consumed reagents and sample was 2.54 mL.

ACKNOWLEDGMENTS

Dr. Idris thanks the Department of Chemistry, College of Science, King Faisal University, where a part of this work was conducted. Mr. Naheid also thanks the Atomic Energy Council, Sudan Academy of Science, Khartoum, Sudan for permitting him to study for a Master's Degree in Chemistry.

Notes

a sildenafil; b excipients.

Fixed conditions: C18 column (4.6 × 25 mm), UV detection at 240 nm, mobile phase included 0.3 M ammonium acetate, analyte concentration 50 µg/L, and flow rate 40 µL/s.

a concentration (µg/mL).

b number of repetitions.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 583.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.