Measuring ocular aberrations and image quality in peripheral vision with a clinical wavefront aberrometer

& , PhD
Pages 212-222 | Received 31 Oct 2008, Accepted 27 Feb 2009, Published online: 15 Apr 2021
 

Abstract

Background:  Clinical aberrometers are accurate, robust instruments for measuring wavefront aberrations for foveal vision but several practical concerns arise when performing aberrometry of the peripheral field. The purpose of this study was to evaluate these concerns experimentally using a physical eye model.

Methods:  A physical model eye was constructed to provide a stable test case that resembled a human eye. Aberrations were measured with a commercial Shack‐Hartmann aberrometer along lines‐of‐sight ranging from zero to 45° of eccentricity. Commercial software for wavefront reconstruction and Zernike analysis was adapted for use with elliptical entrance pupils encountered off‐axis.

Results:  Pupil dimensions estimated from the array of Shack‐Hartmann spots captured by the wavefront sensor followed geometrical optics predictions up to 30° eccentricity. With careful attention to detail, aberration analysis over an elliptical pupil was verified with alternative software. Retinal image quality declined slowly as eccentricity increased due to the eye model's spherical aberration. The total RMS computed from Zernike coefficients overestimated the total RMS computed based on the wavefront error of the elliptical pupil.

Conclusion:  Measurement of off‐axis wavefront aberrations of a model eye over a restricted range of eccentricities is possible with the COAS clinical wavefront aberrometer and auxiliary lenses to correct astigmatism. When central image quality is good, the off‐axis aberrations will have a powerful effect on peripheral image quality. When central image quality is poor, the additional effect of off‐axis aberrations will be minor.

ACKNOWLEDGEMENTS

We appreciate the assistance of Mr. Glenn Herringshaw and his staff in constructing the model eye used in this study.

GRANTS AND FINANCIAL SUPPORT

This work was supported by NIH grant R01‐EY05109. We thank Wavefront Sciences for access to their analysis software CLAS‐2D.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Purchase Issue

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 84.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.