940
Views
2
CrossRef citations to date
0
Altmetric
Original Articles

Hyperspectral photoacoustic spectroscopy of highly-absorbing samples for diagnostic ocular imaging applications

ORCID Icon & ORCID Icon

Figures & data

Figure 1. HS-PAS setup for (a) ocular measurement with eye and OAC standard and (b) validation.

Figure 1. HS-PAS setup for (a) ocular measurement with eye and OAC standard and (b) validation.

Figure 2. (a) UST and (b) photodiode signals of OAC standard under 500 nm excitation.

Figure 2. (a) UST and (b) photodiode signals of OAC standard under 500 nm excitation.

Figure 3. (a) PV(λ) and (b) FV(λ) of the OAC standard and sample.

Figure 3. (a) PV(λ) and (b) FV(λ) of the OAC standard and sample.

Figure 4. (a) Assumed behavior of light in OAC standard. Experimental setup to measure (b) transmittance, and (c) reflectance of OAC standard.

Figure 4. (a) Assumed behavior of light in OAC standard. Experimental setup to measure (b) transmittance, and (c) reflectance of OAC standard.

Figure 5. Normalized OAC spectrum of standard μstd_N(λ).

Figure 5. Normalized OAC spectrum of standard μstd_N(λ).

Figure 6. μsam_N(λ) of Red fluorescent microsphere suspension.

Figure 6. μsam_N(λ) of Red fluorescent microsphere suspension.

Figure 7. Validation results using (a) Red, (b) Crimson and (c) Nile Red fluorescent microsphere suspensions.

Figure 7. Validation results using (a) Red, (b) Crimson and (c) Nile Red fluorescent microsphere suspensions.

Figure 8. Measured normalized OAC spectrum of iris in porcine eye sample.

Figure 8. Measured normalized OAC spectrum of iris in porcine eye sample.

Figure 9. (a) Schematic of the eye. B-scan images across the center of the eye using (b) 465 (c) 750, and (d) 870 nm.

Figure 9. (a) Schematic of the eye. B-scan images across the center of the eye using (b) 465 (c) 750, and (d) 870 nm.

Table 1. Selected wavelengths and measured pulse energy.

Supplemental material

uopt_a_1392661_sm6697.docx

Download MS Word (47.8 KB)