Table 1. CEN/TS 14793 pro forma for determining and summarising test data in comparing if an alternative method is equivalent to an SRM.
Figure 1. Orthogonal regression of mean AM (TGN M22/FTIR) vs. mean SRM (EN 14790/gravimetry) measurements of water vapor in the NPL Stack Simulator facility in a changing gas matrix of CO/NO/SO2/HCl at 180°C and 12 m s−1 velocity.
![Figure 1. Orthogonal regression of mean AM (TGN M22/FTIR) vs. mean SRM (EN 14790/gravimetry) measurements of water vapor in the NPL Stack Simulator facility in a changing gas matrix of CO/NO/SO2/HCl at 180°C and 12 m s−1 velocity.](/cms/asset/6d18ca33-f1c5-486e-a5c7-e2e1835b740d/uawm_a_1058868_f0001_b.gif)
Table 2. CEN/TS 14793 test of equivalence of TGN M22/FTIR to EN 14790/gravimetry for measurements of water vapour across a concentration range of 0–14 vol%.
Table 3. CEN/TS 14793 test of equivalence of TGN M22/FTIR to EN 14791/impinger train and ion chromatography for measurements of SO2 across a concentration range of 0–75 mg m−3.
Table 4. CEN/TS 14793 test of equivalence of TGN M22/FTIR to EN 14791/impinger train and ion chromatography for measurements of SO2 across a concentration range of 0–200 mg m−3.
Table 5. CEN/TS 14793 test of equivalence of TGN M22/FTIR to EN 14792/chemiluminescence for measurements of NO across a concentration range of 0–200 mg m−3.
Table 6. CEN/TS 14793 test of equivalence of TGN M22/FTIR to EN 14792/chemiluminescence for measurements of NO across a concentration range of 0–300 mg m−3.
Table 7. CEN/TS 14793 test of equivalence of TGN M22/FTIR to EN 15058/NDIR for measurements of CO across a concentration range of 0–75 mg m−3.
Table 8. CEN/TS 14793 test of equivalence of TGN M22/FTIR to EN 15058/NDIR for measurements of CO across a concentration range of 0–100 mg m−3.
Table 9. CEN/TS 14793 test of equivalence of TGN M22/FTIR to EN 1911/impinger train and ion chromatography for measurements of HCl across a concentration range of 0–15 mg m−3.
Table 10. CEN/TS 14793 test of equivalence of TGN M22/FTIR to EN 1911/impinger train and ion chromatography for measurements of HCl across a concentration range of 0–60 mg m−3.
Table 11. Summary of CEN/TS 14793 test of equivalence of TGN M22/FTIR to the European Standard Reference Methods for the Monitoring of Emissions from Stacks and Flues of H2O, SO2, NO, CO, and HCl.