Interocular symmetry of retinal nerve fibre layer thickness in healthy eyes: a spectral‐domain optical coherence tomographic study

, MD, , MD, , MD PhD, , MD & , MD PhD
Pages 550-554 | Received 23 Dec 2013, Accepted 19 May 2014, Published online: 15 Apr 2021
 

Abstract

Purpose

This study was performed to investigate the interocular symmetry of peripapillary retinal nerve fibre layer (RNFL) thickness, as measured by Cirrus high‐definition spectral‐domain optical coherence tomography (OCT), in healthy eyes. A wide range of subject ages and refractive errors was examined.

Methods

The retinal nerve fibre layer thickness was measured in 1,234 healthy eyes from 617 subjects using OCT. Interocular differences (right eye minus left eye) in global area and quadrant nerve fibre layer thicknesses were measured. The effect of age and refractive error on interocular nerve fibre layer thickness difference was also examined.

Results

Means (and standard deviations) of subjects' ages and average subject refractive errors were 36.4 ± 19.8 years (range: five to 80 years) and ‐2.67 ± 2.91 (range: ‐14.13 to +5.75 D), respectively. Cutoff limits for normal interocular nerve fibre layer thickness differences (2.5th and 97.5th percentiles of normative data) in the global area and in the superior, nasal, inferior and temporal quadrants were 9.5, 23.0, 15.6, 20.0 and 22.6 μm, respectively. Mean interocular retinal nerve fibre layer thickness differences in global area and in superior, nasal, inferior and temporal quadrants were 0.7 (p < 0.001), ‐3.9 (p < 0.001), 2.6 (p < 0.001), 1.1 (p = 0.007) and 3.0 (p < 0.001) μm, respectively. Interocular nerve fibre layer thickness differences were not significantly correlated with age or refractive error (average of right and left eyes, both p > 0.05).

Conclusions

Significant interocular differences in peripapillary retinal nerve fibre layer thickness exist in healthy eyes, particularly in the superior quadrant. This finding should be considered when comparing retinal nerve fibre layer thickness between right and left eyes.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Purchase Issue

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 84.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.